2. CONTENTS
ISOELECTIC FOCUSSING
PRINCIPLE
THEORETICAL ASPECTS
PREPARATION OF IEF GEL
DETERMINATION OF ISOELECTRIC POINT OF
PROTEIN
APPLICATION
ISOTACHOPHORESIS
PRINCIPLE
APPLICATION.
3. ISOELECTRIC FOCUSSING
Electrophoretic method that separates proteins
according to the iso-electric points.
Is ideal for separation of amphoteric substances.
Separation is achieved by applying a potential
difference across a gel that contain a pH gradient.
Isoelectric focusing requires solid support such as
agarose gel and polyacrylamide.
4. Isoelectric focusing gels contain synthetic buffers
called ampholytes that smooth the pH gradients.
Ampholytes are complex mixtures of synthetic
polyamino-polycarboxylic acids
Commercially available ampholytes are-
BIO-LYTE
PHARMALYTE
5. It gives good separation with a high resolution
compared to any other method
Resolution depends on
I. The pH gradient,
II. The thickness of the gel
III. Time of electrophoresis,
IV. The applied voltage,
V. Diffusion of the protein into the gel
6. pI:
Isoelectric focusing uses the theory of
protein pI.
pI is the pH at which a given protein has a
neutral overall charge.
The pI is dependant on which type of
residues are present and how many bases
make proteins positive and acids
negative.
pI is very specific for each protein.
7. PRINCIPLE
Isoionic point is the point at which the net charge of protein
molecule is zero.
In isoelectric focusing a stable pH gradient established.
pH increases gradually from anode to cathode.
A protein introduced at a point where the pH is lower than
the isoionic point will posses a net positive charge & will
migrate in the direction of cathode.
Due to the presence of protein gradient, the protein will
migrate to an environment of higher pH values .
8. Iso electric point of protein varies with the
environment.
Isoelectric point is taken to be the pH where all the an
ions on the protein are exactly balanced by equal
number of cations;the net charge is zero & the protein
does not migrate in the electric field.
The value of pH provides a rough idea about the
amino acid composition of a given protein.
•Finally protein will encounter a pH where its net charge
is zero & will stop migrating.
•This is the isoelectric point of protein.
9.
10. THEORETICAL ASPECTS.
The pH gradient forces a protein to remain in its isoelectric
point position, thus concentrating it ; this concentrating effect
is called "focusing".
Increasing the applied voltage or reducing the sample load
result in improved separation of bands. The applied voltage is
limited by the heat generated, which must be dissipated.
The use of thin gels and an efficient cooling plate controlled by
a thermostatic circulator prevents the burning of the gel whilst
allowing sharp focusing.
The separation is estimated by determining the minimum pI
difference (ΔpI), which is necessary to separate 2 neighboring
bands:
13. PREPARATION OF IEF GEL
Carrier ampholytes (suitable pH) and riboflavin mixed with acrylamide
solution.
Mixture is poured over a glass plate which contain spacer
Second glass plate is placed on first
Gel is polymerized
14. This takes 2-3 hr
After the gel has set glass plates are prised apart
Electrode wicks are laid along the long length of each side
of the gel
Potential difference is applied
Ampholytes form a pH gradient between anode and
cathode
15. The power is then turned off
Samples applied by laying on gel filter paper soaked in the
sample
Voltage is again applied for 30 min
Proteins having positive charge will migrates towards the
cathode. negatively charged protein will migrates towards
anode
Become stationary when they reaches isoelectric point
16. The gel is washed with trichloroacetic acid
This precipitaes the proteins and allows smaller ampholytes to
be washed out
Gel is stained with Coomasie Brilliant Blue
Destained
18. DETERMINATION OF ISOELECTRIC
POINT OF PROTEIN.
To determine isoelectric point of protein (polypeptide),one has
to:
i. Measure the mobility of protein at several different pH values.
ii. Plot the mobility values so obtained against the pH values.
iii. Take reading of pH of that point where the point intersects
zero mobility line.
The pH at this intersection is then taken to be the isoelectric
point of the concerned protein.
20. APPLICATION
Useful for separating, purifying & characterizing
proteins.
Identification of iso enzymes.
Separation & identification of serum proteins.
Used by food & agricultural industries, forensic &
human genetic laboratories.
Used for research in enzymology, immunology &
membrane biochemistry.
21. ISO-TACHOPHORESIS
This method is based on the principle of moving boundary
electrophoresis.
Separation is achieved either in a horizontal or vertical way
based on the equipment employed.
The name iso-tachophoresis is derived from Greek , refers
to the fact that the ions being separated all travel (phoresis)
at the same( iso)speed( tacho).
Separation of ionic components of sample is achieved
through stacking them into discrete zones in order of their
mobilities, producing very high resolution.
22. Migration of small ions
Discontinuous electrolyte system
leading electrolyte (L- ions) &
trailing electrolyte (T- ions)
Apply sample solution at interphase of L & T
Apply electric field -> each type of ion arrange
between L and T ions -> discrete zones
Separates small anions, cations, organic & amino
acids, peptides, nucleotides, nucleosides, proteins
23. Separation of mixture of an ions involves:
A leading anion (chloride) with higher mobility than the sample ion.
A trailing an ion (glutamate) with lower mobility than the sample ion.
A common cat ion for all the anions.
When the current is switched on, the leading ions will move towards
appropriate electrode.
The sample ions will follow in order of their mobilities .
The trailing will follow the sample ions.
Once equilibrium is achieved ions will all move at the same speed in
discrete bands in order of their mobilities.
24.
25. Where the mobilities of sample ions are similar, the
resolution can be enhanced by including with the
sample, synthetic ampholytes called spacer ions.
These have mobilities intermediate to that of sample
ions and help to separate them by taking up positions
between the sample ions.
The spacer ions are similar to ampholytes used in
isoelectric focusing.
26. APPLICATION
Used for separating charged substances ranging from inorganic to
organic acids to proteins and nucleic acids.
Used extensively in research laboratories.
Find applications in industries in pollution control(detecting detergents
and inorganic ions in effluent water)
In quality control in the food, brewing and pharmaceutical industries.
Samples as small as few micrograms can be separated quantitatively,&
large quantities of sample can be separated preparatively .