ATOMIC ABSORPTION 
SPECTROPHOTOMETER (AAS) 
Presentation Author: 
G.TEJASRI 
1
CONTENTS: 
 History 
 Introduction 
 Principle 
 Instrumentation 
 Interferences and control measures 
 Precautions 
 Data analyzing 
 Applications 
 Limitations 
 High sensitivity techniques 
 Graphite furnace 
 Comparison of AAS with GFAAS and ICP-MS. 
 References 2
HISTORY: 
 The first AAS was presented by Walsh and co-workers in 
Melbourne in 1954, was a double beam atomic absorption 
spectrophotometer. 
 Walsh worked with Perkin-Elmer, 
the first AAS instrument developed 
by the company was MODEL 303. 
Alan Walsh 
(1916-1998) 
3
INTRODUCTION 
 Atomic Absorption Spectroscopy (AAS) is a quantitative 
method of analysis that is applicable to many metals and 
a few non-metals. 
 Almost every metallic element can be determined 
quantitatively by using the spectral absorption 
characteristics of atoms. 
 It is a very common technique for detecting and 
measuring concentration of metals in the samples. 
 It can analyze over 62 elements. 
4
THE DETECTABLE ELEMENTS ARE THOSE 
WHICH ARE COLORED IN PINK 
5
PRINCIPLE: 
 “When a beam of monochromatic radiation is passed 
through the atoms of an element, the rate of decrease of 
intensity of radiation is proportional to the intensity of 
incident radiation as well as the concentration of the 
solution.” 
 This technique basically uses the principle that free 
atoms (gas) generated in an atomizer can absorb 
radiation at specific frequency. 
 The atoms absorb UV or visible light and make 
transitions to higher electronic levels. AAS quantifies the 
absorption of ground state atoms in the gaseous state. 
 The analyte concentration is calculated from the amount 
of absorption . 6
INSTRUMENTATION: 
 The basic requirements are: 
(1) a light source; 
(2) a sample cell; and 
(3) a means of specific light measurement. 
7
LIGHT SOURCE: 
 The two most common line sources used in atomic 
absorption are : 
1. ‘‘Hollow Cathode Lamp(HCL)’’ and 
2. ‘‘Electrodeless Discharge Lamp(EDL).’’ 
8
HOLLOW CATHODE LAMP(HCL) 
 Its an excellent, bright line source for most of the 
elements determinable by atomic absorption. 
 Each HCL will have a particular current for optimum 
performance. 
 In general, higher currents will produce brighter 
emission and less baseline noise. As the current 
continues to increase, lamp life may shorten and spectral 
line broadening may occur, resulting in a reduction in 
sensitivity and linear working range. 
9
HCL–CONSTRUCTION 
10
CATHODE AND ANODE 
 The anode and cathode are sealed in a glass cylinder 
normally filled with either neon or argon at low pressure. 
At the end of the glass cylinder is a window transparent 
to the emitted radiation. 
 The cathode of the lamp frequently is a hollowed-out 
cylinder of the metal whose spectrum is to be produced. 
Its constructed from a highly pure metal resulting in a 
very pure emission spectrum. 
 The ‘‘multi-element’’ lamp may provide superior 
performance for a single element or, with some 
combinations, may be used as a source for all of the 
elements contained in the cathode alloy. 
 Anode is made up of tungesten. 11
DISADVANTAGES OF HCL: 
 A finite lifetime – due to depletion of the analyte element 
from the cathode 
 Adsorption of fill gas atoms onto the inner surfaces of 
the lamp – the primary cause for lamp failure 
 Some cathode materials can slowly evolve hydrogen 
when heated – a background continuum emission 
contaminates the purity of the line spectrum of the 
element, resulting in a reduction of atomic absorption 
sensitivity and poor calibration linearity. 
12
ELECTRODELESS DISCHARGE LAMP(EDL) 
 A small amount of the metal or salt of the element for which 
the source is to be used is sealed inside a quartz bulb. 
 This bulb is placed inside a small, self-contained RF generator 
or ‘‘driver’’. When power is applied to the driver, an RF field 
is created. 
 The coupled energy will vaporize and excite the atoms inside 
the bulb, causing them to emit their characteristic spectrum. 
 They are typically much more intense and, in some cases, 
more sensitive than comparable hollow cathode lamps. Hence, 
better precision and lower detection limits where an analysis 
is intensity limited. 
 EDL are available for a wide variety of elements, including 
Sb, As, Bi, Cd, Cs, Ge, Pb, Hg, P, K, Rb, Se, Te, Th, Sn and 
Zn. 13
CONSTRUCTION OF AN EDL 
14
PHOTOMETERS 
 The portion of an atomic absorption spectrometer’s 
optical system which conveys the light from the source 
to the monochromator is referred to as the photometer. 
 Types: single and double beam photometers 
15
SINGLE BEAM PHOTOMETER: 
 It is called ‘‘single-beam’’ because all measurements are based 
on the varying intensity of a single beam of light in a single 
optical path . 
 The primary advantage is – it has fewer components and is 
less complicated than alternative designs. 
 Hence, its easier to construct and less expensive. 
 With a single light path and a minimum number of optical 
components, single-beam systems typically provide very high 
light throughput. 
 The primary limitation is – it provides no means to 
compensate for instrumental variations during an analysis, 
such as changes in source intensity. The resulting signal 
variability can limit the performance capabilities of a single-beam 
system. 16
DOUBLE BEAM PHOTOMETERS 
 It uses additional optics to divide the light from the lamp 
into a ‘‘sample beam’’ (directed through the sample cell) 
and a ‘‘reference beam’’ (directed around the sample 
cell) 
 The reference beam serves as a monitor of lamp intensity 
and the response characteristics of common electronic 
circuitry. 
 Therefore, the observed absorbance, determined from a 
ratio of sample beam and reference beam readings, is 
more free of effects due to drifting lamp intensities and 
other electronic anomalies which similarly affect both 
sample and reference beams. 17
SINGLE BEAM PHOTOMETER 
DOUBLE BEAM PHOTOMETER 
18
OPTICS AND THE MONOCHROMATOR SYSTEM 
 An important factor in determining the baseline noise in an 
atomic absorption instrument is the amount of light energy 
reaching the photomultiplier (PMT). Lamp intensity is 
optimized to be as bright as possible while avoiding line 
broadening problems. 
 Light from the source must be focused on the sample cell and 
directed to the monochromator, where the wavelengths of 
light are dispersed and the analytical line of interest is focused 
onto the detector. 
 Light from the source enters the monochromator at the 
entrance slit and is directed to the grating where dispersion 
takes place. The diverging wavelengths of light are directed 
toward the exit slit. By adjusting the angle of the grating, a 
selected emission line from the source can be allowed to pass 
through the exit slit and fall onto the detector. 
19
MONOCHROMATOR 
20
PRE – MIX BURNER SYSTEM 
 In ‘‘premix’’ design, sample solution is aspirated through 
a nebulizer and sprayed as a fine aerosol into the mixing 
chamber. 
 Here the sample aerosol is mixed with fuel and oxidant 
gases and carried to the burner head, where combustion 
and sample atomization occur. 
 Fuel gas is introduced into the mixing chamber through 
the fuel inlet, and oxidant enters through the nebulizer 
sidearm. 
 Mixing of the fuel and oxidant in the burner chamber 
eliminates the need to have combustible fuel/oxidant in 
the gas lines, a potential safety hazard. 21
 Only a portion of the sample solution is introduced into 
the burner chamber by the nebulizer. 
 The finest droplets of sample aerosol are carried with the 
combustion gases to the burner head, where atomization 
takes place. 
 The excess sample is removed from the premix chamber 
through a drain which uses a liquid trap to prevent 
combustion gases from escaping through the drain line. 
 The inside of the burner chamber is coated with a 
wettable inert plastic material to provide free drainage of 
excess sample and prevent burner chamber ‘‘memory.’’ 
 A free draining burner chamber rapidly reaches 
equilibrium, usually requiring less than two seconds for 
the absorbance to respond fully to sample changes. 22
PRE-MIX BURNER – CONSTRUCTION 
23
IMPACT DEVICES 
 Impact devices are used to reduce droplet size further 
and to cause remaining larger droplets to be deflected 
from the gas stream and removed from the burner 
through the drain. 
 Classified as: impact beads and flow spoilers. 
IMPACT BEAD FLOW SPOILER 
24
IMPACT BEADS 
 Spherical bead made of glass, silica or ceramic. 
 Normally used to improve nebulization efficiency. 
 Positioned directly in the nebulizer spray as it exits the 
nebulizer. 
25
FLOW SPOILERS 
 Normally do not improve nebulization efficiency. 
 The primary use is to remove the remaining large 
droplets from the sample aerosol. 
 They are used in atomic absorption burner systems 
normally are placed between the nebulizer and the burner 
head. 
 They typically have three or more large vanes 
constructed from or coated with a corrosion resistant 
material. 
 Smaller droplets are transported through the open areas 
between the vanes while larger droplets contact the vanes 
and are removed from the aerosol. 
26
NEBULIZERS 
 Classified as : 
HIGH SENSITIVITY NEBULIZER STEEL NEBULIZER 
27
BURNER HEADS AND MOUNTING SYSTEMS 
 Burner heads typically are constructed of stainless steel 
or titanium. All-titanium heads are preferred as they 
provide extreme resistance to heat and corrosion 
 A ten-centimeter single-slot burner head is recommended 
for air-acetylene flames. 
 A special five-centimeter burner head with a narrower 
slot is required when a nitrous oxide-acetylene flame is 
to be used. 
 A ‘‘quick change’’ mount may reduce or eliminate 
entirely the need for realignment of the atomizer when it 
is replaced in the sample compartment. 
28
FLAME SELECTION 
Legend: 
HVG – Hydride Generation 
MVU – Mercury Vapor 
29
STEPS INVOLVED IN FLAME ATOMIZATION 
30
DETECTOR: 
 The intensity of the light is fairly low, so a 
photomultiplier tube (PMT) is used to boost the signal 
intensity 
 A detector (a special type of transducer) is used to 
generate voltage from the impingement of electrons 
generated by the photomultiplier tube 
31
PHOTO MULTIPLIER TUBE 
32
ELECTRONICS 
 Precision in Atomic Absorption Measurements: 
Precision will improve with the period of 
time over which each sample is read. Where analyte 
concentrations are not approaching detection limits, 
integration times of 1 to 3 sec will usually provide 
acceptable precision. When approaching instrument 
detection limits where repeatability is poor, precision can 
be improved by using even longer integration times, up 
to 10 sec. 
33
CALIBRATION OF THE SPECTROMETER: 
 In the linear region, data on as little as one standard and a 
blank may be sufficient for defining the relationship 
between concentration and absorbance. However, 
additional standards are usually used to verify calibration 
accuracy. 
 Where the relationship becomes nonlinear, more 
standards are required for which the accuracy of a 
calibration depends on the number of standards and the 
equations used for calibration. 
34
DATA ANALYZING 
 By comparing the light intensity that 
has passed through the sample with 
that of the same light after it has 
passed through a blank, the 
absorbance is measured. 
 The absorbance of different standard 
solutions of a compound of the 
element are also measured and a 
calibration curve is constructed. 
 Absorbance is plotted against 
concentration. We then use the 
calibration curve to determine the 
unknown concentration 
Concentration (ppm) 
A 
b 
s 
o 
r 
b 
a 
n 
c 
e 
35
INTERFERENCES & 
CONTROL MEASURES 
NON SPECTRAL 
Matrix 
Method of 
Standard 
Additions 
Chemical 
add an excess 
of another 
element or 
compound 
which 
will form a 
thermally 
stable 
compound 
with the 
interferent 
using a 
hotter 
flame. 
Ionization 
adding an 
excess of 
an element 
which 
is very 
easily 
ionized 
SPECTRAL 
Background 
Absorption 
Continuum 
Source 
Background 
Correction 
Zeeman 
Background 
Correction 
36
PRECAUTIONS 
 Exhaust System: AAS flames produce large amounts of heat 
& the resultant fumes & vapours may be toxic. 
 Gas Cylinders: should be located outside of the laboratory in 
a cool well-ventilated area. 
 Flammable Solvents: The combination of flame & solvent is 
a hazardous situation. Always use a solvent with the highest 
flashpoint consistent with the analysis being conducted. Use 
covered containers & the smallest practical volume. 
 Burners: Keep burners clear & do not allow them to block. 
 UV Radiation: Hazardous UV radiation is emitted by flames, 
hollow cathode lamps, analytical furnaces. Never look directly 
at any of these. Operate the AAS with the door or flame shield 
closed and wear appropriate safety glasses. 
37
APPLICATIONS: 
38
LIMITATIONS OF FLAME AAS 
 Determinations of analyte concentrations in the mg/L 
concentration region are routine for most elements. 
 The absorbance depends on the number of atoms in the optical 
path of the spectrometer at a given instant. 
 The nebulization process draws sample into the burner 
chamber at approximately 3-8 ml/min limiting the sample 
introduction rate, and, therefore, the amount of sample 
available for transport to the flame. 
 Only a small fraction of the sample nebulized ever reaches the 
flame, with the remainder being directed to the drain. 
 Sample which is introduced into the flame is resident in the 
light path for only a fleeting moment as it is propelled 
upwards through the flame. 
39
HIGH SENSITIVITY TECHNIQUES: 
 All these limitations in the flame AAS lead to the invent 
of high sensitivity techniques. 
1. The cold vapor mercury technique 
2. Hydride generation technique 
3. Graphite furnace atomic absorption 
40
THE COLD VAPOR MERCURY TECHNIQUE 
 Hg is chemically reduced to the free atomic state by 
reacting the sample with a strong reducing agent like 
SnCl2 or NaBH4 in a closed reaction system. 
 The volatile free Hg is then driven from the reaction 
flask by bubbling air or Ar through the solution. 
 Hg atoms are carried in the gas stream through tubing 
connected to an absorption cell, which is placed in the 
light path of the AAS. 
 Sometimes the cell is heated slightly to avoid water 
condensation but otherwise the cell is completely 
unheated. 
41
ADVANTAGES: LIMITATIONS: 
 Improved sensitivity is 
achieved through a 100% 
sampling efficiency (which 
can be further increased by 
using very large sample 
volumes) 
 All of the mercury in the 
sample solution placed in 
the reaction flask is 
chemically atomized and 
transported to the sample 
cell for measurement. 
 The detection limit for Hg 
approximately 0.02 mg/L 
 Its limited to Hg, since no 
other element offers the 
possibility of chemical 
reduction to a volatile free 
atomic state at room 
temperature. 
 The theoretical limit to this 
technique would be that 
imposed by background or 
contamination levels of 
mercury in the reagents or 
system hardware. 42
HYDRIDE GENERATION TECHNIQUE: 
 Samples are reacted in an external system with a 
reducing agent, usually NaBH4. 
 Gaseous reaction products(volatile hydrides) are then 
carried to a sampling cell in the light path of the AA 
spectrometer. 
 To dissociate the hydride gas into free atoms, the sample 
cell must be heated. 
 The cell is either heated by an air-acetylene flame or by 
electricity 
43
ADVANTAGES LIMITATIONS: 
 For As, Bi ,Ge ,Pb ,Sb ,Se 
,Sn ,Te the detection limits 
well below the mg/L range 
are achievable. 
 The extremely low detection 
limits result from a much 
higher sampling efficiency. 
 Separation of the analyte 
element from the matrix by 
hydride generation is used 
to eliminate matrix-related 
interferences. 
 Its limited to As, Bi ,Ge, Pb, 
Sb, Se, Sn, Te elements only. 
 Results depend on the 
valence state of the analyte, 
reaction time, gas pressures, 
acid concentration, and cell 
temperature. 
 The formation of the analyte 
hydrides is suppressed by a 
number of common matrix 
components, leaving the 
technique subject to 
chemical interference. 
44
GRAPHITE FURNACE ATOMIC ABSORPTION: 
 In this technique, a tube of graphite is located in the 
sample compartment of the AAS, with the light path 
passing through it. 
 A small volume of sample solution is quantitatively 
placed into the tube, normally through a sample injection 
hole located in the center of the tube wall. 
 The tube is heated through a programmed temperature 
sequence until finally the analyte present in the sample is 
dissociated into atoms and atomic absorption occurs. 
45
ADVANTAGES 
 Detection limits fall in the ng/L range for most elements. 
 The sample is atomized in a very short period of time, 
concentrating the available atoms in the heated cell and 
resulting in the observed increased sensitivity. 
 This uses only micro litre sample volumes, which is 
compensated by long atom residence time in the light 
path. 
 Much more automated than the other techniques. 
 Wide applicability. 
 It can determine most elements measurable by AA in a 
wide variety of matrices. 46
GRAPHITE FURNACE TECHNIQUE 
 It’s a non flame technique 
 Main components: 
1. The atomizer 
2. The power supply and 
3. The programmer 
Graphite tube 
• Electrical contacts 
Enclosed water cooled housing 
• Inert purge gas controls 
 The atomizer is located in the sampling compartment of 
the atomic absorption spectrometer, where sample 
atomization and light absorption occur. The power 
supply controls power and gas flows to the atomizer 
under the direction of the programmer, which is usually 
built into the power supply or spectrometer 
47
LONGITUDINALLY-HEATED GRAPHITE FURNACE 
ATOMIZER 
48
A GRAPHITE TUBE FOR A TRANSVERSELY-HEATED 
FURNACE. 
49
 The power supply and programmer perform the 
following functions: 
1. Electrical power control 
2. Temperature program control 
3. Gas flow control 
4. Spectrometer function control 
50
STEPS MAKE UP THE TYPICAL GRAPHITE 
FURNACE PROGRAM 
1) Drying 
2) Pyrolysis 
3) Cool down (optional) 
4) Atomization 
5) Clean out 
6) Cool down 
Legend: 
1) Temperature: final temperature during step 
2) Ramp time: time for temperature increase 
3) Hold time: time for maintaining final 
temperature 
4) Internal gas: gas type and flow rate 51
COMPARISION WITH OTHER TECHNIQUES: 
52
Parameters AAS GFAAS ICP – MS 
Temperature 2300°C – 2700°C 3000°C 6000°C 
Radiation used UV ,VISIBLE UV, VISIBLE ---- 
Detection limit Ppm Ppb Ppt 
Elements applicable 
68+ 50+ 82 
to 
Sample throughput 10-15 sec per element 3-4 min per element All elements <1 
minute 
Sample volume 
required 
Large Very small Very small to medium 
Isotopic analysis No No Yes 
typical consumable 
items and utilities 
required 
acetylene/nitrous 
oxide gases 
(compressed air 
source) 
hollow cathode lamps 
reagents and 
standards 
power 
argon gas 
hollow cathode lamps 
graphite tubes and 
cones 
reagents and 
standards 
power 
cooling water 
argon gas 
quartz torches 
sampling and 
skimmer cones 
reagents and 
standards 
pump tubing 
power 
cooling water 
53
REFERNCES 
 Shriver and Atkins’ Inorganic Chemistry, Fifth Edition 
 Concepts, Instrumentation and Techniques in Atomic Absorption 
Spectrophotometry by Richard D. Beaty and Jack D. Kerber 
 An elementary overview of elemental analysis by thermo elemental 
54
Queries 
55
56

More Related Content

What's hot

Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopyWasla Anum
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopyMahendra G S
 
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...Muhammad Asif Shaheeen
 
Applications of Atomic Absorption Spectrometry (AAS)
Applications of Atomic Absorption Spectrometry (AAS)Applications of Atomic Absorption Spectrometry (AAS)
Applications of Atomic Absorption Spectrometry (AAS)Maharishi Dayanand University
 
Atomic emission spectroscopy PPT
Atomic emission spectroscopy PPTAtomic emission spectroscopy PPT
Atomic emission spectroscopy PPTHumnaMehmood
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopysharmasuriti
 
DIODE ARRAY AND FLUORESCENCE DETECTOR
DIODE ARRAY AND FLUORESCENCE DETECTORDIODE ARRAY AND FLUORESCENCE DETECTOR
DIODE ARRAY AND FLUORESCENCE DETECTORsharmeenkhan15
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopymanishasoral
 
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY.pptx
INDUCTIVELY COUPLED PLASMA MASS  SPECTROMETRY.pptxINDUCTIVELY COUPLED PLASMA MASS  SPECTROMETRY.pptx
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY.pptxSakshi Patil
 
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptFlame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptSachin G
 
GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]
GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]
GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]Shikha Popali
 
Atomic absorption spectrophotometer
Atomic absorption spectrophotometerAtomic absorption spectrophotometer
Atomic absorption spectrophotometerTapeshwar Yadav
 
Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)Sajjad Ullah
 
Atomic absorption spectroscopy (aas)
Atomic absorption spectroscopy (aas)Atomic absorption spectroscopy (aas)
Atomic absorption spectroscopy (aas)karimbscdu
 
Inductively coupled mass spectrometry
Inductively coupled mass spectrometryInductively coupled mass spectrometry
Inductively coupled mass spectrometryAnup Patil
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyManikandan R
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyAnshul2593
 
Atomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and InstrumentsAtomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and InstrumentsVasiliy V. Rosen (Ph.D.)
 

What's hot (20)

Emission spectroscopy
Emission spectroscopyEmission spectroscopy
Emission spectroscopy
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
 
Applications of Atomic Absorption Spectrometry (AAS)
Applications of Atomic Absorption Spectrometry (AAS)Applications of Atomic Absorption Spectrometry (AAS)
Applications of Atomic Absorption Spectrometry (AAS)
 
Atomic emission spectroscopy PPT
Atomic emission spectroscopy PPTAtomic emission spectroscopy PPT
Atomic emission spectroscopy PPT
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
 
DIODE ARRAY AND FLUORESCENCE DETECTOR
DIODE ARRAY AND FLUORESCENCE DETECTORDIODE ARRAY AND FLUORESCENCE DETECTOR
DIODE ARRAY AND FLUORESCENCE DETECTOR
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY.pptx
INDUCTIVELY COUPLED PLASMA MASS  SPECTROMETRY.pptxINDUCTIVELY COUPLED PLASMA MASS  SPECTROMETRY.pptx
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY.pptx
 
Flame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy pptFlame emission spectroscopy and atomic absorption spectroscopy ppt
Flame emission spectroscopy and atomic absorption spectroscopy ppt
 
GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]
GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]
GAS CHROMATOGRAPHY-MASS SPECTROSCOPY [GC-MS]
 
Atomic absorption spectrophotometer
Atomic absorption spectrophotometerAtomic absorption spectrophotometer
Atomic absorption spectrophotometer
 
Atomic absorption spectrometry (aas)
Atomic absorption spectrometry (aas)Atomic absorption spectrometry (aas)
Atomic absorption spectrometry (aas)
 
Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)
 
Atomic absorption spectroscopy (aas)
Atomic absorption spectroscopy (aas)Atomic absorption spectroscopy (aas)
Atomic absorption spectroscopy (aas)
 
Inductively coupled mass spectrometry
Inductively coupled mass spectrometryInductively coupled mass spectrometry
Inductively coupled mass spectrometry
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and InstrumentsAtomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and Instruments
 

Viewers also liked

ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYMusa Khan
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyUlivin Al Farisi
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry Sailee Gurav
 
Jablonski diagram physical chemistry
Jablonski diagram  physical chemistry Jablonski diagram  physical chemistry
Jablonski diagram physical chemistry AZCPh
 
Hydride Generation Atomic Absorption Spectrophotometer
Hydride Generation Atomic Absorption SpectrophotometerHydride Generation Atomic Absorption Spectrophotometer
Hydride Generation Atomic Absorption SpectrophotometerYusrizal Azmi
 
Diatom - Overview
Diatom - OverviewDiatom - Overview
Diatom - OverviewFaiz Taimur
 
Ultracentrifugation
Ultracentrifugation Ultracentrifugation
Ultracentrifugation Aswathi K S
 
Classification of pollen and spores
Classification of pollen and sporesClassification of pollen and spores
Classification of pollen and sporesTahir Shahzad
 
Forensic Palynology
Forensic PalynologyForensic Palynology
Forensic Palynologyvlee86
 
Forensic Anthropology الاستعراف
Forensic Anthropology الاستعراف Forensic Anthropology الاستعراف
Forensic Anthropology الاستعراف Mohammad Ihmeidan
 
Analytical Ultracentrifugation
Analytical UltracentrifugationAnalytical Ultracentrifugation
Analytical UltracentrifugationAkumpaul
 
Forensic palynology
Forensic palynologyForensic palynology
Forensic palynologyfbeer1
 
Types Of Evidence Presentation
Types Of Evidence PresentationTypes Of Evidence Presentation
Types Of Evidence PresentationMaria Donohue
 
7 Forensic Science Powerpoint Chapter 07 Forensic Anthropology
7  Forensic Science Powerpoint Chapter 07 Forensic Anthropology7  Forensic Science Powerpoint Chapter 07 Forensic Anthropology
7 Forensic Science Powerpoint Chapter 07 Forensic AnthropologyGrossmont College
 

Viewers also liked (20)

ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Flame Photometer
Flame PhotometerFlame Photometer
Flame Photometer
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry
 
Jablonski diagram physical chemistry
Jablonski diagram  physical chemistry Jablonski diagram  physical chemistry
Jablonski diagram physical chemistry
 
Hydride Generation Atomic Absorption Spectrophotometer
Hydride Generation Atomic Absorption SpectrophotometerHydride Generation Atomic Absorption Spectrophotometer
Hydride Generation Atomic Absorption Spectrophotometer
 
ICP / AAS Comparison
ICP / AAS ComparisonICP / AAS Comparison
ICP / AAS Comparison
 
Flame emission spectroscopy
Flame emission spectroscopyFlame emission spectroscopy
Flame emission spectroscopy
 
Diatom - Overview
Diatom - OverviewDiatom - Overview
Diatom - Overview
 
Ultracentrifugation
Ultracentrifugation Ultracentrifugation
Ultracentrifugation
 
Classification of pollen and spores
Classification of pollen and sporesClassification of pollen and spores
Classification of pollen and spores
 
Diatoms - Diaonoflagellets
Diatoms - Diaonoflagellets Diatoms - Diaonoflagellets
Diatoms - Diaonoflagellets
 
Forensic Palynology
Forensic PalynologyForensic Palynology
Forensic Palynology
 
Forensic Anthropology الاستعراف
Forensic Anthropology الاستعراف Forensic Anthropology الاستعراف
Forensic Anthropology الاستعراف
 
Analytical Ultracentrifugation
Analytical UltracentrifugationAnalytical Ultracentrifugation
Analytical Ultracentrifugation
 
Study of diatoms
Study of diatomsStudy of diatoms
Study of diatoms
 
Soil Evidence
Soil EvidenceSoil Evidence
Soil Evidence
 
Forensic palynology
Forensic palynologyForensic palynology
Forensic palynology
 
Types Of Evidence Presentation
Types Of Evidence PresentationTypes Of Evidence Presentation
Types Of Evidence Presentation
 
7 Forensic Science Powerpoint Chapter 07 Forensic Anthropology
7  Forensic Science Powerpoint Chapter 07 Forensic Anthropology7  Forensic Science Powerpoint Chapter 07 Forensic Anthropology
7 Forensic Science Powerpoint Chapter 07 Forensic Anthropology
 

Similar to Atomic absorption spectrophotometer

Atomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptxAtomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptxPriyankashah645680
 
Atomic Absorption Spectroscopy (www.Redicals.com)
Atomic Absorption Spectroscopy (www.Redicals.com)Atomic Absorption Spectroscopy (www.Redicals.com)
Atomic Absorption Spectroscopy (www.Redicals.com)Goa App
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometry Atomic absorption spectrophotometry
Atomic absorption spectrophotometry ZeravanAli
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometrysadia muhammad din
 
Atomic Absorption Spectroscopy
Atomic Absorption Spectroscopy Atomic Absorption Spectroscopy
Atomic Absorption Spectroscopy GaneshBhagure1
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometrySurabhi Pal
 
Flame Emission Spectroscopy (FES) Instrumentation
Flame Emission Spectroscopy (FES) InstrumentationFlame Emission Spectroscopy (FES) Instrumentation
Flame Emission Spectroscopy (FES) InstrumentationAnkitVasoya5
 
Atomic Absorption spectroscopy
Atomic Absorption spectroscopyAtomic Absorption spectroscopy
Atomic Absorption spectroscopyVishak Perumal
 
FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPYFLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPYAbhishekPatil387
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyShaikhSaniya2
 
Nidhi singh(004)
Nidhi singh(004)Nidhi singh(004)
Nidhi singh(004)nidhi0309
 
Atomic absorption spectroscopy lec
Atomic absorption spectroscopy lecAtomic absorption spectroscopy lec
Atomic absorption spectroscopy lecZainab&Sons
 
Atomic_Absorption_and_Emission_Spectrometry.pptx
Atomic_Absorption_and_Emission_Spectrometry.pptxAtomic_Absorption_and_Emission_Spectrometry.pptx
Atomic_Absorption_and_Emission_Spectrometry.pptxthedigitalthief2020
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometryMuhammad Amien
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometryMUsama10
 

Similar to Atomic absorption spectrophotometer (20)

Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)
 
Mridu- AAS.pptx
Mridu- AAS.pptxMridu- AAS.pptx
Mridu- AAS.pptx
 
Atomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptxAtomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptx
 
Atomic Absorption Spectroscopy (www.Redicals.com)
Atomic Absorption Spectroscopy (www.Redicals.com)Atomic Absorption Spectroscopy (www.Redicals.com)
Atomic Absorption Spectroscopy (www.Redicals.com)
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometry Atomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
APA seminar
APA seminarAPA seminar
APA seminar
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic Absorption Spectroscopy
Atomic Absorption Spectroscopy Atomic Absorption Spectroscopy
Atomic Absorption Spectroscopy
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
Atomic spectroscopy
Atomic spectroscopy Atomic spectroscopy
Atomic spectroscopy
 
Flame Emission Spectroscopy (FES) Instrumentation
Flame Emission Spectroscopy (FES) InstrumentationFlame Emission Spectroscopy (FES) Instrumentation
Flame Emission Spectroscopy (FES) Instrumentation
 
Atomic Absorption spectroscopy
Atomic Absorption spectroscopyAtomic Absorption spectroscopy
Atomic Absorption spectroscopy
 
FLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPYFLAME EMISSION SPECTROSCOPY
FLAME EMISSION SPECTROSCOPY
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Nidhi singh(004)
Nidhi singh(004)Nidhi singh(004)
Nidhi singh(004)
 
Atomic absorption spectroscopy lec
Atomic absorption spectroscopy lecAtomic absorption spectroscopy lec
Atomic absorption spectroscopy lec
 
Atomic_Absorption_and_Emission_Spectrometry.pptx
Atomic_Absorption_and_Emission_Spectrometry.pptxAtomic_Absorption_and_Emission_Spectrometry.pptx
Atomic_Absorption_and_Emission_Spectrometry.pptx
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometry
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometry
 

More from Nanyang Technological University, Singapore (14)

Ielts – quick tips for over night learners
Ielts – quick tips for over night learnersIelts – quick tips for over night learners
Ielts – quick tips for over night learners
 
Interview hacks
Interview hacksInterview hacks
Interview hacks
 
Video presentation template
Video presentation templateVideo presentation template
Video presentation template
 
World water day
World water dayWorld water day
World water day
 
Air pollution notes
Air pollution notesAir pollution notes
Air pollution notes
 
Environmental impact assessment
Environmental impact assessmentEnvironmental impact assessment
Environmental impact assessment
 
Ecosystem and biodiversity.compressed
Ecosystem and biodiversity.compressedEcosystem and biodiversity.compressed
Ecosystem and biodiversity.compressed
 
Air pollution
Air pollutionAir pollution
Air pollution
 
Swm notes
Swm notesSwm notes
Swm notes
 
7 r's
7 r's7 r's
7 r's
 
A monograph on sw&m
A monograph on sw&mA monograph on sw&m
A monograph on sw&m
 
Trenchless technology
Trenchless technologyTrenchless technology
Trenchless technology
 
Pozzocrete
PozzocretePozzocrete
Pozzocrete
 
Introduction to air pollution
Introduction to air pollutionIntroduction to air pollution
Introduction to air pollution
 

Recently uploaded

办理英属哥伦比亚大学毕业证成绩单|购买加拿大UBC文凭证书
办理英属哥伦比亚大学毕业证成绩单|购买加拿大UBC文凭证书办理英属哥伦比亚大学毕业证成绩单|购买加拿大UBC文凭证书
办理英属哥伦比亚大学毕业证成绩单|购买加拿大UBC文凭证书zdzoqco
 
9873940964 High Profile Call Girls Delhi |Defence Colony ( MAYA CHOPRA ) DE...
9873940964 High Profile  Call Girls  Delhi |Defence Colony ( MAYA CHOPRA ) DE...9873940964 High Profile  Call Girls  Delhi |Defence Colony ( MAYA CHOPRA ) DE...
9873940964 High Profile Call Girls Delhi |Defence Colony ( MAYA CHOPRA ) DE...Delhi Escorts
 
Thessaly master plan- WWF presentation_18.04.24.pdf
Thessaly master plan- WWF presentation_18.04.24.pdfThessaly master plan- WWF presentation_18.04.24.pdf
Thessaly master plan- WWF presentation_18.04.24.pdfTheaMan11
 
Call Girls Ahmedabad 7397865700 Ridhima Hire Me Full Night
Call Girls Ahmedabad 7397865700 Ridhima Hire Me Full NightCall Girls Ahmedabad 7397865700 Ridhima Hire Me Full Night
Call Girls Ahmedabad 7397865700 Ridhima Hire Me Full Nightssuser7cb4ff
 
Dwarka Call Girls 9643097474 Phone Number 24x7 Best Services
Dwarka Call Girls 9643097474 Phone Number 24x7 Best ServicesDwarka Call Girls 9643097474 Phone Number 24x7 Best Services
Dwarka Call Girls 9643097474 Phone Number 24x7 Best Servicesnajka9823
 
Making a Difference: Understanding the Upcycling and Recycling Difference
Making a Difference: Understanding the Upcycling and Recycling DifferenceMaking a Difference: Understanding the Upcycling and Recycling Difference
Making a Difference: Understanding the Upcycling and Recycling DifferenceSwag Cycle
 
9873940964 Full Enjoy 24/7 Call Girls Near Shangri La’s Eros Hotel, New Delhi
9873940964 Full Enjoy 24/7 Call Girls Near Shangri La’s Eros Hotel, New Delhi9873940964 Full Enjoy 24/7 Call Girls Near Shangri La’s Eros Hotel, New Delhi
9873940964 Full Enjoy 24/7 Call Girls Near Shangri La’s Eros Hotel, New Delhidelih Escorts
 
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170simranguptaxx69
 
原版1:1复刻塔夫斯大学毕业证Tufts毕业证留信学历认证
原版1:1复刻塔夫斯大学毕业证Tufts毕业证留信学历认证原版1:1复刻塔夫斯大学毕业证Tufts毕业证留信学历认证
原版1:1复刻塔夫斯大学毕业证Tufts毕业证留信学历认证jdkhjh
 
Düsseldorf U学位证,杜塞尔多夫大学毕业证书1:1制作
Düsseldorf U学位证,杜塞尔多夫大学毕业证书1:1制作Düsseldorf U学位证,杜塞尔多夫大学毕业证书1:1制作
Düsseldorf U学位证,杜塞尔多夫大学毕业证书1:1制作f3774p8b
 
Limnology and Wetland Management 2023 NaRM.pptx
Limnology and Wetland Management 2023 NaRM.pptxLimnology and Wetland Management 2023 NaRM.pptx
Limnology and Wetland Management 2023 NaRM.pptxTesfahunTesema
 
Water Conservation.pptxfgggggggggggggggggggggggggggggggggggggg
Water Conservation.pptxfggggggggggggggggggggggggggggggggggggggWater Conservation.pptxfgggggggggggggggggggggggggggggggggggggg
Water Conservation.pptxfggggggggggggggggggggggggggggggggggggggkhulekanimkhize2
 
World Environment Day PPT slides for Earth DAy arpil 2022
World Environment Day PPT slides for Earth DAy arpil 2022World Environment Day PPT slides for Earth DAy arpil 2022
World Environment Day PPT slides for Earth DAy arpil 2022herebasit
 
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量sehgh15heh
 
Determination of antibacterial activity of various broad spectrum antibiotics...
Determination of antibacterial activity of various broad spectrum antibiotics...Determination of antibacterial activity of various broad spectrum antibiotics...
Determination of antibacterial activity of various broad spectrum antibiotics...Open Access Research Paper
 
Science, Technology and Nation Building.pptx
Science, Technology and Nation Building.pptxScience, Technology and Nation Building.pptx
Science, Technology and Nation Building.pptxgrandmarshall132
 
BIODIVERSITY QUIZ ELIMINATION ROUND.pptx
BIODIVERSITY QUIZ ELIMINATION ROUND.pptxBIODIVERSITY QUIZ ELIMINATION ROUND.pptx
BIODIVERSITY QUIZ ELIMINATION ROUND.pptxROLANARIBATO3
 

Recently uploaded (20)

办理英属哥伦比亚大学毕业证成绩单|购买加拿大UBC文凭证书
办理英属哥伦比亚大学毕业证成绩单|购买加拿大UBC文凭证书办理英属哥伦比亚大学毕业证成绩单|购买加拿大UBC文凭证书
办理英属哥伦比亚大学毕业证成绩单|购买加拿大UBC文凭证书
 
Escort Service Call Girls In Shakti Nagar, 99530°56974 Delhi NCR
Escort Service Call Girls In Shakti Nagar, 99530°56974 Delhi NCREscort Service Call Girls In Shakti Nagar, 99530°56974 Delhi NCR
Escort Service Call Girls In Shakti Nagar, 99530°56974 Delhi NCR
 
9873940964 High Profile Call Girls Delhi |Defence Colony ( MAYA CHOPRA ) DE...
9873940964 High Profile  Call Girls  Delhi |Defence Colony ( MAYA CHOPRA ) DE...9873940964 High Profile  Call Girls  Delhi |Defence Colony ( MAYA CHOPRA ) DE...
9873940964 High Profile Call Girls Delhi |Defence Colony ( MAYA CHOPRA ) DE...
 
Thessaly master plan- WWF presentation_18.04.24.pdf
Thessaly master plan- WWF presentation_18.04.24.pdfThessaly master plan- WWF presentation_18.04.24.pdf
Thessaly master plan- WWF presentation_18.04.24.pdf
 
Call Girls Ahmedabad 7397865700 Ridhima Hire Me Full Night
Call Girls Ahmedabad 7397865700 Ridhima Hire Me Full NightCall Girls Ahmedabad 7397865700 Ridhima Hire Me Full Night
Call Girls Ahmedabad 7397865700 Ridhima Hire Me Full Night
 
Dwarka Call Girls 9643097474 Phone Number 24x7 Best Services
Dwarka Call Girls 9643097474 Phone Number 24x7 Best ServicesDwarka Call Girls 9643097474 Phone Number 24x7 Best Services
Dwarka Call Girls 9643097474 Phone Number 24x7 Best Services
 
Making a Difference: Understanding the Upcycling and Recycling Difference
Making a Difference: Understanding the Upcycling and Recycling DifferenceMaking a Difference: Understanding the Upcycling and Recycling Difference
Making a Difference: Understanding the Upcycling and Recycling Difference
 
9873940964 Full Enjoy 24/7 Call Girls Near Shangri La’s Eros Hotel, New Delhi
9873940964 Full Enjoy 24/7 Call Girls Near Shangri La’s Eros Hotel, New Delhi9873940964 Full Enjoy 24/7 Call Girls Near Shangri La’s Eros Hotel, New Delhi
9873940964 Full Enjoy 24/7 Call Girls Near Shangri La’s Eros Hotel, New Delhi
 
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
Call Girls Sarovar Portico Naraina Hotel, New Delhi 9873777170
 
原版1:1复刻塔夫斯大学毕业证Tufts毕业证留信学历认证
原版1:1复刻塔夫斯大学毕业证Tufts毕业证留信学历认证原版1:1复刻塔夫斯大学毕业证Tufts毕业证留信学历认证
原版1:1复刻塔夫斯大学毕业证Tufts毕业证留信学历认证
 
Düsseldorf U学位证,杜塞尔多夫大学毕业证书1:1制作
Düsseldorf U学位证,杜塞尔多夫大学毕业证书1:1制作Düsseldorf U学位证,杜塞尔多夫大学毕业证书1:1制作
Düsseldorf U学位证,杜塞尔多夫大学毕业证书1:1制作
 
Limnology and Wetland Management 2023 NaRM.pptx
Limnology and Wetland Management 2023 NaRM.pptxLimnology and Wetland Management 2023 NaRM.pptx
Limnology and Wetland Management 2023 NaRM.pptx
 
Hot Sexy call girls in Nehru Place, 🔝 9953056974 🔝 escort Service
Hot Sexy call girls in Nehru Place, 🔝 9953056974 🔝 escort ServiceHot Sexy call girls in Nehru Place, 🔝 9953056974 🔝 escort Service
Hot Sexy call girls in Nehru Place, 🔝 9953056974 🔝 escort Service
 
Water Conservation.pptxfgggggggggggggggggggggggggggggggggggggg
Water Conservation.pptxfggggggggggggggggggggggggggggggggggggggWater Conservation.pptxfgggggggggggggggggggggggggggggggggggggg
Water Conservation.pptxfgggggggggggggggggggggggggggggggggggggg
 
Health Facility Electrification: State of Play
Health Facility Electrification: State of PlayHealth Facility Electrification: State of Play
Health Facility Electrification: State of Play
 
World Environment Day PPT slides for Earth DAy arpil 2022
World Environment Day PPT slides for Earth DAy arpil 2022World Environment Day PPT slides for Earth DAy arpil 2022
World Environment Day PPT slides for Earth DAy arpil 2022
 
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
原版定制copy澳洲詹姆斯库克大学毕业证JCU毕业证成绩单留信学历认证保障质量
 
Determination of antibacterial activity of various broad spectrum antibiotics...
Determination of antibacterial activity of various broad spectrum antibiotics...Determination of antibacterial activity of various broad spectrum antibiotics...
Determination of antibacterial activity of various broad spectrum antibiotics...
 
Science, Technology and Nation Building.pptx
Science, Technology and Nation Building.pptxScience, Technology and Nation Building.pptx
Science, Technology and Nation Building.pptx
 
BIODIVERSITY QUIZ ELIMINATION ROUND.pptx
BIODIVERSITY QUIZ ELIMINATION ROUND.pptxBIODIVERSITY QUIZ ELIMINATION ROUND.pptx
BIODIVERSITY QUIZ ELIMINATION ROUND.pptx
 

Atomic absorption spectrophotometer

  • 1. ATOMIC ABSORPTION SPECTROPHOTOMETER (AAS) Presentation Author: G.TEJASRI 1
  • 2. CONTENTS:  History  Introduction  Principle  Instrumentation  Interferences and control measures  Precautions  Data analyzing  Applications  Limitations  High sensitivity techniques  Graphite furnace  Comparison of AAS with GFAAS and ICP-MS.  References 2
  • 3. HISTORY:  The first AAS was presented by Walsh and co-workers in Melbourne in 1954, was a double beam atomic absorption spectrophotometer.  Walsh worked with Perkin-Elmer, the first AAS instrument developed by the company was MODEL 303. Alan Walsh (1916-1998) 3
  • 4. INTRODUCTION  Atomic Absorption Spectroscopy (AAS) is a quantitative method of analysis that is applicable to many metals and a few non-metals.  Almost every metallic element can be determined quantitatively by using the spectral absorption characteristics of atoms.  It is a very common technique for detecting and measuring concentration of metals in the samples.  It can analyze over 62 elements. 4
  • 5. THE DETECTABLE ELEMENTS ARE THOSE WHICH ARE COLORED IN PINK 5
  • 6. PRINCIPLE:  “When a beam of monochromatic radiation is passed through the atoms of an element, the rate of decrease of intensity of radiation is proportional to the intensity of incident radiation as well as the concentration of the solution.”  This technique basically uses the principle that free atoms (gas) generated in an atomizer can absorb radiation at specific frequency.  The atoms absorb UV or visible light and make transitions to higher electronic levels. AAS quantifies the absorption of ground state atoms in the gaseous state.  The analyte concentration is calculated from the amount of absorption . 6
  • 7. INSTRUMENTATION:  The basic requirements are: (1) a light source; (2) a sample cell; and (3) a means of specific light measurement. 7
  • 8. LIGHT SOURCE:  The two most common line sources used in atomic absorption are : 1. ‘‘Hollow Cathode Lamp(HCL)’’ and 2. ‘‘Electrodeless Discharge Lamp(EDL).’’ 8
  • 9. HOLLOW CATHODE LAMP(HCL)  Its an excellent, bright line source for most of the elements determinable by atomic absorption.  Each HCL will have a particular current for optimum performance.  In general, higher currents will produce brighter emission and less baseline noise. As the current continues to increase, lamp life may shorten and spectral line broadening may occur, resulting in a reduction in sensitivity and linear working range. 9
  • 11. CATHODE AND ANODE  The anode and cathode are sealed in a glass cylinder normally filled with either neon or argon at low pressure. At the end of the glass cylinder is a window transparent to the emitted radiation.  The cathode of the lamp frequently is a hollowed-out cylinder of the metal whose spectrum is to be produced. Its constructed from a highly pure metal resulting in a very pure emission spectrum.  The ‘‘multi-element’’ lamp may provide superior performance for a single element or, with some combinations, may be used as a source for all of the elements contained in the cathode alloy.  Anode is made up of tungesten. 11
  • 12. DISADVANTAGES OF HCL:  A finite lifetime – due to depletion of the analyte element from the cathode  Adsorption of fill gas atoms onto the inner surfaces of the lamp – the primary cause for lamp failure  Some cathode materials can slowly evolve hydrogen when heated – a background continuum emission contaminates the purity of the line spectrum of the element, resulting in a reduction of atomic absorption sensitivity and poor calibration linearity. 12
  • 13. ELECTRODELESS DISCHARGE LAMP(EDL)  A small amount of the metal or salt of the element for which the source is to be used is sealed inside a quartz bulb.  This bulb is placed inside a small, self-contained RF generator or ‘‘driver’’. When power is applied to the driver, an RF field is created.  The coupled energy will vaporize and excite the atoms inside the bulb, causing them to emit their characteristic spectrum.  They are typically much more intense and, in some cases, more sensitive than comparable hollow cathode lamps. Hence, better precision and lower detection limits where an analysis is intensity limited.  EDL are available for a wide variety of elements, including Sb, As, Bi, Cd, Cs, Ge, Pb, Hg, P, K, Rb, Se, Te, Th, Sn and Zn. 13
  • 15. PHOTOMETERS  The portion of an atomic absorption spectrometer’s optical system which conveys the light from the source to the monochromator is referred to as the photometer.  Types: single and double beam photometers 15
  • 16. SINGLE BEAM PHOTOMETER:  It is called ‘‘single-beam’’ because all measurements are based on the varying intensity of a single beam of light in a single optical path .  The primary advantage is – it has fewer components and is less complicated than alternative designs.  Hence, its easier to construct and less expensive.  With a single light path and a minimum number of optical components, single-beam systems typically provide very high light throughput.  The primary limitation is – it provides no means to compensate for instrumental variations during an analysis, such as changes in source intensity. The resulting signal variability can limit the performance capabilities of a single-beam system. 16
  • 17. DOUBLE BEAM PHOTOMETERS  It uses additional optics to divide the light from the lamp into a ‘‘sample beam’’ (directed through the sample cell) and a ‘‘reference beam’’ (directed around the sample cell)  The reference beam serves as a monitor of lamp intensity and the response characteristics of common electronic circuitry.  Therefore, the observed absorbance, determined from a ratio of sample beam and reference beam readings, is more free of effects due to drifting lamp intensities and other electronic anomalies which similarly affect both sample and reference beams. 17
  • 18. SINGLE BEAM PHOTOMETER DOUBLE BEAM PHOTOMETER 18
  • 19. OPTICS AND THE MONOCHROMATOR SYSTEM  An important factor in determining the baseline noise in an atomic absorption instrument is the amount of light energy reaching the photomultiplier (PMT). Lamp intensity is optimized to be as bright as possible while avoiding line broadening problems.  Light from the source must be focused on the sample cell and directed to the monochromator, where the wavelengths of light are dispersed and the analytical line of interest is focused onto the detector.  Light from the source enters the monochromator at the entrance slit and is directed to the grating where dispersion takes place. The diverging wavelengths of light are directed toward the exit slit. By adjusting the angle of the grating, a selected emission line from the source can be allowed to pass through the exit slit and fall onto the detector. 19
  • 21. PRE – MIX BURNER SYSTEM  In ‘‘premix’’ design, sample solution is aspirated through a nebulizer and sprayed as a fine aerosol into the mixing chamber.  Here the sample aerosol is mixed with fuel and oxidant gases and carried to the burner head, where combustion and sample atomization occur.  Fuel gas is introduced into the mixing chamber through the fuel inlet, and oxidant enters through the nebulizer sidearm.  Mixing of the fuel and oxidant in the burner chamber eliminates the need to have combustible fuel/oxidant in the gas lines, a potential safety hazard. 21
  • 22.  Only a portion of the sample solution is introduced into the burner chamber by the nebulizer.  The finest droplets of sample aerosol are carried with the combustion gases to the burner head, where atomization takes place.  The excess sample is removed from the premix chamber through a drain which uses a liquid trap to prevent combustion gases from escaping through the drain line.  The inside of the burner chamber is coated with a wettable inert plastic material to provide free drainage of excess sample and prevent burner chamber ‘‘memory.’’  A free draining burner chamber rapidly reaches equilibrium, usually requiring less than two seconds for the absorbance to respond fully to sample changes. 22
  • 23. PRE-MIX BURNER – CONSTRUCTION 23
  • 24. IMPACT DEVICES  Impact devices are used to reduce droplet size further and to cause remaining larger droplets to be deflected from the gas stream and removed from the burner through the drain.  Classified as: impact beads and flow spoilers. IMPACT BEAD FLOW SPOILER 24
  • 25. IMPACT BEADS  Spherical bead made of glass, silica or ceramic.  Normally used to improve nebulization efficiency.  Positioned directly in the nebulizer spray as it exits the nebulizer. 25
  • 26. FLOW SPOILERS  Normally do not improve nebulization efficiency.  The primary use is to remove the remaining large droplets from the sample aerosol.  They are used in atomic absorption burner systems normally are placed between the nebulizer and the burner head.  They typically have three or more large vanes constructed from or coated with a corrosion resistant material.  Smaller droplets are transported through the open areas between the vanes while larger droplets contact the vanes and are removed from the aerosol. 26
  • 27. NEBULIZERS  Classified as : HIGH SENSITIVITY NEBULIZER STEEL NEBULIZER 27
  • 28. BURNER HEADS AND MOUNTING SYSTEMS  Burner heads typically are constructed of stainless steel or titanium. All-titanium heads are preferred as they provide extreme resistance to heat and corrosion  A ten-centimeter single-slot burner head is recommended for air-acetylene flames.  A special five-centimeter burner head with a narrower slot is required when a nitrous oxide-acetylene flame is to be used.  A ‘‘quick change’’ mount may reduce or eliminate entirely the need for realignment of the atomizer when it is replaced in the sample compartment. 28
  • 29. FLAME SELECTION Legend: HVG – Hydride Generation MVU – Mercury Vapor 29
  • 30. STEPS INVOLVED IN FLAME ATOMIZATION 30
  • 31. DETECTOR:  The intensity of the light is fairly low, so a photomultiplier tube (PMT) is used to boost the signal intensity  A detector (a special type of transducer) is used to generate voltage from the impingement of electrons generated by the photomultiplier tube 31
  • 33. ELECTRONICS  Precision in Atomic Absorption Measurements: Precision will improve with the period of time over which each sample is read. Where analyte concentrations are not approaching detection limits, integration times of 1 to 3 sec will usually provide acceptable precision. When approaching instrument detection limits where repeatability is poor, precision can be improved by using even longer integration times, up to 10 sec. 33
  • 34. CALIBRATION OF THE SPECTROMETER:  In the linear region, data on as little as one standard and a blank may be sufficient for defining the relationship between concentration and absorbance. However, additional standards are usually used to verify calibration accuracy.  Where the relationship becomes nonlinear, more standards are required for which the accuracy of a calibration depends on the number of standards and the equations used for calibration. 34
  • 35. DATA ANALYZING  By comparing the light intensity that has passed through the sample with that of the same light after it has passed through a blank, the absorbance is measured.  The absorbance of different standard solutions of a compound of the element are also measured and a calibration curve is constructed.  Absorbance is plotted against concentration. We then use the calibration curve to determine the unknown concentration Concentration (ppm) A b s o r b a n c e 35
  • 36. INTERFERENCES & CONTROL MEASURES NON SPECTRAL Matrix Method of Standard Additions Chemical add an excess of another element or compound which will form a thermally stable compound with the interferent using a hotter flame. Ionization adding an excess of an element which is very easily ionized SPECTRAL Background Absorption Continuum Source Background Correction Zeeman Background Correction 36
  • 37. PRECAUTIONS  Exhaust System: AAS flames produce large amounts of heat & the resultant fumes & vapours may be toxic.  Gas Cylinders: should be located outside of the laboratory in a cool well-ventilated area.  Flammable Solvents: The combination of flame & solvent is a hazardous situation. Always use a solvent with the highest flashpoint consistent with the analysis being conducted. Use covered containers & the smallest practical volume.  Burners: Keep burners clear & do not allow them to block.  UV Radiation: Hazardous UV radiation is emitted by flames, hollow cathode lamps, analytical furnaces. Never look directly at any of these. Operate the AAS with the door or flame shield closed and wear appropriate safety glasses. 37
  • 39. LIMITATIONS OF FLAME AAS  Determinations of analyte concentrations in the mg/L concentration region are routine for most elements.  The absorbance depends on the number of atoms in the optical path of the spectrometer at a given instant.  The nebulization process draws sample into the burner chamber at approximately 3-8 ml/min limiting the sample introduction rate, and, therefore, the amount of sample available for transport to the flame.  Only a small fraction of the sample nebulized ever reaches the flame, with the remainder being directed to the drain.  Sample which is introduced into the flame is resident in the light path for only a fleeting moment as it is propelled upwards through the flame. 39
  • 40. HIGH SENSITIVITY TECHNIQUES:  All these limitations in the flame AAS lead to the invent of high sensitivity techniques. 1. The cold vapor mercury technique 2. Hydride generation technique 3. Graphite furnace atomic absorption 40
  • 41. THE COLD VAPOR MERCURY TECHNIQUE  Hg is chemically reduced to the free atomic state by reacting the sample with a strong reducing agent like SnCl2 or NaBH4 in a closed reaction system.  The volatile free Hg is then driven from the reaction flask by bubbling air or Ar through the solution.  Hg atoms are carried in the gas stream through tubing connected to an absorption cell, which is placed in the light path of the AAS.  Sometimes the cell is heated slightly to avoid water condensation but otherwise the cell is completely unheated. 41
  • 42. ADVANTAGES: LIMITATIONS:  Improved sensitivity is achieved through a 100% sampling efficiency (which can be further increased by using very large sample volumes)  All of the mercury in the sample solution placed in the reaction flask is chemically atomized and transported to the sample cell for measurement.  The detection limit for Hg approximately 0.02 mg/L  Its limited to Hg, since no other element offers the possibility of chemical reduction to a volatile free atomic state at room temperature.  The theoretical limit to this technique would be that imposed by background or contamination levels of mercury in the reagents or system hardware. 42
  • 43. HYDRIDE GENERATION TECHNIQUE:  Samples are reacted in an external system with a reducing agent, usually NaBH4.  Gaseous reaction products(volatile hydrides) are then carried to a sampling cell in the light path of the AA spectrometer.  To dissociate the hydride gas into free atoms, the sample cell must be heated.  The cell is either heated by an air-acetylene flame or by electricity 43
  • 44. ADVANTAGES LIMITATIONS:  For As, Bi ,Ge ,Pb ,Sb ,Se ,Sn ,Te the detection limits well below the mg/L range are achievable.  The extremely low detection limits result from a much higher sampling efficiency.  Separation of the analyte element from the matrix by hydride generation is used to eliminate matrix-related interferences.  Its limited to As, Bi ,Ge, Pb, Sb, Se, Sn, Te elements only.  Results depend on the valence state of the analyte, reaction time, gas pressures, acid concentration, and cell temperature.  The formation of the analyte hydrides is suppressed by a number of common matrix components, leaving the technique subject to chemical interference. 44
  • 45. GRAPHITE FURNACE ATOMIC ABSORPTION:  In this technique, a tube of graphite is located in the sample compartment of the AAS, with the light path passing through it.  A small volume of sample solution is quantitatively placed into the tube, normally through a sample injection hole located in the center of the tube wall.  The tube is heated through a programmed temperature sequence until finally the analyte present in the sample is dissociated into atoms and atomic absorption occurs. 45
  • 46. ADVANTAGES  Detection limits fall in the ng/L range for most elements.  The sample is atomized in a very short period of time, concentrating the available atoms in the heated cell and resulting in the observed increased sensitivity.  This uses only micro litre sample volumes, which is compensated by long atom residence time in the light path.  Much more automated than the other techniques.  Wide applicability.  It can determine most elements measurable by AA in a wide variety of matrices. 46
  • 47. GRAPHITE FURNACE TECHNIQUE  It’s a non flame technique  Main components: 1. The atomizer 2. The power supply and 3. The programmer Graphite tube • Electrical contacts Enclosed water cooled housing • Inert purge gas controls  The atomizer is located in the sampling compartment of the atomic absorption spectrometer, where sample atomization and light absorption occur. The power supply controls power and gas flows to the atomizer under the direction of the programmer, which is usually built into the power supply or spectrometer 47
  • 49. A GRAPHITE TUBE FOR A TRANSVERSELY-HEATED FURNACE. 49
  • 50.  The power supply and programmer perform the following functions: 1. Electrical power control 2. Temperature program control 3. Gas flow control 4. Spectrometer function control 50
  • 51. STEPS MAKE UP THE TYPICAL GRAPHITE FURNACE PROGRAM 1) Drying 2) Pyrolysis 3) Cool down (optional) 4) Atomization 5) Clean out 6) Cool down Legend: 1) Temperature: final temperature during step 2) Ramp time: time for temperature increase 3) Hold time: time for maintaining final temperature 4) Internal gas: gas type and flow rate 51
  • 52. COMPARISION WITH OTHER TECHNIQUES: 52
  • 53. Parameters AAS GFAAS ICP – MS Temperature 2300°C – 2700°C 3000°C 6000°C Radiation used UV ,VISIBLE UV, VISIBLE ---- Detection limit Ppm Ppb Ppt Elements applicable 68+ 50+ 82 to Sample throughput 10-15 sec per element 3-4 min per element All elements <1 minute Sample volume required Large Very small Very small to medium Isotopic analysis No No Yes typical consumable items and utilities required acetylene/nitrous oxide gases (compressed air source) hollow cathode lamps reagents and standards power argon gas hollow cathode lamps graphite tubes and cones reagents and standards power cooling water argon gas quartz torches sampling and skimmer cones reagents and standards pump tubing power cooling water 53
  • 54. REFERNCES  Shriver and Atkins’ Inorganic Chemistry, Fifth Edition  Concepts, Instrumentation and Techniques in Atomic Absorption Spectrophotometry by Richard D. Beaty and Jack D. Kerber  An elementary overview of elemental analysis by thermo elemental 54
  • 56. 56