Introduction to The Decontamination area
(Zone 1)
Day 1
Presented by Talal Albudayri
Learining Outcomes
o
o
o Describe why policies and procedures are required for an
effective decontamination unit.
Describe international standards available for decontamination units.
Describe the function of a decontamination unit and key
requirements for unit design.
Decontamination (Zone 1)
o
oCombination of process which are cleaning and disinfection to
make RMD safe for handling by staff and safe to use on patient.
The returned RMD to the decontamination area are
considered potentially infected and contaminated.
Personal Protective Equipment (PPE)
o
o
o
o
o
o
Cover gown with
sleeves. Face mask.
Eye protection.
Gloves.
Shoe covers.
Headcover.
Requirements tool for cleaning
o
o
o
o
o
o
o
o
o
Instrument cleaning sinks.
Brushes.
Air and water guns.
Detergents. (Enzymatic Detergents).
Cart washer.
Ultrasonic washer and automated washer
disinfectors. Disinfectants.
Environmental monitors.
Spill kit.
Decontamination (Zone 1)
All RMDs that come into contact with the patient or surgical field should be
systematically decontaminated after each surgical procedure.
o
o
o
o
The reusable medical devices reprocessed to a level appropriate for their intended use.
Depends on the body sites where the device will be used and the risk associated with a
particular procedure.
The minimum levels of processing based on three risk categories of use, are shown in a
classifications system called the Spaulding classification system.
In the reprocessing world, Spaulding Classification is the standard from which we learn what
to disinfect or sterilize and how much to do it.
The appropriate level of decontamination will depend on the
body sites where the device will be used and the risk
associated with the procedure.
The History of Spaulding Classification
o
o
o
o
Spaulding Classification goes way back to 1939 when a microbiologist named
Earle Spaulding wrote a paper for Temple University.
In it, he proposed a system for determining which patient care items needed
disinfection or sterilization. Spaulding separated patient care items into three
categories: non-critical, semi-critical, and critical.
He also recognized three levels of germicidal activity: high, intermediate, and
low. By 1957, this system was called the Spaulding Criteria and was widely being
used to further establish these norms.
The CDC currently defines Spaulding Classification as, “a strategy for
sterilization or disinfection of inanimate objects and surfaces based on
the degree of risk involved in their use.”
Critical
Ø A medical device that comes into contact with the vascular system
or sterile tissue and that shall be sterile at the time of use.
Clean as soon as possible after using.
Sterilize by moist heat after cleaning.
If RMD is heat or moisture sensitive, sterilize suing an alternative process,
e.g. automated low temperature chemical sterilizing process, liquid
chemical sterilizing process.
Process:
o
o
o
Semi-Critical
Ø A medical device that comes into contact with mucous
membranes or nonintact skin.
Clean as soon as possible after using.
Sterilize by moist heat after cleaning.
If RMD will not tolerate moist heat sterilization use a low temperature
sterilization process or disinfection using a high level instrument grade
.
Process:
chemical disinfectant
o
o
o
Non-Critical
Ø
Clean as necessary with detergent solution.
If further treatment is necessary,
A medical device that only comes into contact with intact skin
but not mucous membranes.
Process:
disinfect with compatible low level or
intermediate level instrument grade disinfectant after cleaning.
o
o
Surgical Instrument Processing Steps
Keeping items Moist
Dirty instruments returned to the
OR Why we should keep it moist ?!
Sorting and Disassembly of Contaminated RMD
Ø
The manufacturer’s instructions for cleaning.
1. Manual cleaning process .
2. Automated cleaning process, and there two types:
To ensure the device is not damaged and is cleaned adequately.
Reusable medical device decontamination must be carried out in a
timely manner.
Are followed:
DEPENDS ON:
Washer Disinfectors.
Ultrasonic Cleaners.
RMD Sorted according to :
o
o
Disassembly of RMD
o
o
o
o
o
Open all reusable medical device box locks.
Place the medical devices in a mesh basket in a way
that ensures effective cleaning of the device.
Do not place reusable medical devices on top of
one another.
Arrange medical devices in an orderly fashion in
mesh trays so that all surfaces are exposed to the
cleaning action when using an automated cleaner.
Overloaded baskets will result in ineffective cleaning.
Disassembly of RMD
o
o
o
If extra mesh baskets are required
for cleaning a device set,
in the extra basketsto identify the
set name and number.
Place heavy retractors and other
heavy medical devices on the
bottom or in a separate tray.
Place each jointed medical device in
the open position in the mesh basket.
a marker is placed
Disassembly of RMD
o Secure small and light items with
a hold down screen or by other
means, to ensure they are not
free to move around during the
cleaning process. Place scissors,
light-weight medical devices, and
microsurgical devices next.
Disassembly of RMD
o
o
Receivers and gallipots (bowl-shaped items) must not be placed
over any other reusable medical devices, as they may cause
shadowing and interfere with the cleaning process .
Separate all sharp devices from general devices. This is to ensure
ease of identification for personnel assembling the devices after
cleaning, in order to prevent sharps injury.
Disassembly of RMD
o
o
o
o
o
o
Open, flush and brush taps that do not come apart.
Stylets are used to unblock devices during use. Remove them, and
flush and brush the devices.
Open ratcheted devices for cleaning to ensure the box joints and
jaws are being cleaned thoroughly and all debris removed.
Dismantle any devices with valves, ports or multiple parts as far as
possible for cleaning, following manufacturer’s instructions.
For medical devices with one or more lumens (tubes), connect each
lumen to the appropriate flushing system provided for that purpose.
Ensure the tips of devices align and meet at the tip, and are not
hooked or snagged—items that are misaligned, damaged or worn
need to be sent for repair.
Camera Head Laying on top of Flex scope tip
As we have learned
o
o
o
In other words--if the item can't be cleaned, it can't be
disinfected or sterilized .
cleaning is an essential prerequisite for all effective disinfection and
sterilization processes, as organic residue may prevent the
disinfectant or sterilant from contacting the surfaces of the item being
processed and may also bind and inactivate chemical disinfectants.
cleaning is the initial and most crucial step in breaking the chain of
disease transmission.
Summery of the Decontamination (Zone 1)
Q: So, how do I know I am choosing the right method
of decontamination for the devices I’m dealing with?
o
o
o
o
Always remember—the appropriate level of decontamination will depend on the
body sites where the device will be used and the risk associated with the
procedure.
So for a device that enters the body, like surgical scissors, it is considered a
critical
risk and must be sterilized.
An endoscope, that comes in contact with intact skin or mucous membranes, is
considered a semi-critical risk, and a high level of disinfection is needed.
A blood pressure cuff that comes in contact with healthy skin, is considered non-
critical, and a good cleaning is needed.
Manual Cleaning
Day 2
Presented by Talal Albudayri
Manual Cleaning
Ø The use of manual cleaning methods for washing certain delicate or
complex devices, for example when:
Medical devices cannot be immersed in water ,power tools.
Medical devices have an operating temperature lower than is achieved in
automated washing.
Mechanical damage may result from the impact of the water jets or other items
in the load.
o
o
o
IN Manual Cleaning
o Carefully wash and rinse these devices according to the
manufacturer’s instructions. Non-immersion manual cleaning
methods are appropriate for certain devices like electrical tools
that may be compromised by soaking in wet solutions.
Immersion Method Essential Requirements
o
o
o
o
o
o
o
Always use brushes under water to avoid splashing and aerosols.
After cleaning
Place devices in
Sinkor a receptacle which will hold enough water and detergent to fully immerse the item
to be cleaned.
Detergentand a validated method of dispensing a measured quantity, following
manufacturers’ instructions for dilution and temperature of warm detergent solution.
Temperature controlin the wash and rinse sinks (usually <40C), thermostatic mixer
taps are preferred.
Brushesa selection of non-damaging brushes in a range of diameters and lengths for
cleaning both the external surfaces and The internal surfaces of devices. Single-use is
desirable,If not using single-use, there must be a protocol in place to decontaminate the
brushes at least daily.
, dry the manually washed devices.
a drying cabinet . W here a drying cabinet is not available use a clean,
disposable lint-free, absorbent wipe to dry the devices.
To ensure effective manual cleaning, follow written
procedures
o
o
o
o
Specifically designed to clean RMD.
Household soap must never be used.
A mild detergent is preferred for manual cleaning (pH range 8.0–11.5).
Follow the
Fill the sink with water to a predetermined level, at the specified temperature and
with the appropriate amount of detergent .
It is recommended to use three sinks-washing, potable or tap water rinse, and
purified water rinse.
Use detergents:
o
ĂŒ
ĂŒ
manufacturer’s instructions and local policies and procedures for
detergent dilution and water TEMP.
The gloves IN each stage IT must be changed
THE STEPS
o
o
o
o
o
o Inspect devices and equipment to establish that they are
clean before further processing or storage.
Immerse the item, carefully, in the solution in order to displace
trapped air—it is important to ensure that the cleaning solution
reaches all surfaces including those of lumened (tubed) devices.
Remove all visible soiling, including lumens and valves. Remove
stubborn staining by using inside non-abrasive scouring pad or
soaking in an approved stain-removing solution.
Dry mechanically in a drying cabinet or hand dry with a
clean, lint- free cloth and single use.
Note:items should not be left to dry in ambient air .
Flush all lumened devices with a jet-gun (discharge under
water to avoid splashing and aerosols).
Rinse the item finally in warm-to-hot water.
Ø
o
o
Take special care when you are
rinsing your instruments.
Here's a note of caution:
If not rinsed properly, low pH detergents may
breakdown the stainless protective surface and
cause black staining. High pH detergents may cause
surface deposits of brown stains, which can
interfere
with the smooth operation of the instrument.
Ø
o A mild detergent is preferred for manual
cleaning (pH range 8.0–11.5).
THE STEPS
o
o
Name of device or device set.
Name of processor.
Date.
Type of cleaning.
Type of detergent and detergent dilution used.
Keep a record of each device and device set that has been manually
cleaned. The records should contain, at a minimum:
§
§
§
§
§
THE STEPS
Automated Cleaning
(Including Pre-Cleaning)
Day 2
Presented by Talal Albudayri
Automated Cleaning (Including Pre-Cleaning)
o
o
o
o
o
As we have learned, cleaning is an essential prerequisite for all effective
disinfection and sterilization processes.
if the item can’t be cleaned, it can’t be disinfected or sterilized.
cleaning is the initial and most crucial step in breaking the chain of disease
transmission.
The advantage of using automated cleaning equipment is that it provides an
efficient, validated, reproducible process which is more easily controlled than
manual methods. always reproduces the same result.
Automated processes are generally more convenient and also provide
protection for the user in reducing exposure to contaminated devices and
chemicals.
o
o
Ideally, manual cleaning
should be used only where
no pre-treatment has been
carried out by the user, and
gross soiling remains on
the device.
All soiling must be removed
before automated cleaning
takes place.
Automated Cleaning (Including Pre-Cleaning)
Washer-Disinfectors
o All washer-disinfectors used for
decontamination of reusable medical
devices conform to the medicines and
healthcare products regulatory
authority (MHRA) requirements, are
compliant with EU legislation and
have
CE Marking (Conformité Européene),
and are approved by the US Food and
Drug Administration (FDA) or an
equivalent competent authority.
Questions to consider when Determining if the Device
is Compatiaable with the washer-disinfector included
o
o
o
o
o
What does the manufacturer’s IFU indicate?
Is it water submersible?
What is the maximum operating temperature?
Will mechanical damage occur from the impact of the water jets or other items in the load?
Is the device compatible with the process chemicals?
When ready to load the reusable medical devices there
are a few key points to note
4. Use the load carrier and racks for items they
were designed to hold.
1. Ensure the washer-disinfector and all
services are operational.
2. The washer-disinfector should not start if
any anomalies are present including blocked or
occluded spray arms.
3. Wearing PPE, load the rack or machine ensuring that
the arrangement of devices does not get in the way of the
cleaning process, and the rotary spray arms can rotate.
6.Load the load carrier into the washer-disinfector.
7.Secure the door (if fitted), and select and
start the appropriate cycle.
8.When the cycle is done, make sure that all
stages and parameters have been achieved.
9.When the automated cleaning process is complete,
inspect the processed devices.
5.Keep a record of each device and device set processed
in each washer-disinfector, and each cycle in order to
trace them through the decontamination process.
When ready to load the reusable medical devices there
are a few key points to note
Phases of a Typical Cycle
o
o
o
o
o
o
Drying –this phase purges the load and chamber with filtered heated air to remove
residual moisture.
Record information for each washer-disinfector cycle.
Cold flushing/pre-rinsing–chamber is filled with cold water (<40C), to prevent coagulation
of blood and protein, soil fixation-devices are rinsed to remove gross soil before the
chamber drains.
Warm washing –chamber fills again with warm fresh water plus detergent and is heated to
about 60C-this wash removes any remaining soil with the physical and chemical action of
detergent and again drains when completed.
Rinsing –chamber fills again, but this time with filtered or reverse osmosis (RO) water to
rinse and remove any chemical residues from the previous cleaning phase-in many washers
this phase is repeated a second time.
Disinfection –the chamber is filled for a final time again with filtered or RO water but this
time the temperature is raised to 80-93C and held for a predetermined time (usually 90C for
5 minutes) to ensure effective disinfection.
Documentation is required for every washer-disinfector
cycle and at a minimum should contain the following
o
o
o
o
o
o
o
o
o
o
o
Operator’s name.
Date and time of start of cycle.
Washer-disinfector identification number Type of washer-
disinfector.
Type of cycle used.
Cycle number.
Load contents, for example: general device set, stitch set, mayo
scissors.
Critical parameters for the specific washer-disinfector cycle.
Results of washer-disinfector process.
Signature of a qualified person (decontamination) confirming
whether or not the process cycle was within recommended
parameters.
Any notes or observation for the process cycle.
Document aborted cycles in a log book, recording the reasons and
actions taken.
Ultrasonic Cleaners
o
o
The Ultrasonic cleanser is used to
remove fine soil from surgical devices
after they have been manually cleaned
and before they are sterilized.
materials cannot be successfully
processed by this method:
Plastics and other similar materials.
Cemented glass syringes and lenses will
be damaged if repeatedly subjected to
this process.
Ø
Ø
General Guidelines
o
o
o
o
o
Water temperature is usually <35C
Thermal disinfection option-temperatures between 80C -93C with holding time >20
seconds to 10 minutes.
Document, scan and record all cycles and processes—preferably using automated
track and trace systems
Follow the manufacturer’s IFUs and standard operating procedures (SOP) Older
machines - de-gas time is between 5 and 15 minutes (de-gas is the removal of air from
the water to allow the ultrasonic waves to pass more freely).
Cycle time depends on the machine, cycle program and how much soil is on the Medical
device.
How Ultrasonic cleaners Works?!
o
o The high-frequency energy causes microscopic bubbles to form on the surface
of the devices and as the bubbles implode, minute vacuum areas are created,
drawing out the tiniest particles of debris from the crevices of the devices, This
process is called cavitation.
Converting high-frequency sound waves into mechanical vibrations that
free soil from the surface of devices.
How Ultrasonic cleaners Works?!
Thank you

Weitere Àhnliche Inhalte

Was ist angesagt?

Was ist angesagt? (7)

Pleomorphic adenoma of minor salivary gland revised
Pleomorphic adenoma of minor salivary gland revisedPleomorphic adenoma of minor salivary gland revised
Pleomorphic adenoma of minor salivary gland revised
 
Hacking with Remote Admin Tools (RAT)
 Hacking with Remote Admin Tools (RAT) Hacking with Remote Admin Tools (RAT)
Hacking with Remote Admin Tools (RAT)
 
Odontogenic tumours part 1
Odontogenic tumours part 1Odontogenic tumours part 1
Odontogenic tumours part 1
 
Seminar On Trojan Horse
Seminar On Trojan HorseSeminar On Trojan Horse
Seminar On Trojan Horse
 
Extravasation Mucocele– A Case Report
Extravasation Mucocele– A Case ReportExtravasation Mucocele– A Case Report
Extravasation Mucocele– A Case Report
 
Odontogenic tumours part 3
Odontogenic tumours part 3Odontogenic tumours part 3
Odontogenic tumours part 3
 
Oral candidiasis by hasan aziz ali
Oral candidiasis by hasan aziz aliOral candidiasis by hasan aziz ali
Oral candidiasis by hasan aziz ali
 

Ähnlich wie Introduction to The Decontamination area.pdf

Cleaning and disinfection of p atient care equipment
Cleaning and disinfection of p atient care equipmentCleaning and disinfection of p atient care equipment
Cleaning and disinfection of p atient care equipmentMEEQAT HOSPITAL
 
Reusable Medical Devices (SOP).docx
Reusable Medical Devices (SOP).docxReusable Medical Devices (SOP).docx
Reusable Medical Devices (SOP).docxanjalatchi
 
Cleaning and Disinfection of Medical Instruments
Cleaning and Disinfection of Medical InstrumentsCleaning and Disinfection of Medical Instruments
Cleaning and Disinfection of Medical InstrumentsAli Kermanjani, PhD
 
INSTRUCTIONS FOR REPROCESSING OF REUSABLE SURGICAL INSTRUMENTS
INSTRUCTIONS FOR REPROCESSING OF REUSABLE SURGICAL INSTRUMENTSINSTRUCTIONS FOR REPROCESSING OF REUSABLE SURGICAL INSTRUMENTS
INSTRUCTIONS FOR REPROCESSING OF REUSABLE SURGICAL INSTRUMENTSzackwilliam
 
DECONTAMINATION AND Sterilization OF INSTRUMENTS.pptx
DECONTAMINATION AND Sterilization OF INSTRUMENTS.pptxDECONTAMINATION AND Sterilization OF INSTRUMENTS.pptx
DECONTAMINATION AND Sterilization OF INSTRUMENTS.pptxAnthonyMatu1
 
ASA DENTAL ITALY Reprocessing
ASA DENTAL ITALY Reprocessing ASA DENTAL ITALY Reprocessing
ASA DENTAL ITALY Reprocessing Shimaa Hamdy
 
Safety precautions in the clinic and laboratory.pptx
Safety precautions in the clinic and laboratory.pptxSafety precautions in the clinic and laboratory.pptx
Safety precautions in the clinic and laboratory.pptxMustafa Al-Ali
 
HLD Training Course ppt 10-09-2015
HLD Training Course ppt 10-09-2015HLD Training Course ppt 10-09-2015
HLD Training Course ppt 10-09-2015Anibal (Luis) Acevedo
 
STERILISATION AND DISINFECTION IN DENTISTRY.pptx
STERILISATION   AND DISINFECTION   IN DENTISTRY.pptxSTERILISATION   AND DISINFECTION   IN DENTISTRY.pptx
STERILISATION AND DISINFECTION IN DENTISTRY.pptxDrRutikaNaik
 
Cross Infection Control in operative dentistry and endodontics .pptx
Cross Infection Control in operative dentistry and endodontics .pptxCross Infection Control in operative dentistry and endodontics .pptx
Cross Infection Control in operative dentistry and endodontics .pptxamiramna351
 
Behaviour_and_Practices_in_Cleanrooms.pptx
Behaviour_and_Practices_in_Cleanrooms.pptxBehaviour_and_Practices_in_Cleanrooms.pptx
Behaviour_and_Practices_in_Cleanrooms.pptxssuserba5705
 
BASICS TO FOOT CARE ppt FINAL2014.pdf
BASICS TO FOOT CARE ppt FINAL2014.pdfBASICS TO FOOT CARE ppt FINAL2014.pdf
BASICS TO FOOT CARE ppt FINAL2014.pdfMarcAndreo2
 
Dental infection control post treatment last
Dental infection control post treatment lastDental infection control post treatment last
Dental infection control post treatment lastAmin Abusallamah
 
Infection Control Guidelines for Dental Clinics [compatibility mode]
Infection Control Guidelines for Dental Clinics [compatibility mode]Infection Control Guidelines for Dental Clinics [compatibility mode]
Infection Control Guidelines for Dental Clinics [compatibility mode]drnahla
 
Infection Control Guidelines for Dental Clinics [compatibility mode]
Infection Control Guidelines for Dental Clinics [compatibility mode]Infection Control Guidelines for Dental Clinics [compatibility mode]
Infection Control Guidelines for Dental Clinics [compatibility mode]drnahla
 
theater steri final uma (2) (1).pptx
theater steri final uma (2) (1).pptxtheater steri final uma (2) (1).pptx
theater steri final uma (2) (1).pptxUmaKumar14
 
control hazards and risks.pptx
control hazards and risks.pptxcontrol hazards and risks.pptx
control hazards and risks.pptxnerissadollente1
 
Asepsis, sterilization and infection control
Asepsis, sterilization and infection controlAsepsis, sterilization and infection control
Asepsis, sterilization and infection controlDr. Meenal Atharkar
 

Ähnlich wie Introduction to The Decontamination area.pdf (20)

Cleaning and disinfection of p atient care equipment
Cleaning and disinfection of p atient care equipmentCleaning and disinfection of p atient care equipment
Cleaning and disinfection of p atient care equipment
 
OT
OTOT
OT
 
Reusable Medical Devices (SOP).docx
Reusable Medical Devices (SOP).docxReusable Medical Devices (SOP).docx
Reusable Medical Devices (SOP).docx
 
Cleaning and Disinfection of Medical Instruments
Cleaning and Disinfection of Medical InstrumentsCleaning and Disinfection of Medical Instruments
Cleaning and Disinfection of Medical Instruments
 
INSTRUCTIONS FOR REPROCESSING OF REUSABLE SURGICAL INSTRUMENTS
INSTRUCTIONS FOR REPROCESSING OF REUSABLE SURGICAL INSTRUMENTSINSTRUCTIONS FOR REPROCESSING OF REUSABLE SURGICAL INSTRUMENTS
INSTRUCTIONS FOR REPROCESSING OF REUSABLE SURGICAL INSTRUMENTS
 
DECONTAMINATION AND Sterilization OF INSTRUMENTS.pptx
DECONTAMINATION AND Sterilization OF INSTRUMENTS.pptxDECONTAMINATION AND Sterilization OF INSTRUMENTS.pptx
DECONTAMINATION AND Sterilization OF INSTRUMENTS.pptx
 
ASA DENTAL ITALY Reprocessing
ASA DENTAL ITALY Reprocessing ASA DENTAL ITALY Reprocessing
ASA DENTAL ITALY Reprocessing
 
Safety precautions in the clinic and laboratory.pptx
Safety precautions in the clinic and laboratory.pptxSafety precautions in the clinic and laboratory.pptx
Safety precautions in the clinic and laboratory.pptx
 
Cleaning Decontamination, Disinfection and Sterilization Process
Cleaning Decontamination, Disinfection and Sterilization ProcessCleaning Decontamination, Disinfection and Sterilization Process
Cleaning Decontamination, Disinfection and Sterilization Process
 
HLD Training Course ppt 10-09-2015
HLD Training Course ppt 10-09-2015HLD Training Course ppt 10-09-2015
HLD Training Course ppt 10-09-2015
 
STERILISATION AND DISINFECTION IN DENTISTRY.pptx
STERILISATION   AND DISINFECTION   IN DENTISTRY.pptxSTERILISATION   AND DISINFECTION   IN DENTISTRY.pptx
STERILISATION AND DISINFECTION IN DENTISTRY.pptx
 
Cross Infection Control in operative dentistry and endodontics .pptx
Cross Infection Control in operative dentistry and endodontics .pptxCross Infection Control in operative dentistry and endodontics .pptx
Cross Infection Control in operative dentistry and endodontics .pptx
 
Behaviour_and_Practices_in_Cleanrooms.pptx
Behaviour_and_Practices_in_Cleanrooms.pptxBehaviour_and_Practices_in_Cleanrooms.pptx
Behaviour_and_Practices_in_Cleanrooms.pptx
 
BASICS TO FOOT CARE ppt FINAL2014.pdf
BASICS TO FOOT CARE ppt FINAL2014.pdfBASICS TO FOOT CARE ppt FINAL2014.pdf
BASICS TO FOOT CARE ppt FINAL2014.pdf
 
Dental infection control post treatment last
Dental infection control post treatment lastDental infection control post treatment last
Dental infection control post treatment last
 
Infection Control Guidelines for Dental Clinics [compatibility mode]
Infection Control Guidelines for Dental Clinics [compatibility mode]Infection Control Guidelines for Dental Clinics [compatibility mode]
Infection Control Guidelines for Dental Clinics [compatibility mode]
 
Infection Control Guidelines for Dental Clinics [compatibility mode]
Infection Control Guidelines for Dental Clinics [compatibility mode]Infection Control Guidelines for Dental Clinics [compatibility mode]
Infection Control Guidelines for Dental Clinics [compatibility mode]
 
theater steri final uma (2) (1).pptx
theater steri final uma (2) (1).pptxtheater steri final uma (2) (1).pptx
theater steri final uma (2) (1).pptx
 
control hazards and risks.pptx
control hazards and risks.pptxcontrol hazards and risks.pptx
control hazards and risks.pptx
 
Asepsis, sterilization and infection control
Asepsis, sterilization and infection controlAsepsis, sterilization and infection control
Asepsis, sterilization and infection control
 

Mehr von Talal Albudayri

Introduction to operating room (Part three).pdf
Introduction to operating  room  (Part three).pdfIntroduction to operating  room  (Part three).pdf
Introduction to operating room (Part three).pdfTalal Albudayri
 
Introduction to operating room (Part Two).pdf
Introduction to operating  room (Part Two).pdfIntroduction to operating  room (Part Two).pdf
Introduction to operating room (Part Two).pdfTalal Albudayri
 
Introduction to operating room (Part one).pdf
Introduction to operating  room (Part one).pdfIntroduction to operating  room (Part one).pdf
Introduction to operating room (Part one).pdfTalal Albudayri
 
Outpatient Department (OPD).. overview, introduction and more info
Outpatient Department (OPD).. overview, introduction and more infoOutpatient Department (OPD).. overview, introduction and more info
Outpatient Department (OPD).. overview, introduction and more infoTalal Albudayri
 
Sterile Processing For Ambulatory Surgery and Other Practices...pdf
Sterile Processing For Ambulatory Surgery and Other Practices...pdfSterile Processing For Ambulatory Surgery and Other Practices...pdf
Sterile Processing For Ambulatory Surgery and Other Practices...pdfTalal Albudayri
 
Dental Surgical Instruments.. info and more
Dental Surgical Instruments.. info and moreDental Surgical Instruments.. info and more
Dental Surgical Instruments.. info and moreTalal Albudayri
 
CSSD EXAMS THAT YOU CAN TAKE TO HELP BOOST YOUR PROFESSIONAL CAREER
CSSD EXAMS THAT YOU CAN TAKE TO HELP BOOST YOUR PROFESSIONAL CAREERCSSD EXAMS THAT YOU CAN TAKE TO HELP BOOST YOUR PROFESSIONAL CAREER
CSSD EXAMS THAT YOU CAN TAKE TO HELP BOOST YOUR PROFESSIONAL CAREERTalal Albudayri
 
Quality Tools For CSSD Production .. A quality management system (QMS)
Quality Tools For CSSD Production .. A quality management system (QMS)Quality Tools For CSSD Production .. A quality management system (QMS)
Quality Tools For CSSD Production .. A quality management system (QMS)Talal Albudayri
 
Risk management .. WE WILL SOLVE THE PROBLEMS To understand all about Risk ma...
Risk management .. WE WILL SOLVE THE PROBLEMS To understand all about Risk ma...Risk management .. WE WILL SOLVE THE PROBLEMS To understand all about Risk ma...
Risk management .. WE WILL SOLVE THE PROBLEMS To understand all about Risk ma...Talal Albudayri
 
Surgical procedures & most used Sets.pdf
Surgical procedures & most used Sets.pdfSurgical procedures & most used Sets.pdf
Surgical procedures & most used Sets.pdfTalal Albudayri
 
CSSD EQUIPMENT .. AS AN INCORPORATED UNIT INSIDE CLINICS AND MEDICAL CARE OFF...
CSSD EQUIPMENT .. AS AN INCORPORATED UNIT INSIDE CLINICS AND MEDICAL CARE OFF...CSSD EQUIPMENT .. AS AN INCORPORATED UNIT INSIDE CLINICS AND MEDICAL CARE OFF...
CSSD EQUIPMENT .. AS AN INCORPORATED UNIT INSIDE CLINICS AND MEDICAL CARE OFF...Talal Albudayri
 
REGULATION AND STANDARDS .. Understand the difference between regulations, vo...
REGULATION AND STANDARDS .. Understand the difference between regulations, vo...REGULATION AND STANDARDS .. Understand the difference between regulations, vo...
REGULATION AND STANDARDS .. Understand the difference between regulations, vo...Talal Albudayri
 
Flexible Endoscopes.pdf Risks associated with reprocessing flexible endoscopes
Flexible Endoscopes.pdf Risks associated with reprocessing flexible endoscopesFlexible Endoscopes.pdf Risks associated with reprocessing flexible endoscopes
Flexible Endoscopes.pdf Risks associated with reprocessing flexible endoscopesTalal Albudayri
 
QUALITY ASSURANCE .. quality assurance and quality control.
QUALITY ASSURANCE .. quality assurance and quality control.QUALITY ASSURANCE .. quality assurance and quality control.
QUALITY ASSURANCE .. quality assurance and quality control.Talal Albudayri
 
Sterile storage and transport. Storage of sterile instrument under control en...
Sterile storage and transport. Storage of sterile instrument under control en...Sterile storage and transport. Storage of sterile instrument under control en...
Sterile storage and transport. Storage of sterile instrument under control en...Talal Albudayri
 
Sterilization.pdf Sterilization is the complete destruction of microorganisms...
Sterilization.pdf Sterilization is the complete destruction of microorganisms...Sterilization.pdf Sterilization is the complete destruction of microorganisms...
Sterilization.pdf Sterilization is the complete destruction of microorganisms...Talal Albudayri
 
Cleaning in sterile areas.pdf explain why personal protective equipment (PPE)...
Cleaning in sterile areas.pdf explain why personal protective equipment (PPE)...Cleaning in sterile areas.pdf explain why personal protective equipment (PPE)...
Cleaning in sterile areas.pdf explain why personal protective equipment (PPE)...Talal Albudayri
 
Surgical Instrumentation1.pdf DEFINE BASIC CATEGORIES OF INSTRUMENTS
Surgical Instrumentation1.pdf DEFINE BASIC CATEGORIES OF INSTRUMENTSSurgical Instrumentation1.pdf DEFINE BASIC CATEGORIES OF INSTRUMENTS
Surgical Instrumentation1.pdf DEFINE BASIC CATEGORIES OF INSTRUMENTSTalal Albudayri
 
IAP.pdf INSPECTION, ASSEMBLY AND PACKAGING (IAP)
IAP.pdf INSPECTION, ASSEMBLY AND PACKAGING (IAP)IAP.pdf INSPECTION, ASSEMBLY AND PACKAGING (IAP)
IAP.pdf INSPECTION, ASSEMBLY AND PACKAGING (IAP)Talal Albudayri
 
Microbes and PPE.pdf MICROBES ARE THE MOST NUMEROUS AND DIVERSE LIFE FORMS KN...
Microbes and PPE.pdf MICROBES ARE THE MOST NUMEROUS AND DIVERSE LIFE FORMS KN...Microbes and PPE.pdf MICROBES ARE THE MOST NUMEROUS AND DIVERSE LIFE FORMS KN...
Microbes and PPE.pdf MICROBES ARE THE MOST NUMEROUS AND DIVERSE LIFE FORMS KN...Talal Albudayri
 

Mehr von Talal Albudayri (20)

Introduction to operating room (Part three).pdf
Introduction to operating  room  (Part three).pdfIntroduction to operating  room  (Part three).pdf
Introduction to operating room (Part three).pdf
 
Introduction to operating room (Part Two).pdf
Introduction to operating  room (Part Two).pdfIntroduction to operating  room (Part Two).pdf
Introduction to operating room (Part Two).pdf
 
Introduction to operating room (Part one).pdf
Introduction to operating  room (Part one).pdfIntroduction to operating  room (Part one).pdf
Introduction to operating room (Part one).pdf
 
Outpatient Department (OPD).. overview, introduction and more info
Outpatient Department (OPD).. overview, introduction and more infoOutpatient Department (OPD).. overview, introduction and more info
Outpatient Department (OPD).. overview, introduction and more info
 
Sterile Processing For Ambulatory Surgery and Other Practices...pdf
Sterile Processing For Ambulatory Surgery and Other Practices...pdfSterile Processing For Ambulatory Surgery and Other Practices...pdf
Sterile Processing For Ambulatory Surgery and Other Practices...pdf
 
Dental Surgical Instruments.. info and more
Dental Surgical Instruments.. info and moreDental Surgical Instruments.. info and more
Dental Surgical Instruments.. info and more
 
CSSD EXAMS THAT YOU CAN TAKE TO HELP BOOST YOUR PROFESSIONAL CAREER
CSSD EXAMS THAT YOU CAN TAKE TO HELP BOOST YOUR PROFESSIONAL CAREERCSSD EXAMS THAT YOU CAN TAKE TO HELP BOOST YOUR PROFESSIONAL CAREER
CSSD EXAMS THAT YOU CAN TAKE TO HELP BOOST YOUR PROFESSIONAL CAREER
 
Quality Tools For CSSD Production .. A quality management system (QMS)
Quality Tools For CSSD Production .. A quality management system (QMS)Quality Tools For CSSD Production .. A quality management system (QMS)
Quality Tools For CSSD Production .. A quality management system (QMS)
 
Risk management .. WE WILL SOLVE THE PROBLEMS To understand all about Risk ma...
Risk management .. WE WILL SOLVE THE PROBLEMS To understand all about Risk ma...Risk management .. WE WILL SOLVE THE PROBLEMS To understand all about Risk ma...
Risk management .. WE WILL SOLVE THE PROBLEMS To understand all about Risk ma...
 
Surgical procedures & most used Sets.pdf
Surgical procedures & most used Sets.pdfSurgical procedures & most used Sets.pdf
Surgical procedures & most used Sets.pdf
 
CSSD EQUIPMENT .. AS AN INCORPORATED UNIT INSIDE CLINICS AND MEDICAL CARE OFF...
CSSD EQUIPMENT .. AS AN INCORPORATED UNIT INSIDE CLINICS AND MEDICAL CARE OFF...CSSD EQUIPMENT .. AS AN INCORPORATED UNIT INSIDE CLINICS AND MEDICAL CARE OFF...
CSSD EQUIPMENT .. AS AN INCORPORATED UNIT INSIDE CLINICS AND MEDICAL CARE OFF...
 
REGULATION AND STANDARDS .. Understand the difference between regulations, vo...
REGULATION AND STANDARDS .. Understand the difference between regulations, vo...REGULATION AND STANDARDS .. Understand the difference between regulations, vo...
REGULATION AND STANDARDS .. Understand the difference between regulations, vo...
 
Flexible Endoscopes.pdf Risks associated with reprocessing flexible endoscopes
Flexible Endoscopes.pdf Risks associated with reprocessing flexible endoscopesFlexible Endoscopes.pdf Risks associated with reprocessing flexible endoscopes
Flexible Endoscopes.pdf Risks associated with reprocessing flexible endoscopes
 
QUALITY ASSURANCE .. quality assurance and quality control.
QUALITY ASSURANCE .. quality assurance and quality control.QUALITY ASSURANCE .. quality assurance and quality control.
QUALITY ASSURANCE .. quality assurance and quality control.
 
Sterile storage and transport. Storage of sterile instrument under control en...
Sterile storage and transport. Storage of sterile instrument under control en...Sterile storage and transport. Storage of sterile instrument under control en...
Sterile storage and transport. Storage of sterile instrument under control en...
 
Sterilization.pdf Sterilization is the complete destruction of microorganisms...
Sterilization.pdf Sterilization is the complete destruction of microorganisms...Sterilization.pdf Sterilization is the complete destruction of microorganisms...
Sterilization.pdf Sterilization is the complete destruction of microorganisms...
 
Cleaning in sterile areas.pdf explain why personal protective equipment (PPE)...
Cleaning in sterile areas.pdf explain why personal protective equipment (PPE)...Cleaning in sterile areas.pdf explain why personal protective equipment (PPE)...
Cleaning in sterile areas.pdf explain why personal protective equipment (PPE)...
 
Surgical Instrumentation1.pdf DEFINE BASIC CATEGORIES OF INSTRUMENTS
Surgical Instrumentation1.pdf DEFINE BASIC CATEGORIES OF INSTRUMENTSSurgical Instrumentation1.pdf DEFINE BASIC CATEGORIES OF INSTRUMENTS
Surgical Instrumentation1.pdf DEFINE BASIC CATEGORIES OF INSTRUMENTS
 
IAP.pdf INSPECTION, ASSEMBLY AND PACKAGING (IAP)
IAP.pdf INSPECTION, ASSEMBLY AND PACKAGING (IAP)IAP.pdf INSPECTION, ASSEMBLY AND PACKAGING (IAP)
IAP.pdf INSPECTION, ASSEMBLY AND PACKAGING (IAP)
 
Microbes and PPE.pdf MICROBES ARE THE MOST NUMEROUS AND DIVERSE LIFE FORMS KN...
Microbes and PPE.pdf MICROBES ARE THE MOST NUMEROUS AND DIVERSE LIFE FORMS KN...Microbes and PPE.pdf MICROBES ARE THE MOST NUMEROUS AND DIVERSE LIFE FORMS KN...
Microbes and PPE.pdf MICROBES ARE THE MOST NUMEROUS AND DIVERSE LIFE FORMS KN...
 

KĂŒrzlich hochgeladen

Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Timedelhimodelshub1
 
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...delhimodelshub1
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknowgragteena
 
Dehradun Call Girls Service â€ïžđŸ‘ 9675010100 đŸ‘„đŸ«ŠIndependent Escort Service Dehradun
Dehradun Call Girls Service â€ïžđŸ‘ 9675010100 đŸ‘„đŸ«ŠIndependent Escort Service DehradunDehradun Call Girls Service â€ïžđŸ‘ 9675010100 đŸ‘„đŸ«ŠIndependent Escort Service Dehradun
Dehradun Call Girls Service â€ïžđŸ‘ 9675010100 đŸ‘„đŸ«ŠIndependent Escort Service DehradunNiamh verma
 
Call Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any TimeCall Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any Timedelhimodelshub1
 
Call Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any TimeCall Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any Timedelhimodelshub1
 
Call Girls in Mohali Surbhi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort Service Mohali
Call Girls in Mohali Surbhi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort Service MohaliCall Girls in Mohali Surbhi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort Service Mohali
Call Girls in Mohali Surbhi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort Service MohaliHigh Profile Call Girls Chandigarh Aarushi
 
Russian Call Girls in Chandigarh Ojaswi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort ...
Russian Call Girls in Chandigarh Ojaswi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort ...Russian Call Girls in Chandigarh Ojaswi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort ...
Russian Call Girls in Chandigarh Ojaswi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort ...High Profile Call Girls Chandigarh Aarushi
 
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...Call Girls Noida
 
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012Call Girls Service Gurgaon
 
Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949ps5894268
 
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service MumbaiCollege Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Local Housewife and effective ☎ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
Local Housewife and effective ☎ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...Local Housewife and effective ☎ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
Local Housewife and effective ☎ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...Russian Call Girls Amritsar
 
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service HyderabadCall Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabaddelhimodelshub1
 

KĂŒrzlich hochgeladen (20)

Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service DehradunCall Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
Call Girl Dehradun Aashi 🔝 7001305949 🔝 💃 Independent Escort Service Dehradun
 
Call Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any TimeCall Girls LB Nagar 7001305949 all area service COD available Any Time
Call Girls LB Nagar 7001305949 all area service COD available Any Time
 
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 đŸŽ¶ Independent Escort Service Lucknow
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 đŸŽ¶ Independent Escort Service LucknowVIP Call Girls Lucknow Isha 🔝 9719455033 🔝 đŸŽ¶ Independent Escort Service Lucknow
VIP Call Girls Lucknow Isha 🔝 9719455033 🔝 đŸŽ¶ Independent Escort Service Lucknow
 
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
Russian Call Girls Hyderabad Indira 9907093804 Independent Escort Service Hyd...
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
 
Dehradun Call Girls Service â€ïžđŸ‘ 9675010100 đŸ‘„đŸ«ŠIndependent Escort Service Dehradun
Dehradun Call Girls Service â€ïžđŸ‘ 9675010100 đŸ‘„đŸ«ŠIndependent Escort Service DehradunDehradun Call Girls Service â€ïžđŸ‘ 9675010100 đŸ‘„đŸ«ŠIndependent Escort Service Dehradun
Dehradun Call Girls Service â€ïžđŸ‘ 9675010100 đŸ‘„đŸ«ŠIndependent Escort Service Dehradun
 
Call Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any TimeCall Girls Kukatpally 7001305949 all area service COD available Any Time
Call Girls Kukatpally 7001305949 all area service COD available Any Time
 
Russian Call Girls in Dehradun Komal 🔝 7001305949 🔝 📍 Independent Escort Serv...
Russian Call Girls in Dehradun Komal 🔝 7001305949 🔝 📍 Independent Escort Serv...Russian Call Girls in Dehradun Komal 🔝 7001305949 🔝 📍 Independent Escort Serv...
Russian Call Girls in Dehradun Komal 🔝 7001305949 🔝 📍 Independent Escort Serv...
 
Call Girl Lucknow Gauri 🔝 8923113531 🔝 đŸŽ¶ Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531  🔝 đŸŽ¶ Independent Escort Service LucknowCall Girl Lucknow Gauri 🔝 8923113531  🔝 đŸŽ¶ Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531 🔝 đŸŽ¶ Independent Escort Service Lucknow
 
Call Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any TimeCall Girls Secunderabad 7001305949 all area service COD available Any Time
Call Girls Secunderabad 7001305949 all area service COD available Any Time
 
Call Girls in Mohali Surbhi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort Service Mohali
Call Girls in Mohali Surbhi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort Service MohaliCall Girls in Mohali Surbhi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort Service Mohali
Call Girls in Mohali Surbhi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort Service Mohali
 
Russian Call Girls in Chandigarh Ojaswi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort ...
Russian Call Girls in Chandigarh Ojaswi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort ...Russian Call Girls in Chandigarh Ojaswi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort ...
Russian Call Girls in Chandigarh Ojaswi â€ïžđŸ‘ 9907093804 đŸ‘„đŸ«Š Independent Escort ...
 
Russian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your bookingRussian Call Girls South Delhi 9711199171 discount on your booking
Russian Call Girls South Delhi 9711199171 discount on your booking
 
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
pOOJA sexy Call Girls In Sector 49,9999965857 Young Female Escorts Service In...
 
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
VIP Call Girls Sector 67 Gurgaon Just Call Me 9711199012
 
Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949Low Rate Call Girls In Bommanahalli Just Call 7001305949
Low Rate Call Girls In Bommanahalli Just Call 7001305949
 
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service GuwahatiCall Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
 
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service MumbaiCollege Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
College Call Girls Mumbai Alia 9910780858 Independent Escort Service Mumbai
 
Local Housewife and effective ☎ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
Local Housewife and effective ☎ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...Local Housewife and effective ☎ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
Local Housewife and effective ☎ 8250192130 🍉🍓 Sexy Girls VIP Call Girls Chan...
 
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service HyderabadCall Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
Call Girls Hyderabad Kirti 9907093804 Independent Escort Service Hyderabad
 

Introduction to The Decontamination area.pdf

  • 1. ALL YOU NEED TO KNOW ABOUT CENTRAL STERILE SUPPLY DEPARTMENT
  • 2. Introduction to The Decontamination area (Zone 1) Day 1 Presented by Talal Albudayri
  • 3. Learining Outcomes o o o Describe why policies and procedures are required for an effective decontamination unit. Describe international standards available for decontamination units. Describe the function of a decontamination unit and key requirements for unit design.
  • 4. Decontamination (Zone 1) o oCombination of process which are cleaning and disinfection to make RMD safe for handling by staff and safe to use on patient. The returned RMD to the decontamination area are considered potentially infected and contaminated.
  • 5. Personal Protective Equipment (PPE) o o o o o o Cover gown with sleeves. Face mask. Eye protection. Gloves. Shoe covers. Headcover.
  • 6. Requirements tool for cleaning o o o o o o o o o Instrument cleaning sinks. Brushes. Air and water guns. Detergents. (Enzymatic Detergents). Cart washer. Ultrasonic washer and automated washer disinfectors. Disinfectants. Environmental monitors. Spill kit.
  • 7. Decontamination (Zone 1) All RMDs that come into contact with the patient or surgical field should be systematically decontaminated after each surgical procedure. o o o o The reusable medical devices reprocessed to a level appropriate for their intended use. Depends on the body sites where the device will be used and the risk associated with a particular procedure. The minimum levels of processing based on three risk categories of use, are shown in a classifications system called the Spaulding classification system. In the reprocessing world, Spaulding Classification is the standard from which we learn what to disinfect or sterilize and how much to do it.
  • 8. The appropriate level of decontamination will depend on the body sites where the device will be used and the risk associated with the procedure.
  • 9. The History of Spaulding Classification o o o o Spaulding Classification goes way back to 1939 when a microbiologist named Earle Spaulding wrote a paper for Temple University. In it, he proposed a system for determining which patient care items needed disinfection or sterilization. Spaulding separated patient care items into three categories: non-critical, semi-critical, and critical. He also recognized three levels of germicidal activity: high, intermediate, and low. By 1957, this system was called the Spaulding Criteria and was widely being used to further establish these norms. The CDC currently defines Spaulding Classification as, “a strategy for sterilization or disinfection of inanimate objects and surfaces based on the degree of risk involved in their use.”
  • 10. Critical Ø A medical device that comes into contact with the vascular system or sterile tissue and that shall be sterile at the time of use. Clean as soon as possible after using. Sterilize by moist heat after cleaning. If RMD is heat or moisture sensitive, sterilize suing an alternative process, e.g. automated low temperature chemical sterilizing process, liquid chemical sterilizing process. Process: o o o
  • 11. Semi-Critical Ø A medical device that comes into contact with mucous membranes or nonintact skin. Clean as soon as possible after using. Sterilize by moist heat after cleaning. If RMD will not tolerate moist heat sterilization use a low temperature sterilization process or disinfection using a high level instrument grade . Process: chemical disinfectant o o o
  • 12. Non-Critical Ø Clean as necessary with detergent solution. If further treatment is necessary, A medical device that only comes into contact with intact skin but not mucous membranes. Process: disinfect with compatible low level or intermediate level instrument grade disinfectant after cleaning. o o
  • 15. Dirty instruments returned to the OR Why we should keep it moist ?!
  • 16. Sorting and Disassembly of Contaminated RMD Ø The manufacturer’s instructions for cleaning. 1. Manual cleaning process . 2. Automated cleaning process, and there two types: To ensure the device is not damaged and is cleaned adequately. Reusable medical device decontamination must be carried out in a timely manner. Are followed: DEPENDS ON: Washer Disinfectors. Ultrasonic Cleaners. RMD Sorted according to : o o
  • 17. Disassembly of RMD o o o o o Open all reusable medical device box locks. Place the medical devices in a mesh basket in a way that ensures effective cleaning of the device. Do not place reusable medical devices on top of one another. Arrange medical devices in an orderly fashion in mesh trays so that all surfaces are exposed to the cleaning action when using an automated cleaner. Overloaded baskets will result in ineffective cleaning.
  • 18. Disassembly of RMD o o o If extra mesh baskets are required for cleaning a device set, in the extra basketsto identify the set name and number. Place heavy retractors and other heavy medical devices on the bottom or in a separate tray. Place each jointed medical device in the open position in the mesh basket. a marker is placed
  • 19. Disassembly of RMD o Secure small and light items with a hold down screen or by other means, to ensure they are not free to move around during the cleaning process. Place scissors, light-weight medical devices, and microsurgical devices next.
  • 20. Disassembly of RMD o o Receivers and gallipots (bowl-shaped items) must not be placed over any other reusable medical devices, as they may cause shadowing and interfere with the cleaning process . Separate all sharp devices from general devices. This is to ensure ease of identification for personnel assembling the devices after cleaning, in order to prevent sharps injury.
  • 21. Disassembly of RMD o o o o o o Open, flush and brush taps that do not come apart. Stylets are used to unblock devices during use. Remove them, and flush and brush the devices. Open ratcheted devices for cleaning to ensure the box joints and jaws are being cleaned thoroughly and all debris removed. Dismantle any devices with valves, ports or multiple parts as far as possible for cleaning, following manufacturer’s instructions. For medical devices with one or more lumens (tubes), connect each lumen to the appropriate flushing system provided for that purpose. Ensure the tips of devices align and meet at the tip, and are not hooked or snagged—items that are misaligned, damaged or worn need to be sent for repair.
  • 22. Camera Head Laying on top of Flex scope tip
  • 23. As we have learned o o o In other words--if the item can't be cleaned, it can't be disinfected or sterilized . cleaning is an essential prerequisite for all effective disinfection and sterilization processes, as organic residue may prevent the disinfectant or sterilant from contacting the surfaces of the item being processed and may also bind and inactivate chemical disinfectants. cleaning is the initial and most crucial step in breaking the chain of disease transmission.
  • 24. Summery of the Decontamination (Zone 1) Q: So, how do I know I am choosing the right method of decontamination for the devices I’m dealing with? o o o o Always remember—the appropriate level of decontamination will depend on the body sites where the device will be used and the risk associated with the procedure. So for a device that enters the body, like surgical scissors, it is considered a critical risk and must be sterilized. An endoscope, that comes in contact with intact skin or mucous membranes, is considered a semi-critical risk, and a high level of disinfection is needed. A blood pressure cuff that comes in contact with healthy skin, is considered non- critical, and a good cleaning is needed.
  • 25. Manual Cleaning Day 2 Presented by Talal Albudayri
  • 26. Manual Cleaning Ø The use of manual cleaning methods for washing certain delicate or complex devices, for example when: Medical devices cannot be immersed in water ,power tools. Medical devices have an operating temperature lower than is achieved in automated washing. Mechanical damage may result from the impact of the water jets or other items in the load. o o o
  • 27. IN Manual Cleaning o Carefully wash and rinse these devices according to the manufacturer’s instructions. Non-immersion manual cleaning methods are appropriate for certain devices like electrical tools that may be compromised by soaking in wet solutions.
  • 28. Immersion Method Essential Requirements o o o o o o o Always use brushes under water to avoid splashing and aerosols. After cleaning Place devices in Sinkor a receptacle which will hold enough water and detergent to fully immerse the item to be cleaned. Detergentand a validated method of dispensing a measured quantity, following manufacturers’ instructions for dilution and temperature of warm detergent solution. Temperature controlin the wash and rinse sinks (usually <40C), thermostatic mixer taps are preferred. Brushesa selection of non-damaging brushes in a range of diameters and lengths for cleaning both the external surfaces and The internal surfaces of devices. Single-use is desirable,If not using single-use, there must be a protocol in place to decontaminate the brushes at least daily. , dry the manually washed devices. a drying cabinet . W here a drying cabinet is not available use a clean, disposable lint-free, absorbent wipe to dry the devices.
  • 29. To ensure effective manual cleaning, follow written procedures o o o o Specifically designed to clean RMD. Household soap must never be used. A mild detergent is preferred for manual cleaning (pH range 8.0–11.5). Follow the Fill the sink with water to a predetermined level, at the specified temperature and with the appropriate amount of detergent . It is recommended to use three sinks-washing, potable or tap water rinse, and purified water rinse. Use detergents: o ĂŒ ĂŒ manufacturer’s instructions and local policies and procedures for detergent dilution and water TEMP.
  • 30. The gloves IN each stage IT must be changed
  • 31. THE STEPS o o o o o o Inspect devices and equipment to establish that they are clean before further processing or storage. Immerse the item, carefully, in the solution in order to displace trapped air—it is important to ensure that the cleaning solution reaches all surfaces including those of lumened (tubed) devices. Remove all visible soiling, including lumens and valves. Remove stubborn staining by using inside non-abrasive scouring pad or soaking in an approved stain-removing solution. Dry mechanically in a drying cabinet or hand dry with a clean, lint- free cloth and single use. Note:items should not be left to dry in ambient air . Flush all lumened devices with a jet-gun (discharge under water to avoid splashing and aerosols). Rinse the item finally in warm-to-hot water. Ø
  • 32. o o Take special care when you are rinsing your instruments. Here's a note of caution: If not rinsed properly, low pH detergents may breakdown the stainless protective surface and cause black staining. High pH detergents may cause surface deposits of brown stains, which can interfere with the smooth operation of the instrument. Ø o A mild detergent is preferred for manual cleaning (pH range 8.0–11.5). THE STEPS
  • 33. o o Name of device or device set. Name of processor. Date. Type of cleaning. Type of detergent and detergent dilution used. Keep a record of each device and device set that has been manually cleaned. The records should contain, at a minimum: § § § § § THE STEPS
  • 34. Automated Cleaning (Including Pre-Cleaning) Day 2 Presented by Talal Albudayri
  • 35. Automated Cleaning (Including Pre-Cleaning) o o o o o As we have learned, cleaning is an essential prerequisite for all effective disinfection and sterilization processes. if the item can’t be cleaned, it can’t be disinfected or sterilized. cleaning is the initial and most crucial step in breaking the chain of disease transmission. The advantage of using automated cleaning equipment is that it provides an efficient, validated, reproducible process which is more easily controlled than manual methods. always reproduces the same result. Automated processes are generally more convenient and also provide protection for the user in reducing exposure to contaminated devices and chemicals.
  • 36. o o Ideally, manual cleaning should be used only where no pre-treatment has been carried out by the user, and gross soiling remains on the device. All soiling must be removed before automated cleaning takes place. Automated Cleaning (Including Pre-Cleaning)
  • 37. Washer-Disinfectors o All washer-disinfectors used for decontamination of reusable medical devices conform to the medicines and healthcare products regulatory authority (MHRA) requirements, are compliant with EU legislation and have CE Marking (ConformitĂ© EuropĂ©ene), and are approved by the US Food and Drug Administration (FDA) or an equivalent competent authority.
  • 38. Questions to consider when Determining if the Device is Compatiaable with the washer-disinfector included o o o o o What does the manufacturer’s IFU indicate? Is it water submersible? What is the maximum operating temperature? Will mechanical damage occur from the impact of the water jets or other items in the load? Is the device compatible with the process chemicals?
  • 39. When ready to load the reusable medical devices there are a few key points to note 4. Use the load carrier and racks for items they were designed to hold. 1. Ensure the washer-disinfector and all services are operational. 2. The washer-disinfector should not start if any anomalies are present including blocked or occluded spray arms. 3. Wearing PPE, load the rack or machine ensuring that the arrangement of devices does not get in the way of the cleaning process, and the rotary spray arms can rotate.
  • 40. 6.Load the load carrier into the washer-disinfector. 7.Secure the door (if fitted), and select and start the appropriate cycle. 8.When the cycle is done, make sure that all stages and parameters have been achieved. 9.When the automated cleaning process is complete, inspect the processed devices. 5.Keep a record of each device and device set processed in each washer-disinfector, and each cycle in order to trace them through the decontamination process. When ready to load the reusable medical devices there are a few key points to note
  • 41. Phases of a Typical Cycle o o o o o o Drying –this phase purges the load and chamber with filtered heated air to remove residual moisture. Record information for each washer-disinfector cycle. Cold flushing/pre-rinsing–chamber is filled with cold water (<40C), to prevent coagulation of blood and protein, soil fixation-devices are rinsed to remove gross soil before the chamber drains. Warm washing –chamber fills again with warm fresh water plus detergent and is heated to about 60C-this wash removes any remaining soil with the physical and chemical action of detergent and again drains when completed. Rinsing –chamber fills again, but this time with filtered or reverse osmosis (RO) water to rinse and remove any chemical residues from the previous cleaning phase-in many washers this phase is repeated a second time. Disinfection –the chamber is filled for a final time again with filtered or RO water but this time the temperature is raised to 80-93C and held for a predetermined time (usually 90C for 5 minutes) to ensure effective disinfection.
  • 42. Documentation is required for every washer-disinfector cycle and at a minimum should contain the following o o o o o o o o o o o Operator’s name. Date and time of start of cycle. Washer-disinfector identification number Type of washer- disinfector. Type of cycle used. Cycle number. Load contents, for example: general device set, stitch set, mayo scissors. Critical parameters for the specific washer-disinfector cycle. Results of washer-disinfector process. Signature of a qualified person (decontamination) confirming whether or not the process cycle was within recommended parameters. Any notes or observation for the process cycle. Document aborted cycles in a log book, recording the reasons and actions taken.
  • 43. Ultrasonic Cleaners o o The Ultrasonic cleanser is used to remove fine soil from surgical devices after they have been manually cleaned and before they are sterilized. materials cannot be successfully processed by this method: Plastics and other similar materials. Cemented glass syringes and lenses will be damaged if repeatedly subjected to this process. Ø Ø
  • 44. General Guidelines o o o o o Water temperature is usually <35C Thermal disinfection option-temperatures between 80C -93C with holding time >20 seconds to 10 minutes. Document, scan and record all cycles and processes—preferably using automated track and trace systems Follow the manufacturer’s IFUs and standard operating procedures (SOP) Older machines - de-gas time is between 5 and 15 minutes (de-gas is the removal of air from the water to allow the ultrasonic waves to pass more freely). Cycle time depends on the machine, cycle program and how much soil is on the Medical device.
  • 45. How Ultrasonic cleaners Works?! o o The high-frequency energy causes microscopic bubbles to form on the surface of the devices and as the bubbles implode, minute vacuum areas are created, drawing out the tiniest particles of debris from the crevices of the devices, This process is called cavitation. Converting high-frequency sound waves into mechanical vibrations that free soil from the surface of devices.