SlideShare ist ein Scribd-Unternehmen logo
1 von 91
Prepared by:
Karwan O. Ali
Yousif T. Maaroof
Mzgin. M. Ayoob
UV- VISIBLE SPECTROSCOPY
What is spectroscopy?
The interactions of radiation and matter are the subject of the
science called spectroscopy. Spectroscopic analytical methods
are based on measuring the amount of radiation produced or
absorbed by molecular or atomic species of interest
The Electromagnetic Spectrum
Light exhibits wave property during its propagation and energy particle
during its interaction with matter. The double nature of light (waves and
particles) is known as dualism.
Dual nature of light
• Light consist of energy packets, known as photons.
• The energy (E) of photons is proportional to the frequency i.e.
related to c and . It can be expressed by max plank relation:
• E = h  ( = C /)
• where h = max plank constant = 6.63 x 10-27 erg., sec.)
• i.e. E   or E  1/   
For analytical purposes we use the region of I.R, visible and U.V
radiations.
UV radiation region is classified into : far UV from (10nm-200nm) and
near UV from (200nm-380nm
Why Is a Red Solution Red?
• When a molecule interact with radiant energy, the molecule is said to
be excited , because the outer valence electrons undergo transition
from original energy level ground state (E g) to an exited state (E s).
Interaction of a substance with EMR
Excited state Es
Ground state E g
When a molecule in the ground state absorbs
EMR, 3 energy state transitions will take place.
•These types of transition are:
1) Electronic
2) Vibrational
3) Rotational.
• When the molecule absorb Visible and U.V region.
Raising electrons to a higher energy level, raising the Vibration of
molecule, and increasing rotation of the molecule (Electronic transition
energy + Vibrational transition energy + Rotational transition energy)
1) Electronic
• When the molecule absorb I.R region.
Raising the Vibration of molecule and increasing rotation of the
molecule (Vibrational transition energy + Rotational transition energy)
2) Vibrational
• When the molecule absorbs F.I.R and Microwave regions.
Increasing rotation of the molecule (Rotational transition
energy).
3) Rotational
• Absorption measurements based upon ultraviolet and visible radiation find widespread
application for the quantitative determination of a large variety species [1].
• Beer’s Law:
• A = -logT = logP0/P = abc = bc = 2 - log%T
• Where T = transmittance, T = percentage transmittance, P = transmitted power of
radiation, Po = incident power of radiation, A = absorbance, a = absorptivity,
b = path length, c = concentration,  = molar absorptivity, extinction coefficient
An Introduction to Ultraviolet/Visible
Molecular Absorption Spectrometry
The Quantitative Picture
Where the absorbance A has no units, since A = log10 P0 / P
 Is the molar absorbtivity with units of L mol-1 cm-1
b is the path length of the sample in cm
c is the concentration of the compound in solution, expressed in mol L-1 (or M,
molarity)
The Beer-Lambert Law (Beer’s Law): A =  b c
• Monochromatic incident radiation (all molecules absorb light of one )
• Absorbers independent (Absorbing molecules act independently of one another i.e., low c)
• Path length is uniform (all rays travel the same distance in sample)
• No scattering
• Absorbing medium is optically homogeneous
• Incident beam is not large enough to cause saturation
• All rays should be parallel to each other and perpendicular to surface of medium
Assumptions in derivation of Beer’s Law
Deviations from Beer’s Law
• Real Limitations
• Beer’s law is successful in describing the absorption behavior of dilute solutions only ; in this sense it is a
limiting law. At high concentrations ( > 0.01M ),the average distance between the species responsible for
absorption is diminished to the point where each affects the charge distribution of its neighbors
This interaction, in turn, can alter the species’ ability to absorb at a given wavelength of radiation thus leading
to a deviation from Beer’s law [2].
A similar effect is sometimes encountered in solutions containing low absorber concentrations and high
concentrations of other species, particularly electrolytes.
Beer’s law is valid at low concentrations,
but breaks down at higher concentrations
For linearity, A < 1
Deviations from Beer's law also arise because ε is dependent upon the
refractive index of the solution. Thus, if concentration changes cause
significant alterations in the refractive index η of a solution, departures from
Beer's law are observed. A correction for this effect can be made by
substitution of the quantity εη /(η2 + 2)2 for ε in the Beer equation, as shown
below [2]:
A= [εηb /(η2 + 2)2 ] bC
In general, this correction is never very large and is rarely significant at
concentrations less than 0.01 M.
• Chemical deviations from Beer’s law are caused by shifts in the position of a chemical or physical
equilibrium involving the absorbing species. A common example of this behavior is found with
acid/base indicators. Deviations arising from chemical factors can only be observed when
concentrations are changed
Chemical Deviations
HIn
Ka
H+ + In-
Red,  =600nm Colorless
Phenolphthalein:
If solution is buffered, then pH is constant and [HIn]
is related to absorbance.
But, if un buffered solution,
equilibrium will shift depending on
total analyte concentration
C
HIn
Expected
Actual
• instrument may be caused by fluctuations in the power-supply voltage, an
unstable light source, or a non-linear response of the detector-amplifier system
• Polychromatic Radiation
• All monochromators, regardless of quality and size, have a finite resolving
power and therefore minimum instrumental bandwidth. Polychromatic
radiation (i.e., light of more than one )
Instrumental Factors
Band A shows little deviation, because ε does not change greatly throughout the band. Band
B shows marked deviations because ε undergoes significant changes in this region
The width of the image produced is thus an important measure of the quality of the
performance of a spectrometer. The figure below shows the loss of detail that accompanies the
use of wider slits. It is evident that an increase in slit width brings about a loss of spectral
detail
Effect of Slit Width on Absorbance Measurements
• quantitative measurement of narrow absorption bands demand the use of narrow slits widths. Unfortunately, a
decrease in slit width is accompanied by a second-order power reduction in the radiant energy; at very narrow settings
spectral detail may be lost owing to an increase in the signal-to-noise ratio. In general, it is good practice to narrow
slits no more than is necessary for good resolution for the spectrum at hand
Another effect of slit width is the change of absorbance values that accompany a change in the slit width. The
figure below illustrates this effect. Note that the peak absorbance values increase significantly (by as much as
70% in one instance) as the slit width decreases
The basic components of analytical instruments for absorption spectroscopy,
as well as for emission and fluorescence spectroscopy, are remarkably alike
in function and in general performance requirements whether the
instruments are designed for ultraviolet (UV), visible, or infrared (lR)
radiation. We often call the UV/visible and IR regions of the spectrum the
optical region
Instruments for Optical Spectrometry
1) a stable source of radiant energy
2) a wavelength selector that isolates a limited region of the spectrum for
measurement
3) one or more sample containers
4) a radiation detector, which converts radiant energy to a measurable electrical
signal
5) a signal processing and readout unit, usually consisting of electronic hardware
and, in modern instruments, a computer.
INSTRUMENT COMPONENTS
To be suitable for spectroscopic studies, a source must generate a beam of radiation that is sufficiently powerful
to allow easy detection and measurement.
In addition, its output power should be stable for reasonable periods of time. Typically, for good stability, a well-
regulated power supply must provide electrical power for the source
Spectroscopic Sources
Spectroscopic sources are of two types:
A- Continuum sources, which emit radiation that changes in intensity only
slowly as a function of wavelength.
B- Line sources, which emit a limited number of spectral lines, each of
which spans a very limited wavelength range.
The distinction between these sources is illustrated in Figure bellow. Sources
can also be classified as
continuous sources, which emit radiation continuously with time, or pulsed
sources, which emit radiation in bursts
tungsten lamp of the type used in spectroscopy and its spectrum (b). Intensity of the
tungsten source is usually quite low at wavelengths shorter than about 350 nm. Note that
the intensity reaches a maximum in the near-IR region of the spectrum (<1200 nm in this
case).
Tungsten lamp
• Tungsten/halogen lamp
• Tungsten/halogen lamps, also called quartz/halogen lamps, contain a small amount of iodine
within the quartz envelope that houses the filament. Quartz allows the filament to be operated
at a temperature of about 3500 K, leading to higher intensities and extending the range of the
lamp well into the UV. The lifetime of a tungsten/halogen lamp is more than double that of an
ordinary tungsten lamp, which is limited by sublimation of tungsten from the filament. In the
presence of iodine, the sublimed tungsten reacts to give gaseous WI2 molecules. These
molecules then diffuse back to the hot filament where they decompose, redeposit W atoms on
the filament, and release iodine. Tungsten/halogen lamps are finding ever-increasing use in
spectroscopic instruments because of their extended wavelength range, greater intensity, and
longer life
A deuterium lamp consists of a cylindrical tube containing deuterium at
low pressure, with a quartz window from which the radiation exits
Deuterium lamp
The word laser originally was the upper-case LASER, the acronym from Light
Amplification by Stimulated Emission of Radiation. Lasers have become useful
as sources in certain types of analytical spectroscopy. To understand how a laser
works, consider an assembly of atoms or molecules interacting with an
electromagnetic wave. Laser radiation is highly directional. Spectrally pure,
coherent. and of high intensity. These properties have made possible many
unique research applications that cannot easily be achieved with conventional
sources
Laser Sources
Spectroscopic instruments in the UV and visible regions are usually equipped with one or more
devices to restrict the radiation being measured to a narrow band that is absorbed or emitted by
the analyte. Such devices greatly enhance both the selectivity and the sensitivity of an
instrument. In addition, for absorption measurements, narrow bands of radiation greatly
diminish the chance of Beer's Jaw deviations due to polychromatic radiation. Many
instruments use a monochromator or filter to isolate the desired wavelength band so that only
the band of interest is detected and measured. Others use a spectrograph to spread out, or
disperse, the wavelength so that they can be detected with a multichannel detector
Wavelength Selectors
Monochromators generally have a diffraction grating to disperse the radiation into
its component wavelengths, as shown in Figure bellow
By rotating the grating, different
Wavelengths can be made to pass
through an exit slit
Monochromators consist of:
1- entrance slit
2- collimating mirror or lens
3- a prism or grating
5- focal plane
6- exit slit
Older instruments used prisms for this purpose
Monochromators
Older instruments used prisms for this purpose
Radiation from a source enters the monochromator
via a narrow rectangular opening, or slit.
The radiation is then collimated by a concave
mirror, which produces a parallel beam that
strikes the surface of a reflection grating.
Angular dispersion results from diffraction,
which occurs at the reflective surface
Grating Monochromator
The effective band width of the monochromator depends on the size and quality of the
dispersing element, the slit widths, and the focal length of the monochromator. A high-quality
rnonochromator will exhibit an effective band-width of a few tenths of a nanometer or less in
the ultraviolet/visible region. The effective bandwidth of a monochromator that is satisfactory
for most quantitative applications is about 1 to 20 nm. Many monochromators are equipped
with adjustable slits to permit some control over the bandwidth. A narrow slit decreases the
effective bandwidth but also diminishes the power of the emergent beam. Thus, the minimum
practical bandwidth may be limited by the sensitivity of the detector For qualitative analysis,
narrow slits and minimum effective bandwidths are required if a spectrum is made up of
narrow peaks. For quantitative work. however, wider slits permit operation of the detector
system at lower amplification, which in turn provides greater reproducibility of response.
Most gratings in modern monochromators are replica gratings, which are obtained by
making castings of a master grating. The latter consists of a hard, optically flat, polished
surface on which a suitably shaped diamond tool has created a large number of parallel and
closely spaced grooves. A grating for the ultraviolet and visible region will typically contain
300 to 2000 grooves/mm. with 1200 to 1400 being most common. The construction of a
good master grating is tedious, time consuming, and expensive because the grooves must
be identical in size, exactly parallel, and equally spaced over the length of the grating (3 to
10 cm). Replica gratings are formed from a master grating by a liquid resin casting process
that preserves virtually perfectly the optical accuracy of the original master grating on a clear
resin surface. This surface is ordinarily made reflective by a coating of aluminum or, some
times, gold or platinum
Radiation Filters
Filters operate by absorbing all but a restricted band of radiation from a
continuum source. As shown in Figure bellow, two types of filters are used in
spectroscopy; interference filters and absorption filters. Interference filters are
typically used for absorption measurements, and they generally transmit a much
greater fraction of radiation at their nominal wavelengths than do absorption
filters
Interference Filters
Interference filters are used with ultraviolet and visible radiation, as well as with
wavelengths up to about 14 µm in the infrared region. As the name implies, an
interference filter relies on optical interference to provide a relatively narrow
band of radiation. typically 5 to 20 nm in width. As shown in Figure bellow, an
interference filter consists of a very thin layer of a transparent dielectric material
(frequently calcium fluoride or magnesium fluoride) coated on both sides with a
film of metal that is thin enough to transmit approximately half the radiation
striking it and to reflect the other half.
• the radiant power transmitted, fluoresced, or emitted must be detected in some
manner and converted into a measurable quantity. A detector is a device that
indicates the existence of some physical phenomenon. Familiar examples of
detectors include photographic film (for indicating the presence of
electromagnetic or radioactive radiation.
• The human eye is also a detector; it converts visible radiation into an electrical
signal that is passed to the brain via a chain of neurons in the optic nerve and
produces vision.
Detecting and Measuring Radiant Energy
• The term transducer is used to indicate the type of detector that
converts quantities, such as light intensity, pH, mass, and temperature,
into electrical signals that can be subsequently amplified, manipulated,
and finally converted into numbers proportional to the magnitude of
the original quantity.
• A transducer is a type of detector that converts various types of
chemical and physical quantities into electrical signals such as
electrical charge, current, or voltage.
There are two general types of transducers:
• Photons
All photon detectors are based on the interaction of radiation with a reactive
surface either to produce electrons (photoemission) or to promote electrons to
energy states in which they can conduct electricity (photoconduction). Only
UV, visible, and near-IR radiation possess enough energy to cause
photoemission to occur; thus, photoemissive detectors are limited to
wavelengths shorter than about 2 µm (2000 nm).
Types of Transducers
• Thermal detectors (Heat)
detect a temperature change in a material due to photon absorption
• Thermal detectors can be used over a wide range of wavelengths
.Their main disadvantages are slow response time and lower
sensivity relative to other types of detectors.
Widely used types of photon detectors include phototubes, photomultiplier
tubes, silicon photodiodes, and photodiode arrays.
Photon Detectors
Phototubes
The response of a phototube or a photomultiplier tube is based on the photoelectric effect. a
phototube consists of a semicylindrical photocathode and a wire anode sealed inside an
evacuated transparent glass or quartz envelope. The concave surface of the cathode supports
a layer of photoemissive material. such as an alkali metal or a metal oxide. that emits
electrons when irradiated with light of the appropriate energy. When a voltage is applied
across the electrodes, the emitted photoelectrons are attracted to the positively
charged wire anode, In the complete circuit, a photocurrent then results that is easily
amplified and measured. The number of photoelectrons ejected from the photocathode per
unit time is directly proportional to the radiant power of the beam striking the surface. With
an applied voltage of about 90 V or more, all these photoelectrons are collected at the anode
to give a photocurrent that is also proportional to the radiant power of the beam.
Photoelectrons: are electrons that are ejected
from a photosensitive surface by electromagnetic
radiation.
Photocurrent: is the current in an external
circuit that is limited by the rate of ejection of
photoelectrons.
The photomultiplier tube (PMT) is similar in construction to the phototube but is
significantly more sensitive. Its photocathode is similar to that of the phototube, with electrons
being emitted on exposure to radiation. In place of a single wire anode. however, the PMT has a
series of electrodes called dynodes, The electrons emitted from the cathode are accelerated
toward the first dynode. which is maintained 90 to 100 V positive with respect to the cathode.
Each accelerated photoelectron that strikes the dynode surface produces several electrons, called
secondary electrons, that are then accelerated to dynode 2, which is held 90 to 100 V more
positive than dynode l. Again, electron amplification results. By the time this process has been
repeated at each of the dynodes, 105 to 107 electrons have been produced for each incident
photon. This cascade of electrons is finally collected at the anode to provide an average current
that is further amplified electronically and measured.
One of the major advantages of photomultipliers is their automatic internal amplification.
About 106 to 107 electrons are produced at the anode for each photon that strikes the
photocathode of a photomultiplier tube. Photomultiplier tubes are among the most widely
used types of transducers for detecting ultraviolet/visible radiation. With modern electronic
instrumentation, it is possible to detect the electron pulses resulting from the arrival of
individual photons at the photocathode of a PMT. The pulses are counted, and the
accumulated count is a measure of the intensity of the electromagnetic radiation impinging on
the PMT. Photon counting is advantageous when light intensity, or the frequency of arrival of
photons at t1e photocathode, is low.
Photoconductive transducers consist of a thin film of a semiconductor
material, such as lead sulfide, mercury cadmium telluride (MCT), or
indium antimonide, deposited often on a nonconducting glass surface
and sealed in an evacuated envelope. Absorption of radiation by these
materials promotes nonconducting valence electrons to a higher energy
state, which decreases the electrical resistance of the semiconductor.
Typically, a photoconductor is placed in series with a voltage source and
a load resistor, and the voltage drop across the load resistor serves as a
measure of the radiant power of the beam of radiation. The PbS and InSb
detectors are quite popular in the near-IR region of the spectrum. The
MCT detector is useful in the mid- and far-IR regions when cooled with
liquid N2 to minimize thermal noise.
Photoconductive Cells:
Diode-Array Detectors
Silicon photodiodes have become important recently because 1000 or more can
be fabricated side by side on a single small silicon chip. (The width of individual
diodes is about 0.02 mm). With one or two of the diode-array detectors placed
along the length of the focal plane of a monochromator. All wavelengths can be
monitored simultaneously, thus making high-speed spectroscopy possible. Silicon
photodiode detectors respond extremely rapidly, usually in nanoseconds.
They are more sensitive than a vacuum phototube but considerably less
sensitive than a photomultiplier tube.
Sample containers, which are usually called cells or cuvettes must have windows that are
transparent in the spectral region .Thus as shown in the figure below, quartz or fused silica is
required for the UV region (wavelengths less than 350nm nm) and may be used in the visible
region and out to about 3000 nm (3 µm) in the IR region. Silicate glass is ordinarily used for
the 375 to 2000 nm region because of its low cost compared with quartz. Plastic cells are also
used in the visible region. The most common window material for IR studies is crystalline
sodium chloride, which is soluble in water and in some other solvents.
Sample Containers
The best cells have windows that
are perpendicular to the direction
of the beam in order to minimize
reflection losses. The most
common cell path length for
studies in the UV and visible
regions is 1 cm; matched,
calibrated cells of this size are
available from several
commercial sources. Many other
cells with shorter and longer path
lengths can be purchased.
For reasons of economy, cylindrical cells are sometimes used. Fingerprints, grease, or other
deposits on the walls markedly alter the transmission characteristics of a cell. Thus,
thorough cleaning before and after use is need , and care must be taken to avoid touching
the windows after cleaning is complete. Matched cells should never be dried by heating in
an oven or over a flame because this may cause physical damage or a change in path
length..
Photometers provide simple, relatively inexpensive tools for performing
absorption measurements. Filter photometers are often more convenient and
more rugged and are easier to maintain and use than the more sophisticated
spectrophotometers. Furthermore, photometers characteristically have high
radiant energy throughputs and thus good signal-to-noise ratios even with
relatively simple and inexpensive transducers and circuitry. Photometers have
the advantages of simplicity, ruggedness, and low cost.
Spectrophotometers
offer the considerable advantage that the wavelength can be varied
continuously, thus making it possible to record absorption spectra and is a
scanning instrument (i.e. a spectrophotometer has a monochromator for
separating the individual wavelengths of light). Several dozen models of
spectrophotometers are available commercially. Most spectrophotometers
cover the UV /visible and occasionally the near-infrared region, while
photometers are most often used for the visible region.. Both photometers and
spectrophotometers can be obtained in single- and double-beam varieties.
Single beam spectrometers
Single beam spectrometers are relatively cheap, simple, portable & ideally
suited to quantitative analysis .It is not possible to scan through the entire
spectrum with such an instrument because both the source intensity & the
detector response vary with the wavelength. To record an accurate value of
the absorbance it is necessary to zero the instrument on a reference/blank
before every measurement. Thus, this is essentially a single wavelength
measurement of absorbance.
Voltage fluctuations and changes in light source present a problem
When a heavy load is placed on the electric power system, lights dim and later brighten
*If measurements are being taken on the spectrophotometer at the same time, the
readings will be unreliable
*Aging lamp source may momentarily flicker and cause the readings to be unstable and
errorneous
So, a single-beam instrument requires a stabilized voltage supply to avoid
errors resulting from changes in the beam intensity during the time required to
make the 100% T measurement and determine %T for the analyte.
Disadvantages of single-beam spectrophotometer
The second type of double-beam instrument. Here the beams are separated in time by a
rotating sector mirror that directs the entire beam from the monochromator first through the
reference cell and then through the sample cell. The pulses of radiation are recombined by
another sector mirror, which transmits one pulse and reflects the other to the transducer. As
shown by the insert labeled "front view", the motor-driven sector mirror is made up of pie-
shape segments, half of which are mirrored and half of which are transparent. The mirrored
sections are held in place by blackened metal frames that periodically interrupt the beam and
prevent its reaching the transducer. The double-beam-in-time approach is generally preferred
because of the difficulty in matching the two detectors needed for the double-beam-in-space
design
Double-Beam in time spectrophotometers
Double-Beam in space spectrophotometers
• Many modern photometers and spectrophotometers are based on a double-beam
design. A double-beam-in-space instrument in which two beams are formed in
space by a V-shape mirror called a beamsplitter. One beam passes through the
reference solution to a photodetector, and the second simultaneously traverses
the sample to a second, matched detector. The two outputs are amplified, and
their ratio (or the logarithm of their ratio) is determined electronically or by a
computer and displayed by the readout device.
The determination of an analyte’s concentration based on its absorption of ultraviolet or visible radiation is one of the most
frequently quantitative analytical methods. One reason for its popularity is that many organic and inorganic compounds
have strong absorption bands in the UV/Vis region of the electromagnetic spectrum.
There are many application
• Environmental Applications
• Clinical Applications
• Industrial Analysis
• Forensic Applications
Quantitative Applications
• Methods for the analysis of waters and wastewaters relying on the
absorption of UV/Vis radiation are among some of the most frequently
employed analytical methods.
Environmental Applications
• UV/Vis molecular absorption is one of the most commonly employed
techniques for the analysis of clinical samples, several examples of
which are listed in Table below. The analysis of clinical samples is often
complicated by the complexity of the sample matrix, which may
contribute a significant background absorption at the desired wavelength
.
Clinical Applications
The Application of UV/Vis Molecular Absorption to the
Analysis of Clinical Samples
• UV/Vis molecular absorption is used for the analysis of a diverse array of
industrial samples, including pharmaceuticals, food, paint, glass, and
metals.
• In many cases the methods are Products that have been analyzed in this
fashion include antibiotics, hormones, vitamins, and analgesics.
• One example of the use of UV absorption is in determining the purity of
aspirin tablets.
Industrial Analysis
• UV/Vis molecular absorption is routinely used in the analysis of
narcotics and for drug testing.
• One interesting forensic application is the determination of blood
alcohol using the Breathalyzer test. In this test a 52.5-mL breath
sample is bubbled through an acidified solution of K2Cr2O7. Any
ethanol present in the breath sample is oxidized by the dichromate,
producing acetic acid and Cr3+ as products.
Forensic Applications
• The energy at which the absorption occurs, as well as the intensity of the
absorption, is determined by the chemical environment of the absorbing moiety.
For example, benzene has several ultraviolet absorption bands due to p  p*
transitions. The position and intensity of two of these bands, 203.5 nm (e =
7400) and 254 nm (e = 204), are very sensitive to substitution. For benzoic acid,
in which a carboxylic acid group replaces one of the aromatic hydrogens, the
two bands shift to 230 nm (e = 11,600) and 273 nm (e = 970). Several rules
have been developed to aid in correlating UV/Vis absorption bands to chemical
structure.
Qualitative Applications
1-Limited since few resolved peaks
• Unambiguous identification not usually possible.
Why we cannot use UV / visible Spectroscopy in
Qualitative Analysis ?
Loss of fine structure for
acetaldehyde when
transfer to solvent from
gas phase
Also need to consider
absorbance of solvent
2. Solvent can affect position and shape of curve .
• polar solvents broaden out peaks, eliminates fine structure.
•Loss of fine structure
for 1,2,4,5-tetrazine as
solvent polarity
increases
3.Solvent can also absorb in UV-vis spectrum.
• Solvent for the ultraviolet and visible regions
• Molecular absorption, particularly in the UV/Vis range, has been used for a variety of
different characterization studies, including determining the stoichiometry of metal–
ligand complexes and determining equilibrium constants.
• Stoichiometry of a Metal, Ligand Complex
• The stoichiometry for a metal–ligand complexation reaction of the following general
form
• can be determined by one of three methods: the method of continuous variations,
•the mole-ratio method, and the slope-ratio method
Characterization Applications
• also called Job’s method, is the most popular. In this method a series of
solutions is prepared such that the total moles of metal and ligand, ntot, in
each solution is the same. Thus, if (nM)i and (nL)i are, respectively, the
moles of metal and ligand in the i-th solution, Then
• The relative amount of ligand and metal in each solution is expressed as the mole
• fraction of ligand, (XL)i, and the mole fraction of metal, (XM)i,
Method of continuous variations
A plot of A vs volume ratio (volume ratio =
mole fraction) gives maximum absorbance when
there is a stoichiometric amount of the two.
• A procedure for determining the stoichiometry between two reactants by
preparing solutions containing different mole ratios of two reactants . In
the mole-ratio method the moles of one reactant, usually the metal, are
held constant, while the moles of the other reactant are varied. The
absorbance is monitored at a wavelength at which the metal–ligand
complex absorbs. A plot of absorbance as a function of the ligand-to-
metal mole ratio (nL/nM) has two linear branches that intersect at a mole
ratio corresponding to the formula of the complex.
Mole-ratio method
• Figure (a)shows a mole-ratio plot for the formation of a 1:1 complex
in which the absorbance is monitored at a wavelength at which only
the complex absorbs.
• Figure (b) shows a mole-ratio plot for a 1:2 complex in which the
metal, the ligand, and the complex absorb at the selected wavelength.
• A procedure for determining the stoichiometry between two reactants by
measuring the relative change in absorbance under conditions when each reactant
is the limiting reagent . In the slope-ratio method two sets of solutions are
prepared. The first set consists of a constant amount of metal and a variable
amount of ligand, chosen such that the total concentration of metal, CM, is much
greater than the total concentration of ligand, CL. Under these conditions we
may assume that essentially all the ligand is complexed. The concentration of a
metal–ligand complex of the general form MxLy is
Slope-ratio method
• Determine endpoint by following change in absorbance of:
1) reactant (decrease)
2) product (increase)
3) titrant (increase after endpoint)
• Example Titration curves for
• 𝑺 + 𝑻 → 𝑷
• where S = analyte being titrated, T = titrant, P = product
Photometric Titrations
where S = analyte being titrated, T = titrant, P = product
References
1- Douglas A. skoog, Donald M. West, F. james Holler, Stanley R. Crouch. (2013).
Fundamentals of analytical chemistry. 9th ed . Belmont, USA: Mary finch. pp:651-674.
2- Douglas A. Skoog, F. James Holler, Stanley R. Crouch. (2007). Principles of instrumental
analysis. 6th ed. Belmont, USA: David Harris.pp:335-367.
3- Tony O.(2000). Fundamentals of modern Uv-Visible spectroscopy. 1st ed. Germany:
Agilent technologies .pp:36-43.
4- David H.(2000). Modern analytical chemistry. 1st ed . London: James M smith. Pp: 394-
407.
Uv-visible spectroscopy

Weitere ähnliche Inhalte

Was ist angesagt?

Instrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: DetectorsInstrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: DetectorsVrushali Tambe
 
Uv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationUv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationAnusreeAnu11
 
Uv visible spectroscopy- madan
Uv visible spectroscopy- madanUv visible spectroscopy- madan
Uv visible spectroscopy- madanMadan Sigdel
 
INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPY
INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPYINSTRUMENTATION OF UV-VISIBLE SPECTROSCOPY
INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPYCh. Bhargava krishna
 
INTRODUCTION TO UV-VISIBLE SPECTROSCOPY
INTRODUCTION TO UV-VISIBLE SPECTROSCOPYINTRODUCTION TO UV-VISIBLE SPECTROSCOPY
INTRODUCTION TO UV-VISIBLE SPECTROSCOPYJunaid Khan
 
IR interpretation and sample handling
 IR interpretation and sample handling  IR interpretation and sample handling
IR interpretation and sample handling Afzaye Rasul
 
UV Spectroscopy and Its Applications
UV Spectroscopy  and Its ApplicationsUV Spectroscopy  and Its Applications
UV Spectroscopy and Its ApplicationsVikram Choudhary
 
Instrumentation of IR spectroscopy
Instrumentation of IR spectroscopyInstrumentation of IR spectroscopy
Instrumentation of IR spectroscopyTalha Liaqat
 
C-13 NMR Spectroscopy
C-13 NMR SpectroscopyC-13 NMR Spectroscopy
C-13 NMR SpectroscopyMANISHSAHU106
 
13C-NMR SPECTROSCOPY
13C-NMR SPECTROSCOPY13C-NMR SPECTROSCOPY
13C-NMR SPECTROSCOPYramanbrar09
 
uv -visible spectroscopy
 uv -visible spectroscopy uv -visible spectroscopy
uv -visible spectroscopykeshav pai
 
Infrared spectroscopy
Infrared spectroscopy Infrared spectroscopy
Infrared spectroscopy Prachi Pathak
 
NMR SPECTROSCOPY
NMR SPECTROSCOPYNMR SPECTROSCOPY
NMR SPECTROSCOPYVidyaNani
 
Factors affecting IR absorption frequency
Factors affecting IR absorption frequency Factors affecting IR absorption frequency
Factors affecting IR absorption frequency Vrushali Tambe
 
Flourescence spectroscopy- instrumentation and applications
Flourescence spectroscopy-  instrumentation and applicationsFlourescence spectroscopy-  instrumentation and applications
Flourescence spectroscopy- instrumentation and applicationssinghsnehi01
 
Flame emission spectroscopy
Flame emission spectroscopyFlame emission spectroscopy
Flame emission spectroscopyMehulJain143
 
UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...
UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...
UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...Rajesh Singh
 
Nuclear Magnetic Double Resonance (Decoupling).pptx
Nuclear Magnetic Double Resonance (Decoupling).pptxNuclear Magnetic Double Resonance (Decoupling).pptx
Nuclear Magnetic Double Resonance (Decoupling).pptxRushikeshTidake
 

Was ist angesagt? (20)

Instrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: DetectorsInstrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: Detectors
 
Uv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationUv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentation
 
Uv visible spectroscopy- madan
Uv visible spectroscopy- madanUv visible spectroscopy- madan
Uv visible spectroscopy- madan
 
INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPY
INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPYINSTRUMENTATION OF UV-VISIBLE SPECTROSCOPY
INSTRUMENTATION OF UV-VISIBLE SPECTROSCOPY
 
INTRODUCTION TO UV-VISIBLE SPECTROSCOPY
INTRODUCTION TO UV-VISIBLE SPECTROSCOPYINTRODUCTION TO UV-VISIBLE SPECTROSCOPY
INTRODUCTION TO UV-VISIBLE SPECTROSCOPY
 
IR interpretation and sample handling
 IR interpretation and sample handling  IR interpretation and sample handling
IR interpretation and sample handling
 
UV Spectroscopy and Its Applications
UV Spectroscopy  and Its ApplicationsUV Spectroscopy  and Its Applications
UV Spectroscopy and Its Applications
 
Instrumentation of IR spectroscopy
Instrumentation of IR spectroscopyInstrumentation of IR spectroscopy
Instrumentation of IR spectroscopy
 
C-13 NMR Spectroscopy
C-13 NMR SpectroscopyC-13 NMR Spectroscopy
C-13 NMR Spectroscopy
 
13C-NMR SPECTROSCOPY
13C-NMR SPECTROSCOPY13C-NMR SPECTROSCOPY
13C-NMR SPECTROSCOPY
 
uv -visible spectroscopy
 uv -visible spectroscopy uv -visible spectroscopy
uv -visible spectroscopy
 
Mass Spectroscopy
Mass Spectroscopy Mass Spectroscopy
Mass Spectroscopy
 
Infrared spectroscopy
Infrared spectroscopy Infrared spectroscopy
Infrared spectroscopy
 
NMR SPECTROSCOPY
NMR SPECTROSCOPYNMR SPECTROSCOPY
NMR SPECTROSCOPY
 
Factors affecting IR absorption frequency
Factors affecting IR absorption frequency Factors affecting IR absorption frequency
Factors affecting IR absorption frequency
 
Flourescence spectroscopy- instrumentation and applications
Flourescence spectroscopy-  instrumentation and applicationsFlourescence spectroscopy-  instrumentation and applications
Flourescence spectroscopy- instrumentation and applications
 
Flame emission spectroscopy
Flame emission spectroscopyFlame emission spectroscopy
Flame emission spectroscopy
 
UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...
UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...
UV spectroscopy, Electronic transitions, law of UV, Deviations of UV, chromop...
 
Nuclear Magnetic Double Resonance (Decoupling).pptx
Nuclear Magnetic Double Resonance (Decoupling).pptxNuclear Magnetic Double Resonance (Decoupling).pptx
Nuclear Magnetic Double Resonance (Decoupling).pptx
 
FT NMR
FT NMRFT NMR
FT NMR
 

Andere mochten auch

UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible SpectroscopySudha Reddy
 
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...mariomS7
 
Seminar review 1
Seminar review 1Seminar review 1
Seminar review 1AJAL A J
 
New ppt of uv visible
New ppt of uv visibleNew ppt of uv visible
New ppt of uv visiblekeshav pai
 
Optical Emission Spectrometry OES
Optical Emission Spectrometry  OESOptical Emission Spectrometry  OES
Optical Emission Spectrometry OESGamal Abdel Hamid
 
UV Visible Spectroscopy
UV Visible SpectroscopyUV Visible Spectroscopy
UV Visible SpectroscopyM.T.H Group
 
Go ions v2_021312
Go ions v2_021312Go ions v2_021312
Go ions v2_021312Femi Prince
 
backward wave oscillator
backward wave oscillatorbackward wave oscillator
backward wave oscillatorAJAL A J
 
Application of u.v. spectroscopy
Application of u.v. spectroscopyApplication of u.v. spectroscopy
Application of u.v. spectroscopyManish Kumar
 
Theory of IR spectroscopy
Theory of IR spectroscopyTheory of IR spectroscopy
Theory of IR spectroscopychiranjibi68
 
Microwave Spectroscopy
Microwave SpectroscopyMicrowave Spectroscopy
Microwave Spectroscopykrishslide
 
Spectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationSpectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationBasil "Lexi" Bruno
 
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYINSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYAshwini Somayaji
 
INTRODUCTION TO SPECTROSCOPY
INTRODUCTION TO SPECTROSCOPYINTRODUCTION TO SPECTROSCOPY
INTRODUCTION TO SPECTROSCOPYSWAPNIL NIGAM
 

Andere mochten auch (20)

UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible Spectroscopy
 
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,differ...
 
UV visible spectroscopy
UV visible spectroscopyUV visible spectroscopy
UV visible spectroscopy
 
Seminar review 1
Seminar review 1Seminar review 1
Seminar review 1
 
New ppt of uv visible
New ppt of uv visibleNew ppt of uv visible
New ppt of uv visible
 
Noble Gases
Noble GasesNoble Gases
Noble Gases
 
Optical Emission Spectrometry OES
Optical Emission Spectrometry  OESOptical Emission Spectrometry  OES
Optical Emission Spectrometry OES
 
UV Visible Spectroscopy
UV Visible SpectroscopyUV Visible Spectroscopy
UV Visible Spectroscopy
 
Go ions v2_021312
Go ions v2_021312Go ions v2_021312
Go ions v2_021312
 
Mm wave
Mm waveMm wave
Mm wave
 
backward wave oscillator
backward wave oscillatorbackward wave oscillator
backward wave oscillator
 
Spectroscopy
SpectroscopySpectroscopy
Spectroscopy
 
Molecular Spectroscopy
Molecular SpectroscopyMolecular Spectroscopy
Molecular Spectroscopy
 
Application of u.v. spectroscopy
Application of u.v. spectroscopyApplication of u.v. spectroscopy
Application of u.v. spectroscopy
 
Theory of IR spectroscopy
Theory of IR spectroscopyTheory of IR spectroscopy
Theory of IR spectroscopy
 
Klystron 1
Klystron 1Klystron 1
Klystron 1
 
Microwave Spectroscopy
Microwave SpectroscopyMicrowave Spectroscopy
Microwave Spectroscopy
 
Spectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and applicationSpectrophotometry: basic concepts, instrumentation and application
Spectrophotometry: basic concepts, instrumentation and application
 
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRYINSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
INSTRUMENTATION OF UV-VISIBLE SPECTROPHOTOMETRY
 
INTRODUCTION TO SPECTROSCOPY
INTRODUCTION TO SPECTROSCOPYINTRODUCTION TO SPECTROSCOPY
INTRODUCTION TO SPECTROSCOPY
 

Ähnlich wie Uv-visible spectroscopy

UV visible Specrosopy and applications
UV visible Specrosopy and applicationsUV visible Specrosopy and applications
UV visible Specrosopy and applicationsNAFIHHASSANMEHABANI
 
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...nikhil kadam
 
Uv visible Spectroscopy
Uv visible SpectroscopyUv visible Spectroscopy
Uv visible Spectroscopyknowledge1995
 
Ultraviolet spetroscopy by Dr. Monika Singh part-1 as per PCI syllabus
Ultraviolet spetroscopy by Dr. Monika Singh part-1 as per PCI syllabusUltraviolet spetroscopy by Dr. Monika Singh part-1 as per PCI syllabus
Ultraviolet spetroscopy by Dr. Monika Singh part-1 as per PCI syllabusMonika Singh
 
UV ray spectrophotometer
UV ray spectrophotometerUV ray spectrophotometer
UV ray spectrophotometerGoa App
 
UV rays
UV rays UV rays
UV rays Goa App
 
uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
 uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICSHarevindarsingh
 
UV-visible spectroscopy - 2021
UV-visible spectroscopy - 2021UV-visible spectroscopy - 2021
UV-visible spectroscopy - 2021Ahmed Metwaly
 
uv -visible spectroscopy
uv -visible spectroscopyuv -visible spectroscopy
uv -visible spectroscopyyogitamandlik2
 
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...Dr. Amsavel A
 
Uv/visible spectroscopy
Uv/visible spectroscopyUv/visible spectroscopy
Uv/visible spectroscopySavitaBhatt2
 
Ultraviolet and visible spectroscopy
Ultraviolet and visible  spectroscopyUltraviolet and visible  spectroscopy
Ultraviolet and visible spectroscopyGeetaraniLoushigam
 

Ähnlich wie Uv-visible spectroscopy (20)

UV visible Specrosopy and applications
UV visible Specrosopy and applicationsUV visible Specrosopy and applications
UV visible Specrosopy and applications
 
UV Visible Spectrometry
UV Visible SpectrometryUV Visible Spectrometry
UV Visible Spectrometry
 
Presentation2
Presentation2Presentation2
Presentation2
 
UV-Visible spectroscopy
UV-Visible spectroscopyUV-Visible spectroscopy
UV-Visible spectroscopy
 
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...
 
Uv visible Spectroscopy
Uv visible SpectroscopyUv visible Spectroscopy
Uv visible Spectroscopy
 
Ultraviolet spetroscopy by Dr. Monika Singh part-1 as per PCI syllabus
Ultraviolet spetroscopy by Dr. Monika Singh part-1 as per PCI syllabusUltraviolet spetroscopy by Dr. Monika Singh part-1 as per PCI syllabus
Ultraviolet spetroscopy by Dr. Monika Singh part-1 as per PCI syllabus
 
UV ray spectrophotometer
UV ray spectrophotometerUV ray spectrophotometer
UV ray spectrophotometer
 
UV rays
UV rays UV rays
UV rays
 
Prabhakar singh ii sem-paper v-colorimeter &amp; spectrophotometer
Prabhakar singh  ii sem-paper v-colorimeter &amp; spectrophotometerPrabhakar singh  ii sem-paper v-colorimeter &amp; spectrophotometer
Prabhakar singh ii sem-paper v-colorimeter &amp; spectrophotometer
 
uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
 uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
uv spectroscopy by HARVINDAR SINGH .M.PHARM PHARMACEUTICS
 
UV-visible spectroscopy - 2021
UV-visible spectroscopy - 2021UV-visible spectroscopy - 2021
UV-visible spectroscopy - 2021
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
uv -visible spectroscopy
uv -visible spectroscopyuv -visible spectroscopy
uv -visible spectroscopy
 
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
UV -Vis Spectrophotometry- Principle, Theory, Instrumentation and Application...
 
Uv/visible spectroscopy
Uv/visible spectroscopyUv/visible spectroscopy
Uv/visible spectroscopy
 
Uv
UvUv
Uv
 
Ultraviolet and visible spectroscopy
Ultraviolet and visible  spectroscopyUltraviolet and visible  spectroscopy
Ultraviolet and visible spectroscopy
 
UV / Visible Spectroscopy
UV / Visible SpectroscopyUV / Visible Spectroscopy
UV / Visible Spectroscopy
 

Mehr von Mzgin Mohammed

Organic identification
Organic identificationOrganic identification
Organic identificationMzgin Mohammed
 
preparation of dibenzalacetone
preparation of dibenzalacetonepreparation of dibenzalacetone
preparation of dibenzalacetoneMzgin Mohammed
 
Types of chemical bonds
Types of chemical bondsTypes of chemical bonds
Types of chemical bondsMzgin Mohammed
 
Transition in ultraviolet and visible light
Transition in ultraviolet and visible lightTransition in ultraviolet and visible light
Transition in ultraviolet and visible lightMzgin Mohammed
 
Temperature and heat hazard by mzgin m. ayoob
Temperature and heat hazard by mzgin m. ayoobTemperature and heat hazard by mzgin m. ayoob
Temperature and heat hazard by mzgin m. ayoobMzgin Mohammed
 
Artificial photosynthesis
Artificial photosynthesisArtificial photosynthesis
Artificial photosynthesisMzgin Mohammed
 

Mehr von Mzgin Mohammed (8)

Organic identification
Organic identificationOrganic identification
Organic identification
 
preparation of dibenzalacetone
preparation of dibenzalacetonepreparation of dibenzalacetone
preparation of dibenzalacetone
 
5 benzocaine
5 benzocaine5 benzocaine
5 benzocaine
 
Geometrical isomers
Geometrical isomersGeometrical isomers
Geometrical isomers
 
Types of chemical bonds
Types of chemical bondsTypes of chemical bonds
Types of chemical bonds
 
Transition in ultraviolet and visible light
Transition in ultraviolet and visible lightTransition in ultraviolet and visible light
Transition in ultraviolet and visible light
 
Temperature and heat hazard by mzgin m. ayoob
Temperature and heat hazard by mzgin m. ayoobTemperature and heat hazard by mzgin m. ayoob
Temperature and heat hazard by mzgin m. ayoob
 
Artificial photosynthesis
Artificial photosynthesisArtificial photosynthesis
Artificial photosynthesis
 

Kürzlich hochgeladen

Environmental acoustics- noise criteria.pptx
Environmental acoustics- noise criteria.pptxEnvironmental acoustics- noise criteria.pptx
Environmental acoustics- noise criteria.pptxpriyankatabhane
 
Advances in AI-driven Image Recognition for Early Detection of Cancer
Advances in AI-driven Image Recognition for Early Detection of CancerAdvances in AI-driven Image Recognition for Early Detection of Cancer
Advances in AI-driven Image Recognition for Early Detection of CancerLuis Miguel Chong Chong
 
Q4-Mod-1c-Quiz-Projectile-333344444.pptx
Q4-Mod-1c-Quiz-Projectile-333344444.pptxQ4-Mod-1c-Quiz-Projectile-333344444.pptx
Q4-Mod-1c-Quiz-Projectile-333344444.pptxtuking87
 
Timeless Cosmology: Towards a Geometric Origin of Cosmological Correlations
Timeless Cosmology: Towards a Geometric Origin of Cosmological CorrelationsTimeless Cosmology: Towards a Geometric Origin of Cosmological Correlations
Timeless Cosmology: Towards a Geometric Origin of Cosmological CorrelationsDanielBaumann11
 
BACTERIAL SECRETION SYSTEM by Dr. Chayanika Das
BACTERIAL SECRETION SYSTEM by Dr. Chayanika DasBACTERIAL SECRETION SYSTEM by Dr. Chayanika Das
BACTERIAL SECRETION SYSTEM by Dr. Chayanika DasChayanika Das
 
final waves properties grade 7 - third quarter
final waves properties grade 7 - third quarterfinal waves properties grade 7 - third quarter
final waves properties grade 7 - third quarterHanHyoKim
 
The Sensory Organs, Anatomy and Function
The Sensory Organs, Anatomy and FunctionThe Sensory Organs, Anatomy and Function
The Sensory Organs, Anatomy and FunctionJadeNovelo1
 
Loudspeaker- direct radiating type and horn type.pptx
Loudspeaker- direct radiating type and horn type.pptxLoudspeaker- direct radiating type and horn type.pptx
Loudspeaker- direct radiating type and horn type.pptxpriyankatabhane
 
Science (Communication) and Wikipedia - Potentials and Pitfalls
Science (Communication) and Wikipedia - Potentials and PitfallsScience (Communication) and Wikipedia - Potentials and Pitfalls
Science (Communication) and Wikipedia - Potentials and PitfallsDobusch Leonhard
 
linear Regression, multiple Regression and Annova
linear Regression, multiple Regression and Annovalinear Regression, multiple Regression and Annova
linear Regression, multiple Regression and AnnovaMansi Rastogi
 
6.2 Pests of Sesame_Identification_Binomics_Dr.UPR
6.2 Pests of Sesame_Identification_Binomics_Dr.UPR6.2 Pests of Sesame_Identification_Binomics_Dr.UPR
6.2 Pests of Sesame_Identification_Binomics_Dr.UPRPirithiRaju
 
LAMP PCR.pptx by Dr. Chayanika Das, Ph.D, Veterinary Microbiology
LAMP PCR.pptx by Dr. Chayanika Das, Ph.D, Veterinary MicrobiologyLAMP PCR.pptx by Dr. Chayanika Das, Ph.D, Veterinary Microbiology
LAMP PCR.pptx by Dr. Chayanika Das, Ph.D, Veterinary MicrobiologyChayanika Das
 
6.1 Pests of Groundnut_Binomics_Identification_Dr.UPR
6.1 Pests of Groundnut_Binomics_Identification_Dr.UPR6.1 Pests of Groundnut_Binomics_Identification_Dr.UPR
6.1 Pests of Groundnut_Binomics_Identification_Dr.UPRPirithiRaju
 
Oxo-Acids of Halogens and their Salts.pptx
Oxo-Acids of Halogens and their Salts.pptxOxo-Acids of Halogens and their Salts.pptx
Oxo-Acids of Halogens and their Salts.pptxfarhanvvdk
 
FBI Profiling - Forensic Psychology.pptx
FBI Profiling - Forensic Psychology.pptxFBI Profiling - Forensic Psychology.pptx
FBI Profiling - Forensic Psychology.pptxPayal Shrivastava
 
complex analysis best book for solving questions.pdf
complex analysis best book for solving questions.pdfcomplex analysis best book for solving questions.pdf
complex analysis best book for solving questions.pdfSubhamKumar3239
 
Combining Asynchronous Task Parallelism and Intel SGX for Secure Deep Learning
Combining Asynchronous Task Parallelism and Intel SGX for Secure Deep LearningCombining Asynchronous Task Parallelism and Intel SGX for Secure Deep Learning
Combining Asynchronous Task Parallelism and Intel SGX for Secure Deep Learningvschiavoni
 
Charateristics of the Angara-A5 spacecraft launched from the Vostochny Cosmod...
Charateristics of the Angara-A5 spacecraft launched from the Vostochny Cosmod...Charateristics of the Angara-A5 spacecraft launched from the Vostochny Cosmod...
Charateristics of the Angara-A5 spacecraft launched from the Vostochny Cosmod...Christina Parmionova
 
Introduction of Human Body & Structure of cell.pptx
Introduction of Human Body & Structure of cell.pptxIntroduction of Human Body & Structure of cell.pptx
Introduction of Human Body & Structure of cell.pptxMedical College
 

Kürzlich hochgeladen (20)

Environmental acoustics- noise criteria.pptx
Environmental acoustics- noise criteria.pptxEnvironmental acoustics- noise criteria.pptx
Environmental acoustics- noise criteria.pptx
 
Advances in AI-driven Image Recognition for Early Detection of Cancer
Advances in AI-driven Image Recognition for Early Detection of CancerAdvances in AI-driven Image Recognition for Early Detection of Cancer
Advances in AI-driven Image Recognition for Early Detection of Cancer
 
Q4-Mod-1c-Quiz-Projectile-333344444.pptx
Q4-Mod-1c-Quiz-Projectile-333344444.pptxQ4-Mod-1c-Quiz-Projectile-333344444.pptx
Q4-Mod-1c-Quiz-Projectile-333344444.pptx
 
Timeless Cosmology: Towards a Geometric Origin of Cosmological Correlations
Timeless Cosmology: Towards a Geometric Origin of Cosmological CorrelationsTimeless Cosmology: Towards a Geometric Origin of Cosmological Correlations
Timeless Cosmology: Towards a Geometric Origin of Cosmological Correlations
 
BACTERIAL SECRETION SYSTEM by Dr. Chayanika Das
BACTERIAL SECRETION SYSTEM by Dr. Chayanika DasBACTERIAL SECRETION SYSTEM by Dr. Chayanika Das
BACTERIAL SECRETION SYSTEM by Dr. Chayanika Das
 
final waves properties grade 7 - third quarter
final waves properties grade 7 - third quarterfinal waves properties grade 7 - third quarter
final waves properties grade 7 - third quarter
 
The Sensory Organs, Anatomy and Function
The Sensory Organs, Anatomy and FunctionThe Sensory Organs, Anatomy and Function
The Sensory Organs, Anatomy and Function
 
Loudspeaker- direct radiating type and horn type.pptx
Loudspeaker- direct radiating type and horn type.pptxLoudspeaker- direct radiating type and horn type.pptx
Loudspeaker- direct radiating type and horn type.pptx
 
Science (Communication) and Wikipedia - Potentials and Pitfalls
Science (Communication) and Wikipedia - Potentials and PitfallsScience (Communication) and Wikipedia - Potentials and Pitfalls
Science (Communication) and Wikipedia - Potentials and Pitfalls
 
linear Regression, multiple Regression and Annova
linear Regression, multiple Regression and Annovalinear Regression, multiple Regression and Annova
linear Regression, multiple Regression and Annova
 
6.2 Pests of Sesame_Identification_Binomics_Dr.UPR
6.2 Pests of Sesame_Identification_Binomics_Dr.UPR6.2 Pests of Sesame_Identification_Binomics_Dr.UPR
6.2 Pests of Sesame_Identification_Binomics_Dr.UPR
 
LAMP PCR.pptx by Dr. Chayanika Das, Ph.D, Veterinary Microbiology
LAMP PCR.pptx by Dr. Chayanika Das, Ph.D, Veterinary MicrobiologyLAMP PCR.pptx by Dr. Chayanika Das, Ph.D, Veterinary Microbiology
LAMP PCR.pptx by Dr. Chayanika Das, Ph.D, Veterinary Microbiology
 
6.1 Pests of Groundnut_Binomics_Identification_Dr.UPR
6.1 Pests of Groundnut_Binomics_Identification_Dr.UPR6.1 Pests of Groundnut_Binomics_Identification_Dr.UPR
6.1 Pests of Groundnut_Binomics_Identification_Dr.UPR
 
Let’s Say Someone Did Drop the Bomb. Then What?
Let’s Say Someone Did Drop the Bomb. Then What?Let’s Say Someone Did Drop the Bomb. Then What?
Let’s Say Someone Did Drop the Bomb. Then What?
 
Oxo-Acids of Halogens and their Salts.pptx
Oxo-Acids of Halogens and their Salts.pptxOxo-Acids of Halogens and their Salts.pptx
Oxo-Acids of Halogens and their Salts.pptx
 
FBI Profiling - Forensic Psychology.pptx
FBI Profiling - Forensic Psychology.pptxFBI Profiling - Forensic Psychology.pptx
FBI Profiling - Forensic Psychology.pptx
 
complex analysis best book for solving questions.pdf
complex analysis best book for solving questions.pdfcomplex analysis best book for solving questions.pdf
complex analysis best book for solving questions.pdf
 
Combining Asynchronous Task Parallelism and Intel SGX for Secure Deep Learning
Combining Asynchronous Task Parallelism and Intel SGX for Secure Deep LearningCombining Asynchronous Task Parallelism and Intel SGX for Secure Deep Learning
Combining Asynchronous Task Parallelism and Intel SGX for Secure Deep Learning
 
Charateristics of the Angara-A5 spacecraft launched from the Vostochny Cosmod...
Charateristics of the Angara-A5 spacecraft launched from the Vostochny Cosmod...Charateristics of the Angara-A5 spacecraft launched from the Vostochny Cosmod...
Charateristics of the Angara-A5 spacecraft launched from the Vostochny Cosmod...
 
Introduction of Human Body & Structure of cell.pptx
Introduction of Human Body & Structure of cell.pptxIntroduction of Human Body & Structure of cell.pptx
Introduction of Human Body & Structure of cell.pptx
 

Uv-visible spectroscopy

  • 1.
  • 2. Prepared by: Karwan O. Ali Yousif T. Maaroof Mzgin. M. Ayoob UV- VISIBLE SPECTROSCOPY
  • 3. What is spectroscopy? The interactions of radiation and matter are the subject of the science called spectroscopy. Spectroscopic analytical methods are based on measuring the amount of radiation produced or absorbed by molecular or atomic species of interest
  • 5. Light exhibits wave property during its propagation and energy particle during its interaction with matter. The double nature of light (waves and particles) is known as dualism. Dual nature of light
  • 6. • Light consist of energy packets, known as photons. • The energy (E) of photons is proportional to the frequency i.e. related to c and . It can be expressed by max plank relation: • E = h  ( = C /) • where h = max plank constant = 6.63 x 10-27 erg., sec.) • i.e. E   or E  1/   
  • 7. For analytical purposes we use the region of I.R, visible and U.V radiations. UV radiation region is classified into : far UV from (10nm-200nm) and near UV from (200nm-380nm
  • 8. Why Is a Red Solution Red?
  • 9.
  • 10. • When a molecule interact with radiant energy, the molecule is said to be excited , because the outer valence electrons undergo transition from original energy level ground state (E g) to an exited state (E s). Interaction of a substance with EMR Excited state Es Ground state E g
  • 11. When a molecule in the ground state absorbs EMR, 3 energy state transitions will take place. •These types of transition are: 1) Electronic 2) Vibrational 3) Rotational.
  • 12. • When the molecule absorb Visible and U.V region. Raising electrons to a higher energy level, raising the Vibration of molecule, and increasing rotation of the molecule (Electronic transition energy + Vibrational transition energy + Rotational transition energy) 1) Electronic
  • 13. • When the molecule absorb I.R region. Raising the Vibration of molecule and increasing rotation of the molecule (Vibrational transition energy + Rotational transition energy) 2) Vibrational
  • 14. • When the molecule absorbs F.I.R and Microwave regions. Increasing rotation of the molecule (Rotational transition energy). 3) Rotational
  • 15.
  • 16. • Absorption measurements based upon ultraviolet and visible radiation find widespread application for the quantitative determination of a large variety species [1]. • Beer’s Law: • A = -logT = logP0/P = abc = bc = 2 - log%T • Where T = transmittance, T = percentage transmittance, P = transmitted power of radiation, Po = incident power of radiation, A = absorbance, a = absorptivity, b = path length, c = concentration,  = molar absorptivity, extinction coefficient An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry
  • 17. The Quantitative Picture Where the absorbance A has no units, since A = log10 P0 / P  Is the molar absorbtivity with units of L mol-1 cm-1 b is the path length of the sample in cm c is the concentration of the compound in solution, expressed in mol L-1 (or M, molarity)
  • 18. The Beer-Lambert Law (Beer’s Law): A =  b c
  • 19. • Monochromatic incident radiation (all molecules absorb light of one ) • Absorbers independent (Absorbing molecules act independently of one another i.e., low c) • Path length is uniform (all rays travel the same distance in sample) • No scattering • Absorbing medium is optically homogeneous • Incident beam is not large enough to cause saturation • All rays should be parallel to each other and perpendicular to surface of medium Assumptions in derivation of Beer’s Law
  • 20. Deviations from Beer’s Law • Real Limitations • Beer’s law is successful in describing the absorption behavior of dilute solutions only ; in this sense it is a limiting law. At high concentrations ( > 0.01M ),the average distance between the species responsible for absorption is diminished to the point where each affects the charge distribution of its neighbors This interaction, in turn, can alter the species’ ability to absorb at a given wavelength of radiation thus leading to a deviation from Beer’s law [2]. A similar effect is sometimes encountered in solutions containing low absorber concentrations and high concentrations of other species, particularly electrolytes. Beer’s law is valid at low concentrations, but breaks down at higher concentrations For linearity, A < 1
  • 21. Deviations from Beer's law also arise because ε is dependent upon the refractive index of the solution. Thus, if concentration changes cause significant alterations in the refractive index η of a solution, departures from Beer's law are observed. A correction for this effect can be made by substitution of the quantity εη /(η2 + 2)2 for ε in the Beer equation, as shown below [2]: A= [εηb /(η2 + 2)2 ] bC In general, this correction is never very large and is rarely significant at concentrations less than 0.01 M.
  • 22. • Chemical deviations from Beer’s law are caused by shifts in the position of a chemical or physical equilibrium involving the absorbing species. A common example of this behavior is found with acid/base indicators. Deviations arising from chemical factors can only be observed when concentrations are changed Chemical Deviations HIn Ka H+ + In- Red,  =600nm Colorless Phenolphthalein: If solution is buffered, then pH is constant and [HIn] is related to absorbance. But, if un buffered solution, equilibrium will shift depending on total analyte concentration C HIn Expected Actual
  • 23. • instrument may be caused by fluctuations in the power-supply voltage, an unstable light source, or a non-linear response of the detector-amplifier system • Polychromatic Radiation • All monochromators, regardless of quality and size, have a finite resolving power and therefore minimum instrumental bandwidth. Polychromatic radiation (i.e., light of more than one ) Instrumental Factors
  • 24. Band A shows little deviation, because ε does not change greatly throughout the band. Band B shows marked deviations because ε undergoes significant changes in this region
  • 25. The width of the image produced is thus an important measure of the quality of the performance of a spectrometer. The figure below shows the loss of detail that accompanies the use of wider slits. It is evident that an increase in slit width brings about a loss of spectral detail Effect of Slit Width on Absorbance Measurements
  • 26. • quantitative measurement of narrow absorption bands demand the use of narrow slits widths. Unfortunately, a decrease in slit width is accompanied by a second-order power reduction in the radiant energy; at very narrow settings spectral detail may be lost owing to an increase in the signal-to-noise ratio. In general, it is good practice to narrow slits no more than is necessary for good resolution for the spectrum at hand Another effect of slit width is the change of absorbance values that accompany a change in the slit width. The figure below illustrates this effect. Note that the peak absorbance values increase significantly (by as much as 70% in one instance) as the slit width decreases
  • 27. The basic components of analytical instruments for absorption spectroscopy, as well as for emission and fluorescence spectroscopy, are remarkably alike in function and in general performance requirements whether the instruments are designed for ultraviolet (UV), visible, or infrared (lR) radiation. We often call the UV/visible and IR regions of the spectrum the optical region Instruments for Optical Spectrometry
  • 28. 1) a stable source of radiant energy 2) a wavelength selector that isolates a limited region of the spectrum for measurement 3) one or more sample containers 4) a radiation detector, which converts radiant energy to a measurable electrical signal 5) a signal processing and readout unit, usually consisting of electronic hardware and, in modern instruments, a computer. INSTRUMENT COMPONENTS
  • 29. To be suitable for spectroscopic studies, a source must generate a beam of radiation that is sufficiently powerful to allow easy detection and measurement. In addition, its output power should be stable for reasonable periods of time. Typically, for good stability, a well- regulated power supply must provide electrical power for the source Spectroscopic Sources Spectroscopic sources are of two types: A- Continuum sources, which emit radiation that changes in intensity only slowly as a function of wavelength. B- Line sources, which emit a limited number of spectral lines, each of which spans a very limited wavelength range. The distinction between these sources is illustrated in Figure bellow. Sources can also be classified as continuous sources, which emit radiation continuously with time, or pulsed sources, which emit radiation in bursts
  • 30. tungsten lamp of the type used in spectroscopy and its spectrum (b). Intensity of the tungsten source is usually quite low at wavelengths shorter than about 350 nm. Note that the intensity reaches a maximum in the near-IR region of the spectrum (<1200 nm in this case). Tungsten lamp
  • 31. • Tungsten/halogen lamp • Tungsten/halogen lamps, also called quartz/halogen lamps, contain a small amount of iodine within the quartz envelope that houses the filament. Quartz allows the filament to be operated at a temperature of about 3500 K, leading to higher intensities and extending the range of the lamp well into the UV. The lifetime of a tungsten/halogen lamp is more than double that of an ordinary tungsten lamp, which is limited by sublimation of tungsten from the filament. In the presence of iodine, the sublimed tungsten reacts to give gaseous WI2 molecules. These molecules then diffuse back to the hot filament where they decompose, redeposit W atoms on the filament, and release iodine. Tungsten/halogen lamps are finding ever-increasing use in spectroscopic instruments because of their extended wavelength range, greater intensity, and longer life
  • 32.
  • 33. A deuterium lamp consists of a cylindrical tube containing deuterium at low pressure, with a quartz window from which the radiation exits Deuterium lamp
  • 34. The word laser originally was the upper-case LASER, the acronym from Light Amplification by Stimulated Emission of Radiation. Lasers have become useful as sources in certain types of analytical spectroscopy. To understand how a laser works, consider an assembly of atoms or molecules interacting with an electromagnetic wave. Laser radiation is highly directional. Spectrally pure, coherent. and of high intensity. These properties have made possible many unique research applications that cannot easily be achieved with conventional sources Laser Sources
  • 35. Spectroscopic instruments in the UV and visible regions are usually equipped with one or more devices to restrict the radiation being measured to a narrow band that is absorbed or emitted by the analyte. Such devices greatly enhance both the selectivity and the sensitivity of an instrument. In addition, for absorption measurements, narrow bands of radiation greatly diminish the chance of Beer's Jaw deviations due to polychromatic radiation. Many instruments use a monochromator or filter to isolate the desired wavelength band so that only the band of interest is detected and measured. Others use a spectrograph to spread out, or disperse, the wavelength so that they can be detected with a multichannel detector Wavelength Selectors
  • 36. Monochromators generally have a diffraction grating to disperse the radiation into its component wavelengths, as shown in Figure bellow By rotating the grating, different Wavelengths can be made to pass through an exit slit Monochromators consist of: 1- entrance slit 2- collimating mirror or lens 3- a prism or grating 5- focal plane 6- exit slit Older instruments used prisms for this purpose Monochromators
  • 37. Older instruments used prisms for this purpose
  • 38. Radiation from a source enters the monochromator via a narrow rectangular opening, or slit. The radiation is then collimated by a concave mirror, which produces a parallel beam that strikes the surface of a reflection grating. Angular dispersion results from diffraction, which occurs at the reflective surface Grating Monochromator
  • 39. The effective band width of the monochromator depends on the size and quality of the dispersing element, the slit widths, and the focal length of the monochromator. A high-quality rnonochromator will exhibit an effective band-width of a few tenths of a nanometer or less in the ultraviolet/visible region. The effective bandwidth of a monochromator that is satisfactory for most quantitative applications is about 1 to 20 nm. Many monochromators are equipped with adjustable slits to permit some control over the bandwidth. A narrow slit decreases the effective bandwidth but also diminishes the power of the emergent beam. Thus, the minimum practical bandwidth may be limited by the sensitivity of the detector For qualitative analysis, narrow slits and minimum effective bandwidths are required if a spectrum is made up of narrow peaks. For quantitative work. however, wider slits permit operation of the detector system at lower amplification, which in turn provides greater reproducibility of response.
  • 40. Most gratings in modern monochromators are replica gratings, which are obtained by making castings of a master grating. The latter consists of a hard, optically flat, polished surface on which a suitably shaped diamond tool has created a large number of parallel and closely spaced grooves. A grating for the ultraviolet and visible region will typically contain 300 to 2000 grooves/mm. with 1200 to 1400 being most common. The construction of a good master grating is tedious, time consuming, and expensive because the grooves must be identical in size, exactly parallel, and equally spaced over the length of the grating (3 to 10 cm). Replica gratings are formed from a master grating by a liquid resin casting process that preserves virtually perfectly the optical accuracy of the original master grating on a clear resin surface. This surface is ordinarily made reflective by a coating of aluminum or, some times, gold or platinum
  • 41.
  • 42. Radiation Filters Filters operate by absorbing all but a restricted band of radiation from a continuum source. As shown in Figure bellow, two types of filters are used in spectroscopy; interference filters and absorption filters. Interference filters are typically used for absorption measurements, and they generally transmit a much greater fraction of radiation at their nominal wavelengths than do absorption filters
  • 43.
  • 44. Interference Filters Interference filters are used with ultraviolet and visible radiation, as well as with wavelengths up to about 14 µm in the infrared region. As the name implies, an interference filter relies on optical interference to provide a relatively narrow band of radiation. typically 5 to 20 nm in width. As shown in Figure bellow, an interference filter consists of a very thin layer of a transparent dielectric material (frequently calcium fluoride or magnesium fluoride) coated on both sides with a film of metal that is thin enough to transmit approximately half the radiation striking it and to reflect the other half.
  • 45.
  • 46. • the radiant power transmitted, fluoresced, or emitted must be detected in some manner and converted into a measurable quantity. A detector is a device that indicates the existence of some physical phenomenon. Familiar examples of detectors include photographic film (for indicating the presence of electromagnetic or radioactive radiation. • The human eye is also a detector; it converts visible radiation into an electrical signal that is passed to the brain via a chain of neurons in the optic nerve and produces vision. Detecting and Measuring Radiant Energy
  • 47. • The term transducer is used to indicate the type of detector that converts quantities, such as light intensity, pH, mass, and temperature, into electrical signals that can be subsequently amplified, manipulated, and finally converted into numbers proportional to the magnitude of the original quantity. • A transducer is a type of detector that converts various types of chemical and physical quantities into electrical signals such as electrical charge, current, or voltage.
  • 48. There are two general types of transducers: • Photons All photon detectors are based on the interaction of radiation with a reactive surface either to produce electrons (photoemission) or to promote electrons to energy states in which they can conduct electricity (photoconduction). Only UV, visible, and near-IR radiation possess enough energy to cause photoemission to occur; thus, photoemissive detectors are limited to wavelengths shorter than about 2 µm (2000 nm). Types of Transducers
  • 49. • Thermal detectors (Heat) detect a temperature change in a material due to photon absorption • Thermal detectors can be used over a wide range of wavelengths .Their main disadvantages are slow response time and lower sensivity relative to other types of detectors.
  • 50.
  • 51. Widely used types of photon detectors include phototubes, photomultiplier tubes, silicon photodiodes, and photodiode arrays. Photon Detectors Phototubes The response of a phototube or a photomultiplier tube is based on the photoelectric effect. a phototube consists of a semicylindrical photocathode and a wire anode sealed inside an evacuated transparent glass or quartz envelope. The concave surface of the cathode supports a layer of photoemissive material. such as an alkali metal or a metal oxide. that emits electrons when irradiated with light of the appropriate energy. When a voltage is applied across the electrodes, the emitted photoelectrons are attracted to the positively
  • 52. charged wire anode, In the complete circuit, a photocurrent then results that is easily amplified and measured. The number of photoelectrons ejected from the photocathode per unit time is directly proportional to the radiant power of the beam striking the surface. With an applied voltage of about 90 V or more, all these photoelectrons are collected at the anode to give a photocurrent that is also proportional to the radiant power of the beam. Photoelectrons: are electrons that are ejected from a photosensitive surface by electromagnetic radiation. Photocurrent: is the current in an external circuit that is limited by the rate of ejection of photoelectrons.
  • 53. The photomultiplier tube (PMT) is similar in construction to the phototube but is significantly more sensitive. Its photocathode is similar to that of the phototube, with electrons being emitted on exposure to radiation. In place of a single wire anode. however, the PMT has a series of electrodes called dynodes, The electrons emitted from the cathode are accelerated toward the first dynode. which is maintained 90 to 100 V positive with respect to the cathode. Each accelerated photoelectron that strikes the dynode surface produces several electrons, called secondary electrons, that are then accelerated to dynode 2, which is held 90 to 100 V more positive than dynode l. Again, electron amplification results. By the time this process has been repeated at each of the dynodes, 105 to 107 electrons have been produced for each incident photon. This cascade of electrons is finally collected at the anode to provide an average current that is further amplified electronically and measured.
  • 54.
  • 55. One of the major advantages of photomultipliers is their automatic internal amplification. About 106 to 107 electrons are produced at the anode for each photon that strikes the photocathode of a photomultiplier tube. Photomultiplier tubes are among the most widely used types of transducers for detecting ultraviolet/visible radiation. With modern electronic instrumentation, it is possible to detect the electron pulses resulting from the arrival of individual photons at the photocathode of a PMT. The pulses are counted, and the accumulated count is a measure of the intensity of the electromagnetic radiation impinging on the PMT. Photon counting is advantageous when light intensity, or the frequency of arrival of photons at t1e photocathode, is low.
  • 56. Photoconductive transducers consist of a thin film of a semiconductor material, such as lead sulfide, mercury cadmium telluride (MCT), or indium antimonide, deposited often on a nonconducting glass surface and sealed in an evacuated envelope. Absorption of radiation by these materials promotes nonconducting valence electrons to a higher energy state, which decreases the electrical resistance of the semiconductor. Typically, a photoconductor is placed in series with a voltage source and a load resistor, and the voltage drop across the load resistor serves as a measure of the radiant power of the beam of radiation. The PbS and InSb detectors are quite popular in the near-IR region of the spectrum. The MCT detector is useful in the mid- and far-IR regions when cooled with liquid N2 to minimize thermal noise. Photoconductive Cells:
  • 57. Diode-Array Detectors Silicon photodiodes have become important recently because 1000 or more can be fabricated side by side on a single small silicon chip. (The width of individual diodes is about 0.02 mm). With one or two of the diode-array detectors placed along the length of the focal plane of a monochromator. All wavelengths can be monitored simultaneously, thus making high-speed spectroscopy possible. Silicon photodiode detectors respond extremely rapidly, usually in nanoseconds.
  • 58. They are more sensitive than a vacuum phototube but considerably less sensitive than a photomultiplier tube.
  • 59. Sample containers, which are usually called cells or cuvettes must have windows that are transparent in the spectral region .Thus as shown in the figure below, quartz or fused silica is required for the UV region (wavelengths less than 350nm nm) and may be used in the visible region and out to about 3000 nm (3 µm) in the IR region. Silicate glass is ordinarily used for the 375 to 2000 nm region because of its low cost compared with quartz. Plastic cells are also used in the visible region. The most common window material for IR studies is crystalline sodium chloride, which is soluble in water and in some other solvents. Sample Containers
  • 60. The best cells have windows that are perpendicular to the direction of the beam in order to minimize reflection losses. The most common cell path length for studies in the UV and visible regions is 1 cm; matched, calibrated cells of this size are available from several commercial sources. Many other cells with shorter and longer path lengths can be purchased.
  • 61. For reasons of economy, cylindrical cells are sometimes used. Fingerprints, grease, or other deposits on the walls markedly alter the transmission characteristics of a cell. Thus, thorough cleaning before and after use is need , and care must be taken to avoid touching the windows after cleaning is complete. Matched cells should never be dried by heating in an oven or over a flame because this may cause physical damage or a change in path length..
  • 62. Photometers provide simple, relatively inexpensive tools for performing absorption measurements. Filter photometers are often more convenient and more rugged and are easier to maintain and use than the more sophisticated spectrophotometers. Furthermore, photometers characteristically have high radiant energy throughputs and thus good signal-to-noise ratios even with relatively simple and inexpensive transducers and circuitry. Photometers have the advantages of simplicity, ruggedness, and low cost.
  • 63. Spectrophotometers offer the considerable advantage that the wavelength can be varied continuously, thus making it possible to record absorption spectra and is a scanning instrument (i.e. a spectrophotometer has a monochromator for separating the individual wavelengths of light). Several dozen models of spectrophotometers are available commercially. Most spectrophotometers cover the UV /visible and occasionally the near-infrared region, while photometers are most often used for the visible region.. Both photometers and spectrophotometers can be obtained in single- and double-beam varieties.
  • 64. Single beam spectrometers Single beam spectrometers are relatively cheap, simple, portable & ideally suited to quantitative analysis .It is not possible to scan through the entire spectrum with such an instrument because both the source intensity & the detector response vary with the wavelength. To record an accurate value of the absorbance it is necessary to zero the instrument on a reference/blank before every measurement. Thus, this is essentially a single wavelength measurement of absorbance.
  • 65.
  • 66. Voltage fluctuations and changes in light source present a problem When a heavy load is placed on the electric power system, lights dim and later brighten *If measurements are being taken on the spectrophotometer at the same time, the readings will be unreliable *Aging lamp source may momentarily flicker and cause the readings to be unstable and errorneous So, a single-beam instrument requires a stabilized voltage supply to avoid errors resulting from changes in the beam intensity during the time required to make the 100% T measurement and determine %T for the analyte. Disadvantages of single-beam spectrophotometer
  • 67. The second type of double-beam instrument. Here the beams are separated in time by a rotating sector mirror that directs the entire beam from the monochromator first through the reference cell and then through the sample cell. The pulses of radiation are recombined by another sector mirror, which transmits one pulse and reflects the other to the transducer. As shown by the insert labeled "front view", the motor-driven sector mirror is made up of pie- shape segments, half of which are mirrored and half of which are transparent. The mirrored sections are held in place by blackened metal frames that periodically interrupt the beam and prevent its reaching the transducer. The double-beam-in-time approach is generally preferred because of the difficulty in matching the two detectors needed for the double-beam-in-space design Double-Beam in time spectrophotometers
  • 68.
  • 69. Double-Beam in space spectrophotometers • Many modern photometers and spectrophotometers are based on a double-beam design. A double-beam-in-space instrument in which two beams are formed in space by a V-shape mirror called a beamsplitter. One beam passes through the reference solution to a photodetector, and the second simultaneously traverses the sample to a second, matched detector. The two outputs are amplified, and their ratio (or the logarithm of their ratio) is determined electronically or by a computer and displayed by the readout device.
  • 70.
  • 71. The determination of an analyte’s concentration based on its absorption of ultraviolet or visible radiation is one of the most frequently quantitative analytical methods. One reason for its popularity is that many organic and inorganic compounds have strong absorption bands in the UV/Vis region of the electromagnetic spectrum. There are many application • Environmental Applications • Clinical Applications • Industrial Analysis • Forensic Applications Quantitative Applications
  • 72. • Methods for the analysis of waters and wastewaters relying on the absorption of UV/Vis radiation are among some of the most frequently employed analytical methods. Environmental Applications
  • 73. • UV/Vis molecular absorption is one of the most commonly employed techniques for the analysis of clinical samples, several examples of which are listed in Table below. The analysis of clinical samples is often complicated by the complexity of the sample matrix, which may contribute a significant background absorption at the desired wavelength . Clinical Applications
  • 74. The Application of UV/Vis Molecular Absorption to the Analysis of Clinical Samples
  • 75. • UV/Vis molecular absorption is used for the analysis of a diverse array of industrial samples, including pharmaceuticals, food, paint, glass, and metals. • In many cases the methods are Products that have been analyzed in this fashion include antibiotics, hormones, vitamins, and analgesics. • One example of the use of UV absorption is in determining the purity of aspirin tablets. Industrial Analysis
  • 76. • UV/Vis molecular absorption is routinely used in the analysis of narcotics and for drug testing. • One interesting forensic application is the determination of blood alcohol using the Breathalyzer test. In this test a 52.5-mL breath sample is bubbled through an acidified solution of K2Cr2O7. Any ethanol present in the breath sample is oxidized by the dichromate, producing acetic acid and Cr3+ as products. Forensic Applications
  • 77. • The energy at which the absorption occurs, as well as the intensity of the absorption, is determined by the chemical environment of the absorbing moiety. For example, benzene has several ultraviolet absorption bands due to p  p* transitions. The position and intensity of two of these bands, 203.5 nm (e = 7400) and 254 nm (e = 204), are very sensitive to substitution. For benzoic acid, in which a carboxylic acid group replaces one of the aromatic hydrogens, the two bands shift to 230 nm (e = 11,600) and 273 nm (e = 970). Several rules have been developed to aid in correlating UV/Vis absorption bands to chemical structure. Qualitative Applications
  • 78. 1-Limited since few resolved peaks • Unambiguous identification not usually possible. Why we cannot use UV / visible Spectroscopy in Qualitative Analysis ? Loss of fine structure for acetaldehyde when transfer to solvent from gas phase Also need to consider absorbance of solvent
  • 79. 2. Solvent can affect position and shape of curve . • polar solvents broaden out peaks, eliminates fine structure. •Loss of fine structure for 1,2,4,5-tetrazine as solvent polarity increases
  • 80. 3.Solvent can also absorb in UV-vis spectrum. • Solvent for the ultraviolet and visible regions
  • 81. • Molecular absorption, particularly in the UV/Vis range, has been used for a variety of different characterization studies, including determining the stoichiometry of metal– ligand complexes and determining equilibrium constants. • Stoichiometry of a Metal, Ligand Complex • The stoichiometry for a metal–ligand complexation reaction of the following general form • can be determined by one of three methods: the method of continuous variations, •the mole-ratio method, and the slope-ratio method Characterization Applications
  • 82. • also called Job’s method, is the most popular. In this method a series of solutions is prepared such that the total moles of metal and ligand, ntot, in each solution is the same. Thus, if (nM)i and (nL)i are, respectively, the moles of metal and ligand in the i-th solution, Then • The relative amount of ligand and metal in each solution is expressed as the mole • fraction of ligand, (XL)i, and the mole fraction of metal, (XM)i, Method of continuous variations
  • 83. A plot of A vs volume ratio (volume ratio = mole fraction) gives maximum absorbance when there is a stoichiometric amount of the two.
  • 84. • A procedure for determining the stoichiometry between two reactants by preparing solutions containing different mole ratios of two reactants . In the mole-ratio method the moles of one reactant, usually the metal, are held constant, while the moles of the other reactant are varied. The absorbance is monitored at a wavelength at which the metal–ligand complex absorbs. A plot of absorbance as a function of the ligand-to- metal mole ratio (nL/nM) has two linear branches that intersect at a mole ratio corresponding to the formula of the complex. Mole-ratio method
  • 85. • Figure (a)shows a mole-ratio plot for the formation of a 1:1 complex in which the absorbance is monitored at a wavelength at which only the complex absorbs.
  • 86. • Figure (b) shows a mole-ratio plot for a 1:2 complex in which the metal, the ligand, and the complex absorb at the selected wavelength.
  • 87. • A procedure for determining the stoichiometry between two reactants by measuring the relative change in absorbance under conditions when each reactant is the limiting reagent . In the slope-ratio method two sets of solutions are prepared. The first set consists of a constant amount of metal and a variable amount of ligand, chosen such that the total concentration of metal, CM, is much greater than the total concentration of ligand, CL. Under these conditions we may assume that essentially all the ligand is complexed. The concentration of a metal–ligand complex of the general form MxLy is Slope-ratio method
  • 88. • Determine endpoint by following change in absorbance of: 1) reactant (decrease) 2) product (increase) 3) titrant (increase after endpoint) • Example Titration curves for • 𝑺 + 𝑻 → 𝑷 • where S = analyte being titrated, T = titrant, P = product Photometric Titrations
  • 89. where S = analyte being titrated, T = titrant, P = product
  • 90. References 1- Douglas A. skoog, Donald M. West, F. james Holler, Stanley R. Crouch. (2013). Fundamentals of analytical chemistry. 9th ed . Belmont, USA: Mary finch. pp:651-674. 2- Douglas A. Skoog, F. James Holler, Stanley R. Crouch. (2007). Principles of instrumental analysis. 6th ed. Belmont, USA: David Harris.pp:335-367. 3- Tony O.(2000). Fundamentals of modern Uv-Visible spectroscopy. 1st ed. Germany: Agilent technologies .pp:36-43. 4- David H.(2000). Modern analytical chemistry. 1st ed . London: James M smith. Pp: 394- 407.

Hinweis der Redaktion

  1. How much
  2. Slaw hawreian
  3. An aqueous solution of the complex Fe (SCN)2+ is not red because the complex adds red radiation to the solvent. Instead, it absorbs green from the incoming white radiation and transmits the red component .Thus, in a colorimetric determination of iron based on its thiocyanate complex, the maximum change in absorbance with concentration occurs with green radiation; the absorbance change with red radiation is negligible. In general, then, the radiation used for a colorimetric analysis should be the complementary color of the analyte solution. The following table shows this relationship for various parts of the visible spectrum
  4. Violet = yellow Blue= orange Green = red
  5. Matched cells should be calibrated against each other regularly with an absorbing solution
  6. Photometer in the visible region of EM radiation is called colorimeter